• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMBLEM MRI S-ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMBLEM MRI S-ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number A219
Device Problems Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); Failure to Select Signal (1582); Device Sensing Problem (2917)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/19/2019
Event Type  malfunction  
Manufacturer Narrative
This device is still currently in service and not available for return.
 
Event Description
It was reported that the patient had previously experienced inappropriate shocks due to t-wave oversensing while exercising.The system was then programmed to primary sensing vector.Recently, the patient received two additional inappropriate shocks and the system was subsequently programmed to secondary vector.No additional adverse events.Kn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM MRI S-ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key9921190
MDR Text Key186323443
Report Number2124215-2020-00620
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526584404
UDI-Public00802526584404
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110042/S058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/26/2021
Device Model NumberA219
Device Catalogue NumberA219
Device Lot Number244676
Was Device Available for Evaluation? No
Date Manufacturer Received01/08/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/03/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age29 YR
-
-