This is being filed to report the torn tip.
It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.
One clip was successfully implanted.
A second clip delivery system (cds) was advanced to the mitral valve however the mr was unable to be reduced further.
The leaflets were ungrasped and the clip repositioned several times however the mr was unable to be reduced.
The cds was retracted per the instructions for use (ifu) however the clip became caught on the tip of the steerable guide catheter (sgc).
Standard troubleshooting was performed; however the clip could still not be retracted.
The system was pulled back to the right side and pulled through the septum without issue.
It was noted the clip looked like it had detached from the mandrel even though no deployment had been initiated.
The sgc and cds were retracted to the groin where it was noted one clip arm was inside the sgc and one clip arm was stuck on the vessel wall.
A vascular surgeon was called to perform a cut down.
Once removed, it was confirmed the clip was no longer attached to the mandrel but was still attached to the cds by the lock and gripper lines.
The procedure was completed at this time with the mr reduced to 3.
There was no clinically significant delay in the procedure and no adverse patient sequela.
Return analysis noted the tip of the sgc was torn.
No additional information was provided.
|