Device evaluated by mfr: return product consisted of a runway guide catheter and loose fm.
Analysis of the tip, shaft and hub included microscopic and visual inspection.
Inspection found no damage to the device.
The returned foreign matter (fm) was tangled together and could not be measured accurately.
The fm was identified as fluorocarbon fiber, which is consistent with ptfe lining.
|
It was reported that foreign matter was present on the device.
A model 6f runway jl3.
5 guide catheter was selected for use in a percutaneous coronary intervention (pci).
During the procedure, when the physician inserted the guide catheter, there was a string like a nylon coming out of the guide catheter.
The physician quickly pulled the guide catheter out from the radial.
The procedure was completed with a different device.
No patient complications were reported and the patient was stable post procedure.
|