Model Number URF-P6 |
Device Problem
Material Protrusion/Extrusion (2979)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/12/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The scope has been returned and is pending evaluation.As additional information becomes available following device evaluation, a supplemental report will be filed.
|
|
Event Description
|
The service center was informed that during reprocessing metal was noted to be sticking out of the scope¿s bending section.The only other issue noted during inspection was also the bending rubber.Additionally, it was reported that a leak test is being performed after each use of the scope.Device is reprocessed manually.There was no patient involvement.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.Please see the updates in sections.Upon further review this is not a reportable malfunction.Per the legal manufacturer there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
|
|