Model Number 3721 |
Device Problem
Temperature Problem (3022)
|
Patient Problem
Discomfort (2330)
|
Event Date 03/26/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The investigation for capa 118058 associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.Date of event is estimated the results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
Related manufacturer reference number: 1627487-2020-03763.It was reported that the patient experienced heating while charging.As a result, surgical intervention may be pending.
|
|
Manufacturer Narrative
|
During processing of this complaint, attempts were made to obtain complete patient information.
|
|