Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An exterior visual inspection of the returned cycler showed no signs of physical damage.Upon power up, the function display test failed.The display was dim upon powering on the cycler.It was identified that the cause for the dim display was due to an internal short present transformer (t1) on the inverter board.A known good inverter board was installed, and the display became fully operational.An internal visual inspection of the returned cycler encountered no other discrepancies.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed, and the cause was determined to be an internal short on transformer on the inverter board.The cycler was refurbished following the evaluation.
|
It was reported that the screen of a patient¿s liberty select cycler had a display brightness problem during setup.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.It was reported that an alternate treatment option was not available.Upon follow up, the patient confirmed that there were no adverse events or medical intervention required as a result of the reported event.The patient completed treatment on the cycler.The cycler was returned to the manufacturer and a replacement cycler was provided and received.Upon physical evaluation of the cycler by the manufacturer, evidence of an internal short on the transformer on the inverter board was identified.
|