The device was received for evaluation.Evaluation determined that the a rubber has a cut and was leaking.The bending section was found to have a broken skeleton, metal was sticking out.Leak testing failed due to the cut found on the a rubber.Device was placed for repair.As stated on the ifu and as a preventive measure, the user manual states ,do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, video connector, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.
|