SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL SAW, POWERED, AND ACCESSORIES
|
Back to Search Results |
|
Model Number 72200616 |
Device Problems
Overheating of Device (1437); Electrical Shorting (2926)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/03/2020 |
Event Type
Malfunction
|
Manufacturer Narrative
|
The reported device, used in treatment, was received for evaluation.
There was a relationship found between the returned device and the reported incident.
A visual inspection was performed on the exterior of product and no physical damage was observed.
Mdu failed functional tests with hand piece overload when power cord was bent or twisted.
Power cord assembly grew warm during testing.
After troubleshooting, the cause of the overheating was observed to be a defective power cord assembly.
A visual inspection was performed on the power cords external covering and no physical damage was observed.
It was determined that the power cord has shorted and/or opens internal wiring.
Motor and hall board were tested and passed functional testing.
The complaint investigation has concluded that this unit has succumbed to physical damage to power cord.
Factors which can contribute to a shorted power cord include rolling a heavy cart over the cord or excessive bending of the cord.
A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions.
|
|
Event Description
|
It was reported that the buttons were stuck a few times during knee scope surgery.
A backup device was available to complete the procedure with no delay or patient injuries.
Results of investigation have concluded that this unit presented a short in the cable, which makes it a reportable event.
All available information has been disclosed.
If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|