• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-15
Device Problem Pumping Problem (3016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/09/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement. Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp. The incident occurred in (b)(6). A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported event. He traced the failure back to a defective processor board. He replaced the part and the issue was solved. Subsequent functional verification testing was completed without further issues and the unit was returned to service. The root cause of the event has been identified as a defective cpu board.
 
Event Description
Livanova received a report that a centrifugal pump system with tubing clamp displayed an error code associated to a processor error during a procedure and all buttons on the control panel were not working. There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of DeviceCONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9963141
MDR Text Key195923910
Report Number9611109-2020-00258
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/17/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number60-02-15
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/03/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/09/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-