Model Number 9734716 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The unique identifier was not available at the time of reporting.No parts have been received by the manufacturer for evaluation.The manufacture date was not available at the time of reporting.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation system being used in a sacroiliac and thoracolumbar procedure.It was reported that the site was having issues with the gen 3 spinous process clamps engaging during a case.The site was able to get the clamps to work for the case.There was a 5 min delay in the case.No impact on patient outcome.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received.It was reported that the design of the clamp was not allowing the screw to engage.
|
|