It was reported that this patient presented to the emergency room (er) following a syncopal episode.
A consult review performed by boston scientific technical services (ts) indicated that the left ventricular (lv) output programming on this cardiac resynchronization therapy pacemaker (crt-p) device was below the patients threshold.
Evidence is suggestive that a mistake was made in the lv output programming at the last in-office device check.
Ts also noted as part of the consult review that the right ventricular automatic threshold (rvat) test from the morning the patient presented to the er indicated successful right ventricular (rv) capture below the programmed rv output.
In response, the physician requested that a boston scientific representative (rep) reprogram the lv output.
The patient was subsequently discharged from the hospital.
The device remains in-service.
No additional adverse patient effects were reported.
|