It was reported during cardiosave intra-aortic balloon pump(iabp) therapy, the device spontaneously failed while in operation on a critical care patient.
The intra-aorta balloon pump (iabp) suddenly shutdown.
This spontaneous fail is a duplicate of an occurrence during the preventive maintenance battery test on a full battery 5 weeks ago.
After the preventive maintenance incident, the customer spoke with getinge tech support and completed 2 subsequent 60 minute discharge tests on the battery with no errors, per their instructions.
As all other tests passed during the preventive maintenance (pm), this occurred only during the battery test at the end of the pm function, calibration and operation tests.
Per the nurse the unit was plugged in at the time of failure.
This battery was fully charged both times and failed with 25 to 35 minutes of use.
Patient lost blood pressure with significant desaturation.
There was no reported malfunction on the intra-aortic balloon(iab).
|