Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (false asystole event) has been confirmed.Upon investigation the electrode belt failed an ecg fall-off test.The cause for the failure was isolated to a short in ecg "c".An unused pulse wire in the cable migrated into ecg electrode, causing a short.A root cause investigation determined that the cause of the unused wire migration in the ecg "c&d" cable was the lack of a method to secure the unused wire ends.No adverse event resulted from the defective equipment.
|