BLUE BELT TECHNOLOGIES HANDPIECE, NAVIO (BBT HANDPIECE); ORTHOPEDIC STEREOTAXIC INSTRUMENT
|
Back to Search Results |
|
Catalog Number PFSR110137 |
Device Problems
Failure to Recalibrate (1517); Deformation Due to Compressive Stress (2889)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/27/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that prior to the case starting, after handpiece calibration, the drill would not home, an error appeared indicating that there was a homing failure.All components were checked in order to know if they are correctly assembled but the error persisted.The handpiece was opened but it was not possible to move the gears.The handpiece was replaced to continue with the case normally.No surgical delay or patient injury was reported.Results of the investigation have concluded that the snaplock assembly was damaged, which makes it a reportable event.
|
|
Manufacturer Narrative
|
H10 h3, h6: the navio handpiece, intended for use in treatment, had homing failure during testing.The navio handpiece was not returned for further evaluation and a visual/functional inspection could not be performed.During an internal analysis, the problem was determined to be caused by the material used to manufacture one of the snap lock components.The material used by one of the suppliers is displaying properties after autoclave sterilization that may cause the snap lock assembly to become tight.This handpiece had a snap lock from one of the affected lots.The root cause of this issue was found to be supplier/raw material fault.A device history record review found the device met all manufacturing specifications during release for distribution.A complaint history review found similar reports.
|
|
Search Alerts/Recalls
|
|
|