Model Number CYF-VHR
Device Problem Material Separation (1562)
Patient Problem No Patient Involvement (2645)
Event Type  Malfunction  
Manufacturer Narrative

The device was returned to olympus and evaluated. The user reported problem of "distal sheath separating, distal tip damaged" was confirmed. The distal end cover was found detached. A definitive root cause for the reported problem could not be determined.

 
Event Description

Olympus was informed that a user facility's scope distal end sheath was separating and damaged. The problem reportedly occurred during reprocessing and was found during reprocessing. It was reported there was no patient involvement associated with the problem.

 
Manufacturer Narrative

This supplemental report is being submitted to provide the results of a device history record (dhr) review. The dhr was reviewed for the device involved. As a result of reviewing the dhr, it was confirmed that there were no abnormalities, special adoptions, or variations in the manufacturing process.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCYSTO-NEPHRO VIDEOSCOPE
Type of DeviceCYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key9988878
MDR Text Key222564043
Report Number8010047-2020-02334
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,USER F
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/05/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received04/22/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberCYF-VHR
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer03/23/2020
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/05/2020
Was Device Evaluated By Manufacturer? Yes
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-