• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715K 630G BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715K
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hyperglycemia (1905); Skin Irritation (2076)
Event Date 04/12/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4). Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned. No conclusion can be drawn at this time. We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call that they experienced high blood glucose. The customer¿s blood glucose level was 400 mg/dl. Customer was reporting about skin irritation issues when changing the infusion set. The customer declined troubleshoot for high blood glucose. The insulin pump will not be returned for analysis.
 
Manufacturer Narrative
Device power up properly after battery installation. Device received with operating currents within specification. Device passed selftest, rewind, seating, basic occlusion test, occlusion test, force sensor test and displacement test. Device passed delivery accuracy test. No delivery anomalies noted. (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name630G INSULIN PUMP MMT-1715K 630G BLACK MG
Type of DeviceARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key9989060
MDR Text Key189177707
Report Number2032227-2020-108259
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation
Type of Report Initial,Followup
Report Date 11/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model NumberMMT-1715K
Device Catalogue NumberMMT-1715K
Device Lot NumberHG466CU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/17/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/12/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

Patient Treatment Data
Date Received: 04/22/2020 Patient Sequence Number: 1
-
-