(b)(6).(b)(4).Siemens completed an investigation of the reported event.According to the siemens service specialists the system has been repaired by replacement of the high-voltage cables.A detailed technical investigation of the logfiles confirmed: after the repair no more scan interruptions were found.The ct scanner is end of support since end of 2019.The material consumption in relation to the installed base is monitored by the capa process (values for the last 3 months see below).No further investigation within the complaint process is deemed necessary.Mat# (b)(4) - hv cable set - march 2020: 0%, february 2020: 0%, january 2020: 1,19%.No general design issue has been identified.No further action is warranted at this time.
|