Catalog Number PFSR110004
Device Problem Defective Component (2292)
Patient Problem Injury (2348)
Event Date 02/26/2018
Event Type  Injury  
Event Description
It was reported that the handpiece would not home. The long attachment and bur were re-assembled but this did not work. The gears were moved but were not possible. The customer tried a new handpiece and did not work because the tracker array for the handpiece was bent. A different field report was submitted for that. A troubleshooting test was performed to the handpiece, failing homing as toque was too high. The procedure was changed to manual. No patient injuries reported beyond this event. Results of investigation have concluded that the snaplock assembly was broken, which makes it a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNAVIO HANDPIECE
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth, mn
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth, mn
Manufacturer Contact
richard confer
2828 liberty ave
suite 100
pittsburgh, pa 
6833844
MDR Report Key9993606
MDR Text Key188826598
Report Number3010266064-2020-00628
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 08/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberPFSR110004
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/11/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 04/23/2020 Patient Sequence Number: 1
-
-