Catalog Number NPFS02000
Device Problem Computer Software Problem (1112)
Patient Problem Missing Value Reason (3192)
Event Date 04/14/2015
Event Type  Injury  
Event Description
It was reported that during surgery hip center calculations were showing error and above 1. 0. First three were around 1. 4. Finally got good measurement. However, during normal non stressed range of motion, error presented saying issue with axis. To check measurements, tried to get through stage but had to proceed anyway. Got through remaining registration without issue until it was noticed during free collection, the m-l axis was off kilter at different angle. Continued forward and when at planning stage, the femoral solid surface 3d morphs were at 18 degree angles from the component which was corrected by rotating implant 17 degrees to match corrected bone. At this point navio case was aborted and it was completed as a manual procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNAVIO SURGICAL SYSTEM US
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd
ste 40
plymouth, mn
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest
blvd ste 40
plymouth, mn
Manufacturer Contact
richard confer
2828 liberty ave
suite 100
pittsburgh, pa 
MDR Report Key9994086
MDR Text Key188845551
Report Number3010266064-2020-00255
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue NumberNPFS02000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/29/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No Answer Provided
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 04/23/2020 Patient Sequence Number: 1
-
-