BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM AU ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number NPFS02070
Device Problem Image Orientation Incorrect (1305)
Patient Problem Injury (2348)
Event Date 05/09/2019
Event Type  Injury  
Event Description
It was reported that during tka case, distal femur and lug holes were burred like normally without any issues however when going to position the 4 in 1 cutting block the rotation was completely off (more than 6 degrees which is very unusual). Went back to check the checkpoint pins to make sure that nothing has moved and it was fine. Then returned to the distal burring screen to double check positioning of lug holes with the probe and it clearly shows that the lug holes which has been burred were in the incorrect position. Lug holes were manually done for the correct positioning of the 4 in 1 cutting block.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNAVIO SURGICAL SYSTEM AU
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth, mn
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2828 liberty ave
suite 100
pittsburgh, pa
Manufacturer Contact
richard confer
2828 liberty ave
suite 100
pittsburgh, pa 
6833844
MDR Report Key9994250
MDR Text Key188803967
Report Number3010266064-2020-01102
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue NumberNPFS02070
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/17/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 04/24/2020 Patient Sequence Number: 1
-
-