BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number PFSR110137
Device Problem Mechanical Jam (2983)
Patient Problem No Patient Involvement (2645)
Event Date 08/06/2018
Event Type  malfunction  
Event Description
It was reported that during an inspection, three handpieces were tested and the max torque was 98 for each test.Also, the gears would not move when performing the testing.No patient involved.Results of the investigation have concluded that the snaplock assembly was broken, which makes it a reportable event.
 
Manufacturer Narrative
The navio handpiece intended for use in treatment, had high t1 max torque and gears would not move during testing.A device history record review found no conditions which could contribute to the reported event and the device met all manufacturing specifications during release for distribution.A complaint history review found similar reports.The navio handpiece was returned for further evaluation and the initial visual/functional inspection was performed, which confirmed the reported event.The gears on handpiece could be moved by hand.The snap lock was broken.A broken snap lock nut would prevent the snap lock from moving along the lead screw, causing the handpiece to fail handpiece characterization.The root cause of this issue was found to be mechanical component failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth, mn
MDR Report Key9997187
MDR Text Key188892807
Report Number3010266064-2020-00688
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-