Device |
high level disinfection reprocessing instrument for ultrasonic transducers, liquid |
Regulation Description |
Diagnostic ultrasonic transducer. |
Definition |
To provide high level disinfection of ultrasound transducers If the device is reusable, validated reprocessing instructions and reprocessing validation data for this device type must be included in a 510(k) submission (82 FR 26807, available at https://www.gpo.gov/fdsys/pkg/FR-2017-06-09/pdf/2017-12007.pdf). |
Physical State |
ultrasonic transducer reprocessor, high level disinfectant liquid, and accessories |
Technical Method |
Automatically delivers liquid disinfectant and exposes a suspended ultrasound transducer to the disinfectant at a specified temperature and for a specified time to achieve high level disinfection and then rinses the ultrasound transducer with water to remove the disinfectant |
Target Area |
Ultrasonic transducers |
Regulation Medical Specialty |
Radiology |
Review Panel |
General Hospital |
Product Code | PSW |
Premarket Review |
Office of Surgical and Infection Control Devices
(OHT4)
Infection Control Devices
(DHT4C)
|
Submission Type |
510(k)
|
Regulation Number |
892.1570
|
Device Class |
2
|
Total Product Life Cycle (TPLC) |
TPLC Product Code Report
|
GMP Exempt? |
No
|
Summary Malfunction Reporting |
Ineligible |
Implanted Device? |
No
|
Life-Sustain/Support Device? |
No
|
Third Party Review |
Not Third Party Eligible |