Device |
radiology software for opportunistic evaluation of low bone mineral density |
Definition |
This device is software which opportunistically assesses radiological images to estimate bone mineral density (BMD) intended to assist in a healthcare professional's decision to evaluate patients for possible low BMD within a bone health screening program. The software employs an algorithm that estimates BMD using eligible radiological image data obtained for other clinical purposes. |
Physical State |
Software device that receives radiological image data as an input and identifies patients with possibly low BMD. It cannot be used to rule out low BMD. |
Technical Method |
Analyzes radiological data using locked image processing algorithm to estimate BMD values. Software may implement artificial intelligence algorithms. The algorithm training data typically impacts device performance. |
Target Area |
Human-derived radiological imaging products. |
Regulation Medical Specialty |
Radiology |
Review Panel |
Radiology |
Product Code | SAO |
Premarket Review |
Office of Radiological Health
(OHT8)
Radiological Imaging Devices and Electronic Products
(DHT8B)
|
Submission Type |
510(k)
|
Regulation Number |
892.1171
|
Device Class |
2
|
Total Product Life Cycle (TPLC) |
TPLC Product Code Report
|
GMP Exempt? |
No
|
Summary Malfunction Reporting |
Ineligible |
Implanted Device? |
No
|
Life-Sustain/Support Device? |
No
|
Third Party Review |
Not Third Party Eligible |