• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Premarket Approval (PMA)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceLIPOSORBER LA-15 SYSTEM
Generic NameLipoprotein, low density, removal
ApplicantKANEKA PHARMA AMERICA CORP.
546 FIFTH AVE., 21ST FLOOR
NEW YORK, NY 10036
PMA NumberP910018
Supplement NumberS005
Date Received06/20/2000
Decision Date12/06/2000
Product Code MMY 
Advisory Committee Gastroenterology/Urology
Supplement TypeNormal 180 Day Track
Supplement Reason Change Design/Components/Specifications/Material
Expedited Review Granted? No
Combination ProductNo
Approval Order Statement  
APPROVAL FOR 1) REPLACING THE STERILIZATION DATE WITH THE ACTUAL EXPIRATION DATE, EXPRESSED AS MONTH AND YEAR, FOR THE LIPOSORBER(R) LA-15 SYSTEM COMPONENTS: LIPOSORBER(R) LA-15 COLUMN, SULFLUX(R) FS-05 PLASMA SEPARATOR, AND THE TUBING SYTEM FOR PLASMAPHERESIS LT-MA2; 2) REPLACING THE SLIP-TYPE MALE TUBING CONNECTORS WITH LUER-LOCK TYPE CONNECTORS ON THE SEPARATOR END-CAPS OF TE SULFLUX FS-05 PLASMA SEPARATOR; 3) MAKING THE FOLLOWING CHANGES TO THE TUBING SYSTEM FOR THE PLASMAPHERESIS LT-MA2: A) PROVIDING STERILE TRANSDUCER PROTECTIVE FILTERS (FOUND SUBSTANTIALLY EQUIVALENT UNDER 510(K) K983076) IN PACKAGES OF TEN; B) REDUCING THE NUMBER OF CONNECTION TUBES FROM TEN TO EIGHT, ADDING PINCH CLIPS AND REDUCING THE SHARPNESS OF THE PLASTIC SPIKE ON EACH END OF THE CONNECTION TUBE; C) REPLACING THE STAINLESS STEEL NEEDLES, USED TO CONNECT EACH INFUSION LINE TO A BAG OR BOTTLE OF SALINE OR REGENERATION FLUID, WITH PLASTIC SPIKES; D) ADDING A LUER-LOCK "MALE" AND "FEMALE" CONNECTOR TO THE END OF THE WASTE LINES; E) INCREASING THE LENGTH OF THE TUBING SECTION BETWEEN THE BLOOD PUMP AND CHANBER LEVEL 1 FROM 80 MM TO 90 MM AND THE LENGTH OF THE TUBING SECTION BETWEEN THE WASTE DRIP CHAMBER AND A "T" DCONNECTOR ON THE WASTE LINE FROM 35 M TO 50 MM; AND F) ADDING AN INFUSION LINE.
-
-