| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | SVS APEX PLUS EXCIMER LASER WORKSTATION AND EMPHASIS DISC |
| Generic Name | Excimer laser system |
| Applicant | Alcon Laboratories, Inc. 2501 Discovery Dr., Suite 500 Orlando, FL 32826-3714 |
| PMA Number | P930034 |
| Supplement Number | S009 |
| Date Received | 04/02/1997 |
| Decision Date | 03/11/1998 |
| Withdrawal Date | 05/18/2010 |
| Product Code |
LZS |
| Docket Number | 98M-0329 |
| Notice Date | 05/27/1998 |
| Advisory Committee |
Ophthalmic |
| Supplement Type | Panel Track |
| Supplement Reason | Labeling Change - Indications/instructions/shelf life/tradename |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Approval for the SVS Apex Plus Excimer Laser Workstation and emphasis(R) disc. These devices are indicated to perform myopic astigmatic photorefractive keratectomy (Toric PRK): 1) for the reduction or elimination of mild to moderate myopia (-1.00 to < -6.00 D) and concomitant reduction or elimination of mild to moderate astigmatism (-1.00 to < -4.00D), in which the combined attempted correction must be < -6.00D spherical equivalent at the spectacle plane; 2) in patients with documentation of a stable manifest refraction(+- 1D) over the past year; and, 3) in patients who are 21 years of age or older. |
|
|