General |
Study Status |
Completed |
Application Number / Requirement Number |
P130021 S025/ PAS001 |
Date Original Protocol Accepted |
10/26/2016
|
Date Current Protocol Accepted |
 
|
Study Name |
Continued f/u of premarket cohort
|
Device Name |
Medtronic CoreValve Evolut R XL system
|
Clinical Trial Number(s) |
NCT02207569
NCT02746809
|
General Study Protocol Parameters |
Study Design |
Prospective Cohort Study
|
Data Source |
Sponsor Registry
|
Comparison Group |
No Control
|
Analysis Type |
Descriptive
|
Study Population |
Adult: >21
|
Interim or Final Data Summary |
Actual Number of Patients Enrolled |
72 enrolled; 60 received an attempted implant
|
Actual Number of Sites Enrolled |
10
|
Patient Follow-up Rate |
The follow-up rate through 5 years was approximately 75%. Compliance with visit assessments (e.g., NYHA Classification, KCCQ) ranged from 69% to 72% of subjects who completed the 5Y follow-up visit, and compliance with echocardiographic imaging assessments were approximately 63% of subjects who attended the final 5-year visit.
|
Final Safety Findings |
The Kaplan-Meier rates of CEC-adjudicated safety outcomes at 5 years are summarized as follows: - All-cause mortality: 45.9% (29.8% cardiovascular mortality) - All stroke: 14.6% (6.3% disabling stroke) - Vascular Complications: 3.3% - Bleeding: 22.8% (13.2% life-threatening or disabling) - Acute kidney injury: 0.0% - Myocardial infarction: 13.9% - New permanent pacemaker: 35.7% - Valve thrombosis: 2.1% - Valve endocarditis: 2.8% - Valve-related dysfunction requiring repeat procedure: 2.8% - Coronary artery obstruction: 0.0%
|
Final Effect Findings |
Key effectiveness outcomes are summarized as follows, in subjects who were evaluated for each outcome measure at the 5-year visit (unless otherwise noted): • NYHA Classification: Improvement in NYHA classification was maintained through 5 years. The mean change compared to baseline at 5 years was -1.1 +/- 0.8 (n=23), and 91% of the remaining 23 subjects were Class I/II. • KCCQ: Clinically significant improvement was observed in both overall and clinical summary scores through five years. The change in KCCQ overall summary score at 5 years compared to baseline was 17.0 +/- 22.9 (n=22) • Valve Performance Outcomes: o Mean effective orifice area (EOA) increased from 0.8 +/- 0.2 cm2 at baseline to 2.6 +/- 0.7 cm2 at 30 days. At 5 years, the EOA was 2.6 +/- 0.7 cm2 (n=20) o Overall, mean gradient decreased from 41.5 +/- 10.3 mmHg at baseline to 6.3 +/- 3.2 mmHg at 30 days. At 5 years, the overall mean gradient was 8.3 +/- 5.6 mmHg (n=20). o At 5 years, 100% of subjects with available echocardiographic data (n=20) had mild or less total aortic regurgitation (AR). o At 5 years, 100% of subjects with available echocardiographic data (n=20) had mild or less paravalvular AR.
|
Study Strengths & Weaknesses |
This continued follow-up study provides longer-term data (through 5 years) on the safety and effectiveness of the CoreValve Evolut 34R valve in patients with severe symptomatic aortic stenosis at high or greater surgical risk. Generally, outcomes suggest that safety and effectiveness is maintained through 5 years post procedure, with meaningful improvement in quality of life outcomes and hemodynamic performance. The lack of completion of some assessments of interest including NYHA Classification, KCCQ, and echocardiographic imaging at the 5-year study visit presents challenges to the interpretability of longer-term effectiveness outcomes. The missing data is in part related to challenges with in-person visits during the COVID-19 pandemic. Additionally, it should be noted that some subjects withdrew or were lost to follow-up in this study. Incomplete follow-up may also lead to biased longer-term results.
|
Recommendations for Labeling Changes |
Complete longer-term follow-up data warrants inclusion into the labeling for the device and should be informative to clinicians and patients regarding the longer-term benefit risk profile for the device.
|