| Class 2 Device Recall Ingenuity CT XRay System |  |
Date Initiated by Firm | September 06, 2011 |
Date Posted | December 02, 2011 |
Recall Status1 |
Terminated 3 on March 15, 2013 |
Recall Number | Z-0361-2012 |
Recall Event ID |
59886 |
510(K)Number | K033326 |
Product Classification |
System, X-ray, Tomography, computed - Product Code JAK
|
Product | Ingenuity CT computed Tomography System, Model #728326, 510(k) #K 033326, Serial Numbers: 300010 and 300018.
The Ingenuity CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at difference angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories. |
Code Information |
Model #728326, 510(k) #KK033326, Serial Numbers: 300010 and 300018. |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Road Cleveland OH 44143-2131
|
For Additional Information Contact | 440-483-7000 |
Manufacturer Reason for Recall | Software errors. A number of non-conformances were logged during in-house testing and external validation testing of Brilliance 64 and Ingenuity CT software version 4.0 Beta 2. The non-conformances are due to software errors identified in these device systems. |
FDA Determined Cause 2 | Software design |
Action | Philips Healthcare released a 'Field Change Order' (FCO #72800542) on September 15, 2011. Philips sent an URGENT - Medical Device Correction letter dated October 10, 2011, to all affected customers. The letter listed the affected device systems; described the nature of the software problem; and any possible hazard inherent to the situation. The letter was mailed by Philips to their customers on October 27, 2011. The letter provided information telling the customer how to identify the affected products and what action to take. In this instance, the customer was asked to review the Customer Release Notes which were provided to each customer location once the software update was installed. The letter notified the customer that a Philips representative would be installing a software upgrade in order to address the referenced software issues free-of-charge, through a Field Change Order. The letter closes by providing information which all customers in Canada and North America can use to contact the firm through a toll-free telephone number 1-800-722-9377, option 5; enter side ID or follow the prompts.
For questions regarding this recall call 440-483-4918. |
Quantity in Commerce | 2 units |
Distribution | Worldwide Distribution - USA including Ohio and the countries of Australia, China, Belgium, Israel, and Sweden |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|