|
Class 2 Device Recall Clinac, Trilogy, Trilogy Tx and Novalis linear accelerators |
 |
Date Initiated by Firm |
September 09, 2011 |
Date Posted |
November 09, 2011 |
Recall Status1 |
Terminated 3 on July 24, 2012 |
Recall Number |
Z-0181-2012 |
Recall Event ID |
60022 |
510(K)Number |
K033343 K100890
|
Product Classification |
Accelerator, linear, medical - Product Code IYE
|
Product |
Varian Clinac, Trilogy, Trilogy Tx and Novalis linear accelerators.
Manufactured by Varian Medical System, Palo Alto, CA.
Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
|
Code Information |
Model numbers: H14, H18, H27 and H29. Code numbers: H140001 H140061 H140121 H140225 H140743 H140801 H140859 H140919 H180082 H180209 H180310 H180463 H180536 H140002 H140062 H140122 H140226 H140744 H140802 H140860 H140920 H180083 H180210 H180311 H180466 H180538 H140003 H140063 H140123 H140227 H140745 H140803 H140861 H140921 H180087 H180214 H180312 H180467 H180539 H140004 H140064 H140124 H140228 H140746 H140804 H140862 H140923 H180088 H180215 H180314 H180468 H180540 H140005 H140065 H140125 H140229 H140747 H140805 H140863 H140924 H180093 H180217 H180318 H180469 H180541 H140006 H140066 H140126 H140230 H140748 H140806 H140864 H140925 H180098 H180218 H180319 H180470 H180542 H140007 H140067 H140127 H140231 H140749 H140807 H140865 H140926 H180099 H180220 H180321 H180471 H180543 H140008 H140068 H140128 H140232 H140750 H140808 H140866 H140927 H180102 H180221 H180322 H180472 H180544 H140009 H140069 H140129 H140233 H140751 H140809 H140867 H140929 H180103 H180222 H180323 H180473 H180545 H140010 H140070 H140130 H140234 H140752 H140810 H140868 H140930 H180104 H180225 H180324 H180474 H180548 H140011 H140071 H140131 H140235 H140753 H140811 H140869 H140931 H180107 H180226 H180325 H180475 H180549 H140012 H140072 H140132 H140236 H140754 H140812 H140870 H140932 H180108 H180228 H180326 H180476 H180550 H140014 H140073 H140133 H140237 H140755 H140813 H140871 H140933 H180115 H180229 H180327 H180477 H180551 H140015 H140074 H140134 H140238 H140756 H140814 H140872 H140934 H180116 H180233 H180328 H180478 H180552 H140016 H140075 H140135 H140239 H140757 H140815 H140873 H140935 H180117 H180235 H180329 H180479 H180553 H140017 H140076 H140136 H140240 H140758 H140816 H140874 H140936 H180119 H180236 H180402 H180480 H180554 H140018 H140077 H140137 H140700 H140759 H140817 H140875 H140937 H180122 H180237 H180404 H180482 H180555 H140019 H140078 H140138 H140701 H140760 H140818 H140876 H140938 H180124 H180238 H180405 H180484 H180556 H140021 H140079 H140139 H140702 H140761 H140819 H140877 H140939 H180128 H180240 H180406 H180485 H180557 H140022 H140080 H140140 H140703 H140762 H140820 H140878 H140940 H180129 H180242 H180407 H180486 H180559 H140023 H140082 H140141 H140704 H140763 H140821 H140879 H140941 H180131 H180243 H180409 H180487 H180560 H140024 H140083 H140142 H140705 H140764 H140822 H140881 H140943 H180132 H180245 H180411 H180489 H180561 H140025 H140084 H140143 H140706 H140765 H140823 H140882 H140944 H180133 H180249 H180412 H180490 H180562 H140026 H140085 H140144 H140707 H140766 H140824 H140883 H140945 H180135 H180250 H180414 H180491 H180563 H140027 H140086 H140145 H140708 H140767 H140825 H140884 H140947 H180143 H180251 H180417 H180492 H180564 H140028 H140087 H140146 H140709 H140768 H140826 H140885 H140948 H180144 H180254 H180419 H180493 H180565 H140029 H140088 H140147 H140710 H140769 H140827 H140886 H140950 H180147 H180255 H180420 H180494 H180566 H140030 H140089 H140148 H140711 H140770 H140828 H140887 H140951 H180148 H180256 H180422 H180495 H180567 H140031 H140090 H140149 H140712 H140771 H140829 H140888 H140952 H180151 H180257 H180424 H180496 H180568 H140032 H140091 H140150 H140713 H140772 H140830 H140889 H140953 H180154 H180258 H180425 H180498 H180569 H140033 H140092 H140151 H140714 H140773 H140831 H140890 H180003 H180157 H180260 H180426 H180499 H180570 H140034 H140093 H140152 H140715 H140774 H140832 H140891 H180004 H180159 H180262 H180427 H180501 H180571 H140035 H140094 H140153 H140716 H140775 H140833 H140892 H180007 H180161 H180264 H180432 H180505 H180572 H140036 H140095 H140154 H140717 H140776 H140834 H140893 H180008 H180164 H180267 H180433 H180507 H180573 H140037 H140096 H140201 H140718 H140777 H140835 H140895 H180016 H180166 H180268 H180434 H180508 H180574 H140038 H140097 H140202 H140719 H140778 H140836 H140896 H180017 H180167 H180271 H180436 H180509 H180575 H140039 H140098 H140203 H140720 H140779 H140837 H140897 H180019 H180168 H180272 H180437 H180510 H180576 H140040 H140099 H140204 H140721 H140780 H140838 H140898 H180025 H180170 H180273 H180440 H180511 H180577 H140041 H140100 H140205 H140722 H140781 H140839 H140899 H180026 H180171 H180275 H180441 H180513 H180578 H140042 H140101 H140206 H140723 H140782 H140840 H140900 H180028 H180173 H180283 H180442 H180514 H180579 H140043 H140102 H140207 H140724 H140783 H140841 H140901 H180040 H180175 H180284 H180443 H180515 H180580 H140044 H140103 H140208 H140725 H140784 H140842 H140902 H180043 H180178 H180285 H180444 H180517 H180581 H140045 H140105 H140209 H140726 H140785 H140843 H140903 H180044 H180179 H180287 H180445 H180518 H180582 H140046 H140106 H140210 H140727 H140786 H140844 H140904 H180045 H180181 H180289 H180446 H180519 H180583 H140047 H140107 H140211 H140728 H140787 H140845 H140905 H180048 H180182 H180290 H180447 H180520 H180584 H140048 H140108 H140212 H140729 H140788 H140846 H140906 H180049 H180185 H180291 H180449 H180521 H180585 H140049 H140109 H140213 H140730 H140789 H140847 H140907 H180050 H180186 H180292 H180450 H180522 H180586 H140050 H140110 H140214 H140731 H140790 H140848 H140908 H180052 H180188 H180293 H180451 H180524 H180587 H140051 H140111 H140215 H140732 H140791 H140849 H140909 H180053 H180192 H180294 H180452 H180525 H180588 H140052 H140112 H140216 H140733 H140792 H140850 H140910 H180056 H180193 H180295 H180453 H180527 H180589 H140053 H140113 H140217 H140734 H140793 H140851 H140911 H180058 H180195 H180296 H180454 H180528 H180590 H140054 H140114 H140218 H140735 H140794 H140852 H140912 H180059 H180197 H180298 H180455 H180529 H180591 H140055 H140115 H140219 H140736 H140795 H140853 H140913 H180065 H180199 H180299 H180456 H180530 H180592 H140056 H140116 H140220 H140737 H140796 H140854 H140914 H180066 H180201 H180303 H180457 H180531 H180593 H140057 H140117 H140221 H140738 H140797 H140855 H140915 H180073 H180203 H180304 H180459 H180532 H180594 H140058 H140118 H140222 H140739 H140798 H140856 H140916 H180075 H180204 H180305 H180460 H180533 H180595 H140059 H140119 H140223 H140740 H140799 H140857 H140917 H180079 H180206 H180307 H180461 H180534 H180596 H140060 H140120 H140224 H140742 H140800 H140858 H140918 H180080 H180208 H180309 H180462 H180535 H180597 H180598 H180661 H180725 H180788 H180852 H181012 H181073 H181135 H181199 H270118 H270353 H270482 H270645 H180599 H180662 H180726 H180789 H180853 H181013 H181074 H181136 H181200 H270126 H270355 H270483 H270646 H180600 H180663 H180727 H180791 H180854 H181014 H181075 H181137 H181201 H270128 H270356 H270486 H270647 H180601 H180664 H180729 H180792 H180855 H181015 H181076 H181138 H181202 H270134 H270360 H270487 H270648 H180602 H180665 H180730 H180793 H180856 H181016 H181077 H181139 H181203 H270136 H270364 H270489 H270649 H180603 H180667 H180731 H180794 H180857 H181017 H181078 H181140 H181204 H270138 H270365 H270490 H270651 H180604 H180668 H180733 H180796 H180858 H181018 H181079 H181141 H181205 H270144 H270370 H270491 H270652 H180605 H180669 H180734 H180798 H180859 H181019 H181080 H181142 H181206 H270201 H270371 H270495 H270653 H180606 H180670 H180735 H180799 H180860 H181020 H181081 H181143 H181207 H270209 H270373 H270498 H270654 H180607 H180671 H180736 H180800 H180861 H181021 H181082 H181145 H181208 H270211 H270378 H270499 H270656 H180608 H180672 H180737 H180801 H180862 H181022 H181083 H181146 H181209 H270215 H270383 H270500 H270660 H180610 H180673 H180738 H180802 H180863 H181023 H181084 H181147 H181210 H270216 H270384 H270501 H270661 H180611 H180674 H180739 H180803 H180864 H181024 H181085 H181148 H181211 H270219 H270385 H270502 H270662 H180612 H180675 H180740 H180804 H180865 H181025 H181086 H181149 H181212 H270227 H270386 H270503 H270664 H180614 H180676 H180741 H180805 H180866 H181026 H181087 H181150 H181214 H270228 H270391 H270504 H270665 H180615 H180678 H180742 H180806 H180867 H181027 H181088 H181151 H181215 H270229 H270392 H270505 H270667 H180616 H180679 H180743 H180807 H180868 H181028 H181089 H181152 H181220 H270230 H270395 H270506 H270669 H180617 H180680 H180744 H180808 H180869 H181029 H181090 H181153 H181223 H270232 H270398 H270508 H270670 H180618 H180681 H180745 H180809 H180870 H181030 H181091 H181154 H182000 H270235 H270399 H270511 H270671 H180619 H180682 H180746 H180810 H180871 H181031 H181092 H181155 H182001 H270240 H270400 H270512 H270675 H180620 H180683 H180747 H180811 H180872 H181032 H181093 H181156 H182002 H270244 H270401 H270513 H270680 H180621 H180684 H180748 H180812 H180873 H181033 H181094 H181157 H182003 H270245 H270404 H270516 H270681 H180622 H180685 H180749 H180813 H180874 H181034 H181095 H181158 H182005 H270251 H270409 H270518 H270682 H180623 H180686 H180750 H180814 H180875 H181035 H181096 H181159 H182006 H270252 H270410 H270520 H270684 H180624 H180687 H180751 H180815 H180876 H181036 H181097 H181160 H182007 H270254 H270412 H270522 H270685 H180625 H180688 H180752 H180816 H180877 H181037 H181098 H181161 H182008 H270255 H270416 H270525 H270686 H180626 H180689 H180753 H180817 H180878 H181038 H181099 H181162 H182009 H270256 H270418 H270526 H270690 H180627 H180690 H180754 H180818 H180879 H181039 H181100 H181163 H182010 H270260 H270419 H270529 H270692 H180628 H180691 H180755 H180819 H180880 H181040 H181101 H181164 H182012 H270261 H270421 H270532 H270693 H180629 H180692 H180756 H180820 H180881 H181041 H181102 H181165 H182013 H270264 H270425 H270534 H270697 H180630 H180694 H180757 H180821 H180882 H181042 H181103 H181166 H182016 H270267 H270426 H270535 H270698 H180631 H180695 H180758 H180822 H180883 H181043 H181104 H181167 H182017 H270271 H270427 H270539 H270699 H180632 H180696 H180759 H180823 H180884 H181044 H181105 H181168 H182018 H270273 H270428 H270540 H270700 H180633 H180697 H180760 H180824 H180885 H181045 H181106 H181170 H182019 H270277 H270430 H270542 H270705 H180634 H180698 H180761 H180825 H180886 H181046 H181107 H181171 H182020 H270280 H270431 H270544 H270708 H180635 H180699 H180762 H180826 H180887 H181047 H181108 H181172 H182021 H270281 H270435 H270547 H270709 H180636 H180700 H180763 H180827 H180888 H181048 H181109 H181173 H182022 H270282 H270436 H270550 H270710 H180637 H180701 H180764 H180828 H180890 H181049 H181110 H181174 H182023 H270285 H270437 H270601 H270711 H180638 H180702 H180765 H180829 H180891 H181050 H181111 H181175 H270004 H270290 H270438 H270602 H270712 H180639 H180703 H180766 H180830 H180892 H181051 H181112 H181176 H270006 H270293 H270439 H270605 H270714 H180640 H180704 H180767 H180831 H180894 H181052 H181113 H181177 H270009 H270297 H270443 H270606 H270716 H180641 H180705 H180768 H180832 H180895 H181053 H181114 H181178 H270010 H270302 H270445 H270607 H270718 H180642 H180706 H180769 H180833 H180896 H181055 H181115 H181179 H270021 H270303 H270446 H270613 H270719 H180643 H180707 H180770 H180834 H180897 H181056 H181116 H181180 H270026 H270307 H270447 H270615 H270725 H180644 H180708 H180771 H180835 H180898 H181057 H181117 H181181 H270027 H270310 H270448 H270622 H270726 H180645 H180709 H180772 H180836 H180899 H181058 H181118 H181182 H270029 H270314 H270451 H270623 H270727 H180646 H180710 H180773 H180837 H180901 H181059 H181119 H181183 H270030 H270321 H270452 H270624 H270728 H180647 H180711 H180774 H180838 H180907 H181060 H181120 H181184 H270038 H270323 H270454 H270625 H270732 H180649 H180713 H180775 H180839 H181000 H181061 H181121 H181185 H270045 H270324 H270455 H270626 H270733 H180650 H180714 H180776 H180840 H181001 H181062 H181123 H181187 H270047 H270329 H270458 H270627 H270735 H180651 H180715 H180777 H180841 H181002 H181063 H181124 H181188 H270073 H270332 H270461 H270631 H270736 H180652 H180716 H180778 H180842 H181003 H181064 H181125 H181189 H270079 H270335 H270462 H270632 H270737 H180653 H180717 H180779 H180843 H181004 H181065 H181126 H181191 H270088 H270337 H270464 H270633 H270739 H180654 H180718 H180780 H180844 H181005 H181066 H181127 H181192 H270094 H270338 H270466 H270634 H270740 H180655 H180719 H180781 H180845 H181006 H181067 H181128 H181193 H270096 H270341 H270470 H270635 H270746 H180656 H180720 H180782 H180847 H181007 H181068 H181129 H181194 H270101 H270343 H270473 H270637 H270749 H180657 H180721 H180783 H180848 H181008 H181069 H181130 H181195 H270106 H270344 H270474 H270639 H270751 H180658 H180722 H180784 H180849 H181009 H181070 H181132 H181196 H270107 H270345 H270476 H270641 H270752 H180659 H180723 H180785 H180850 H181010 H181071 H181133 H181197 H270110 H270351 H270477 H270643 H270755 H180660 H180724 H180786 H180851 H181011 H181072 H181134 H181198 H270111 H270352 H270479 H270644 H270757 H270760 H270865 H271051 H271141 H271219 H271293 H271362 H271426 H271499 H271562 H271627 H271692 H271756 H270761 H270866 H271052 H271142 H271222 H271294 H271363 H271428 H271500 H271563 H271628 H271693 H271757 H270762 H270867 H271055 H271143 H271223 H271295 H271364 H271429 H271501 H271564 H271629 H271695 H271758 H270763 H270868 H271056 H271144 H271224 H271296 H271365 H271430 H271502 H271565 H271630 H271696 H271759 H270766 H270869 H271057 H271145 H271225 H271297 H271366 H271431 H271503 H271566 H271631 H271697 H271760 H270767 H270870 H271059 H271149 H271226 H271298 H271367 H271432 H271504 H271567 H271632 H271698 H271761 H270769 H270871 H271061 H271151 H271227 H271300 H271368 H271433 H271505 H271568 H271633 H271699 H271762 H270777 H270872 H271062 H271152 H271228 H271301 H271369 H271434 H271506 H271570 H271634 H271700 H271763 H270778 H270873 H271063 H271153 H271229 H271302 H271370 H271435 H271507 H271571 H271635 H271701 H271764 H270781 H270875 H271065 H271154 H271230 H271303 H271371 H271436 H271508 H271572 H271636 H271702 H271765 H270782 H270877 H271066 H271155 H271231 H271304 H271372 H271437 H271509 H271573 H271637 H271703 H271766 H270783 H270878 H271067 H271157 H271233 H271305 H271373 H271438 H271510 H271574 H271638 H271704 H271767 H270785 H270879 H271068 H271158 H271234 H271306 H271374 H271439 H271511 H271575 H271639 H271705 H271768 H270786 H270880 H271071 H271159 H271235 H271307 H271375 H271440 H271512 H271576 H271640 H271706 H271769 H270787 H270881 H271072 H271160 H271236 H271308 H271376 H271441 H271513 H271577 H271641 H271707 H271770 H270788 H270882 H271073 H271161 H271238 H271309 H271377 H271442 H271514 H271578 H271642 H271708 H271771 H270789 H270883 H271075 H271162 H271239 H271311 H271378 H271443 H271515 H271580 H271643 H271709 H271772 H270791 H270884 H271076 H271165 H271241 H271312 H271379 H271444 H271516 H271581 H271645 H271710 H271773 H270793 H270885 H271079 H271166 H271242 H271314 H271380 H271445 H271517 H271582 H271646 H271711 H271774 H270794 H270886 H271080 H271167 H271243 H271315 H271381 H271447 H271518 H271583 H271647 H271712 H271775 H270796 H270888 H271082 H271168 H271245 H271316 H271382 H271448 H271519 H271586 H271648 H271713 H271776 H270799 H270889 H271083 H271172 H271247 H271317 H271383 H271449 H271520 H271587 H271649 H271714 H271777 H270800 H270890 H271085 H271174 H271248 H271318 H271384 H271450 H271521 H271588 H271650 H271715 H271778 H270801 H270891 H271086 H271176 H271249 H271319 H271385 H271451 H271522 H271589 H271651 H271716 H271779 H270803 H270892 H271088 H271177 H271250 H271320 H271386 H271453 H271523 H271590 H271652 H271717 H271780 H270804 H270894 H271091 H271178 H271251 H271321 H271387 H271454 H271525 H271591 H271653 H271718 H271781 H270805 H270895 H271092 H271179 H271253 H271322 H271388 H271456 H271526 H271592 H271654 H271719 H271782 H270806 H270896 H271093 H271182 H271256 H271323 H271389 H271457 H271527 H271593 H271655 H271720 H271783 H270812 H270897 H271094 H271183 H271257 H271324 H271390 H271458 H271528 H271594 H271656 H271722 H271784 H270813 H270898 H271095 H271184 H271258 H271325 H271392 H271459 H271529 H271595 H271657 H271723 H271785 H270817 H270899 H271096 H271185 H271259 H271327 H271393 H271460 H271530 H271596 H271658 H271724 H271786 H270820 H271000 H271097 H271186 H271260 H271328 H271394 H271461 H271531 H271597 H271659 H271725 H271787 H270821 H271002 H271098 H271188 H271263 H271329 H271395 H271462 H271532 H271598 H271660 H271726 H271788 H270822 H271004 H271099 H271189 H271264 H271330 H271397 H271464 H271533 H271599 H271662 H271727 H271789 H270823 H271005 H271100 H271190 H271265 H271331 H271398 H271465 H271534 H271600 H271663 H271728 H271790 H270825 H271009 H271101 H271191 H271266 H271332 H271399 H271466 H271535 H271601 H271664 H271729 H271791 H270826 H271013 H271102 H271192 H271267 H271333 H271401 H271467 H271536 H271602 H271666 H271731 H271792 H270827 H271015 H271103 H271193 H271268 H271334 H271402 H271468 H271537 H271603 H271667 H271732 H271793 H270828 H271016 H271104 H271194 H271269 H271335 H271403 H271470 H271538 H271604 H271669 H271733 H271794 H270829 H271017 H271105 H271195 H271270 H271336 H271404 H271472 H271539 H271605 H271670 H271734 H271795 H270833 H271019 H271107 H271196 H271272 H271337 H271406 H271473 H271540 H271606 H271671 H271735 H271796 H270835 H271020 H271108 H271197 H271273 H271338 H271407 H271474 H271541 H271607 H271672 H271736 H271797 H270836 H271021 H271109 H271198 H271274 H271339 H271408 H271475 H271542 H271608 H271673 H271737 H271798 H270837 H271023 H271110 H271199 H271275 H271340 H271409 H271476 H271543 H271609 H271674 H271738 H271799 H270838 H271024 H271111 H271200 H271276 H271342 H271410 H271477 H271544 H271610 H271675 H271739 H271800 H270840 H271025 H271112 H271201 H271277 H271343 H271411 H271478 H271545 H271611 H271676 H271740 H271801 H270841 H271026 H271113 H271202 H271278 H271344 H271412 H271479 H271546 H271612 H271677 H271741 H271802 H270842 H271027 H271114 H271204 H271280 H271345 H271413 H271481 H271547 H271613 H271678 H271742 H271803 H270844 H271029 H271115 H271205 H271281 H271347 H271414 H271483 H271548 H271614 H271679 H271743 H271804 H270845 H271031 H271116 H271206 H271282 H271349 H271415 H271484 H271549 H271615 H271680 H271744 H271805 H270846 H271032 H271117 H271207 H271283 H271350 H271416 H271485 H271550 H271616 H271681 H271746 H271806 H270852 H271034 H271120 H271208 H271284 H271351 H271417 H271487 H271551 H271617 H271682 H271747 H271807 H270854 H271035 H271121 H271209 H271285 H271353 H271418 H271488 H271552 H271618 H271683 H271748 H271808 H270855 H271037 H271125 H271210 H271286 H271355 H271419 H271490 H271553 H271619 H271684 H271749 H271809 H270857 H271038 H271127 H271211 H271287 H271356 H271420 H271492 H271554 H271620 H271685 H271750 H271810 H270858 H271039 H271131 H271212 H271288 H271357 H271421 H271493 H271555 H271621 H271686 H271751 H271811 H270859 H271041 H271132 H271214 H271289 H271358 H271422 H271494 H271557 H271622 H271687 H271752 H271812 H270861 H271042 H271135 H271216 H271290 H271359 H271423 H271495 H271558 H271624 H271688 H271753 H271813 H270863 H271043 H271136 H271217 H271291 H271360 H271424 H271497 H271559 H271625 H271689 H271754 H271814 H270864 H271044 H271137 H271218 H271292 H271361 H271425 H271498 H271561 H271626 H271691 H271755 H271815 H271816 H271876 H271937 H271998 H272062 H272124 H272184 H272246 H272308 H272368 H272428 H272492 H272855 H271817 H271877 H271938 H271999 H272063 H272125 H272185 H272247 H272309 H272369 H272429 H272493 H272856 H271818 H271878 H271939 H272000 H272064 H272126 H272186 H272248 H272310 H272370 H272430 H272494 H272857 H271819 H271879 H271940 H272001 H272065 H272127 H272187 H272249 H272311 H272371 H272431 H272495 H272858 H271820 H271880 H271941 H272002 H272066 H272128 H272188 H272250 H272312 H272372 H272432 H272496 H272859 H271821 H271881 H271942 H272003 H272067 H272129 H272189 H272251 H272313 H272373 H272433 H272497 H272860 H271822 H271882 H271943 H272004 H272068 H272130 H272190 H272253 H272314 H272374 H272434 H272500 H272861 H271823 H271883 H271944 H272005 H272069 H272131 H272191 H272254 H272315 H272375 H272435 H272800 H272862 H271824 H271884 H271945 H272006 H272070 H272132 H272192 H272255 H272316 H272376 H272436 H272801 H272863 H271825 H271885 H271946 H272007 H272071 H272133 H272193 H272256 H272317 H272377 H272437 H272802 H272864 H271826 H271886 H271947 H272009 H272072 H272134 H272194 H272257 H272318 H272378 H272438 H272803 H272865 H271827 H271887 H271948 H272010 H272073 H272135 H272195 H272258 H272319 H272379 H272439 H272804 H272866 H271828 H271888 H271949 H272011 H272074 H272136 H272196 H272259 H272320 H272380 H272440 H272805 H272867 H271829 H271889 H271950 H272012 H272075 H272137 H272197 H272260 H272321 H272381 H272441 H272806 H272868 H271830 H271890 H271951 H272013 H272076 H272138 H272198 H272261 H272322 H272382 H272442 H272807 H272869 H271831 H271891 H271952 H272014 H272077 H272139 H272199 H272262 H272323 H272383 H272443 H272808 H272870 H271832 H271892 H271953 H272015 H272078 H272140 H272200 H272263 H272324 H272384 H272444 H272809 H272871 H271833 H271893 H271954 H272016 H272079 H272141 H272201 H272264 H272325 H272385 H272445 H272810 H272872 H271834 H271894 H271955 H272017 H272080 H272142 H272202 H272265 H272326 H272386 H272446 H272811 H272873 H271835 H271895 H271956 H272018 H272081 H272143 H272203 H272266 H272327 H272387 H272447 H272812 H272875 H271836 H271896 H271957 H272019 H272082 H272144 H272204 H272267 H272328 H272388 H272448 H272813 H272876 H271837 H271897 H271958 H272020 H272083 H272145 H272205 H272268 H272329 H272389 H272449 H272814 H272877 H271838 H271898 H271959 H272021 H272084 H272146 H272206 H272269 H272330 H272390 H272450 H272815 H272878 H271839 H271899 H271960 H272022 H272085 H272147 H272207 H272270 H272331 H272391 H272451 H272816 H272879 H271840 H271901 H271961 H272023 H272086 H272148 H272208 H272271 H272332 H272392 H272452 H272817 H272880 H271841 H271902 H271962 H272024 H272087 H272149 H272209 H272272 H272333 H272393 H272453 H272818 H272881 H271842 H271903 H271963 H272025 H272088 H272150 H272210 H272273 H272334 H272394 H272454 H272819 H272882 H271843 H271904 H271964 H272026 H272089 H272151 H272211 H272274 H272335 H272395 H272455 H272820 H272883 H271844 H271905 H271965 H272027 H272090 H272152 H272212 H272276 H272336 H272396 H272456 H272821 H272884 H271845 H271906 H271966 H272028 H272091 H272153 H272213 H272277 H272337 H272397 H272457 H272822 H272885 H271846 H271907 H271967 H272029 H272092 H272154 H272215 H272278 H272338 H272398 H272458 H272823 H272886 H271847 H271908 H271968 H272030 H272093 H272155 H272216 H272279 H272339 H272399 H272459 H272824 H272887 H271848 H271909 H271969 H272031 H272094 H272156 H272217 H272280 H272340 H272400 H272460 H272825 H272888 H271849 H271910 H271970 H272032 H272095 H272157 H272218 H272281 H272341 H272401 H272461 H272826 H272889 H271850 H271911 H271971 H272033 H272096 H272158 H272219 H272282 H272342 H272402 H272462 H272827 H272890 H271851 H271912 H271972 H272034 H272097 H272159 H272220 H272283 H272343 H272403 H272463 H272828 H272891 H271852 H271913 H271973 H272035 H272098 H272160 H272221 H272284 H272344 H272404 H272464 H272829 H272892 H271853 H271914 H271974 H272036 H272099 H272161 H272222 H272285 H272345 H272405 H272465 H272830 H272893 H271854 H271915 H271975 H272038 H272100 H272162 H272223 H272286 H272346 H272406 H272466 H272831 H272894 H271855 H271916 H271976 H272039 H272101 H272163 H272224 H272287 H272347 H272407 H272467 H272833 H272895 H271856 H271917 H271977 H272040 H272102 H272164 H272226 H272288 H272348 H272408 H272468 H272834 H272896 H271857 H271918 H271978 H272041 H272103 H272165 H272227 H272289 H272349 H272409 H272469 H272836 H272897 H271858 H271919 H271979 H272043 H272104 H272166 H272228 H272290 H272350 H272410 H272470 H272837 H272898 H271859 H271920 H271980 H272045 H272106 H272167 H272229 H272291 H272351 H272411 H272471 H272838 H272899 H271860 H271921 H271981 H272046 H272107 H272168 H272230 H272292 H272352 H272412 H272472 H272839 H272900 H271861 H271922 H271982 H272047 H272108 H272169 H272231 H272293 H272353 H272413 H272473 H272840 H272901 H271862 H271923 H271983 H272048 H272109 H272170 H272232 H272294 H272354 H272414 H272475 H272841 H272902 H271863 H271924 H271984 H272049 H272110 H272171 H272233 H272295 H272355 H272415 H272476 H272842 H272903 H271864 H271925 H271985 H272050 H272111 H272172 H272234 H272296 H272356 H272416 H272477 H272843 H272904 H271865 H271926 H271986 H272051 H272112 H272173 H272235 H272297 H272357 H272417 H272478 H272844 H272905 H271866 H271927 H271987 H272052 H272113 H272174 H272236 H272298 H272358 H272418 H272479 H272845 H272906 H271867 H271928 H271988 H272053 H272114 H272175 H272237 H272299 H272359 H272419 H272480 H272846 H272907 H271868 H271929 H271989 H272054 H272115 H272176 H272238 H272300 H272360 H272420 H272482 H272847 H272908 H271869 H271930 H271990 H272055 H272116 H272177 H272239 H272301 H272361 H272421 H272483 H272848 H272909 H271870 H271931 H271991 H272056 H272117 H272178 H272240 H272302 H272362 H272422 H272484 H272849 H272910 H271871 H271932 H271992 H272057 H272118 H272179 H272241 H272303 H272363 H272423 H272485 H272850 H272911 H271872 H271933 H271993 H272058 H272119 H272180 H272242 H272304 H272364 H272424 H272488 H272851 H272912 H271873 H271934 H271994 H272059 H272121 H272181 H272243 H272305 H272365 H272425 H272489 H272852 H272913 H271874 H271935 H271996 H272060 H272122 H272182 H272244 H272306 H272366 H272426 H272490 H272853 H272914 H271875 H271936 H271997 H272061 H272123 H272183 H272245 H272307 H272367 H272427 H272491 H272854 H272915 H272916 H272977 H273040 H273104 H273171 H290125 H290215 H290282 H290346 H290409 H290545 H291045 H291106 H272917 H272978 H273041 H273105 H273172 H290128 H290216 H290283 H290347 H290410 H290546 H291046 H291107 H272918 H272979 H273042 H273106 H273173 H290130 H290217 H290284 H290348 H290411 H290547 H291047 H291108 H272919 H272980 H273043 H273107 H273174 H290132 H290219 H290285 H290349 H290412 H290548 H291048 H291109 H272920 H272981 H273044 H273108 H273175 H290135 H290220 H290286 H290350 H290413 H290549 H291049 H291111 H272921 H272982 H273045 H273109 H273176 H290137 H290221 H290287 H290351 H290414 H290550 H291050 H291112 H272922 H272983 H273046 H273110 H273177 H290138 H290222 H290288 H290352 H290415 H290551 H291051 H291113 H272923 H272984 H273047 H273111 H273178 H290139 H290223 H290289 H290353 H290416 H290552 H291052 H291114 H272924 H272985 H273048 H273112 H273179 H290141 H290224 H290290 H290354 H290417 H290553 H291053 H291115 H272925 H272986 H273049 H273113 H273180 H290143 H290225 H290291 H290355 H290422 H290554 H291054 H291116 H272926 H272988 H273050 H273114 H273181 H290144 H290226 H290292 H290356 H290423 H290555 H291055 H291117 H272927 H272989 H273051 H273115 H273182 H290145 H290227 H290293 H290357 H290424 H290556 H291056 H291118 H272928 H272990 H273052 H273116 H273183 H290146 H290229 H290295 H290358 H290425 H290557 H291057 H291119 H272929 H272991 H273053 H273117 H273184 H290151 H290230 H290296 H290359 H290426 H290558 H291058 H291120 H272930 H272992 H273054 H273118 H273185 H290152 H290231 H290297 H290360 H290427 H290559 H291059 H291121 H272931 H272993 H273055 H273119 H273186 H290154 H290232 H290299 H290362 H290500 H290560 H291060 H291122 H272932 H272994 H273056 H273120 H273187 H290155 H290233 H290301 H290363 H290501 H290561 H291061 H291123 H272933 H272995 H273057 H273121 H273189 H290156 H290234 H290302 H290364 H290502 H290600 H291062 H291124 H272934 H272996 H273058 H273122 H273190 H290157 H290235 H290304 H290365 H290503 H291001 H291063 H291125 H272935 H272997 H273059 H273123 H273191 H290158 H290236 H290305 H290366 H290504 H291003 H291064 H291126 H272936 H272998 H273060 H273124 H273193 H290159 H290237 H290306 H290367 H290505 H291004 H291065 H291127 H272937 H272999 H273061 H273125 H273194 H290160 H290239 H290307 H290368 H290506 H291005 H291066 H291128 H272938 H273000 H273062 H273126 H273195 H290161 H290240 H290308 H290369 H290507 H291006 H291067 H291129 H272939 H273001 H273063 H273127 H273196 H290162 H290241 H290309 H290370 H290508 H291007 H291068 H291130 H272940 H273002 H273065 H273128 H273197 H290164 H290242 H290310 H290371 H290509 H291008 H291069 H291131 H272941 H273003 H273066 H273130 H273198 H290165 H290243 H290311 H290372 H290510 H291009 H291070 H291132 H272942 H273004 H273067 H273131 H273199 H290166 H290244 H290312 H290373 H290511 H291010 H291071 H291133 H272943 H273005 H273068 H273132 H273200 H290168 H290246 H290313 H290374 H290512 H291011 H291072 H291134 H272944 H273006 H273069 H273133 H273201 H290171 H290247 H290314 H290375 H290513 H291012 H291073 H291135 H272946 H273007 H273070 H273134 H273202 H290173 H290248 H290315 H290376 H290514 H291013 H291074 H291136 H272947 H273008 H273071 H273135 H273203 H290175 H290249 H290316 H290377 H290515 H291014 H291075 H291137 H272948 H273010 H273072 H273136 H273204 H290176 H290250 H290317 H290378 H290516 H291015 H291076 H291138 H272949 H273011 H273074 H273137 H273206 H290180 H290253 H290318 H290379 H290517 H291016 H291077 H291139 H272950 H273012 H273075 H273138 H273208 H290181 H290254 H290319 H290380 H290518 H291017 H291078 H291140 H272951 H273013 H273077 H273140 H273209 H290182 H290255 H290320 H290381 H290519 H291018 H291079 H291141 H272952 H273014 H273078 H273141 H273210 H290183 H290256 H290321 H290382 H290520 H291019 H291080 H291142 H272953 H273016 H273079 H273142 H290001 H290185 H290257 H290322 H290383 H290521 H291020 H291081 H291143 H272954 H273017 H273080 H273143 H290003 H290186 H290258 H290323 H290384 H290522 H291021 H291082 H291144 H272955 H273018 H273081 H273144 H290021 H290187 H290259 H290324 H290385 H290523 H291022 H291083 H291145 H272956 H273019 H273082 H273145 H290026 H290189 H290260 H290325 H290386 H290524 H291023 H291084 H291146 H272957 H273020 H273083 H273147 H290027 H290191 H290261 H290326 H290387 H290525 H291024 H291085 H291147 H272958 H273021 H273084 H273148 H290030 H290192 H290262 H290327 H290388 H290526 H291025 H291086 H291148 H272959 H273022 H273085 H273149 H290031 H290193 H290263 H290328 H290389 H290527 H291026 H291087 H291149 H272960 H273023 H273086 H273150 H290032 H290194 H290264 H290329 H290390 H290528 H291027 H291088 H291150 H272961 H273024 H273087 H273151 H290033 H290195 H290265 H290330 H290392 H290529 H291028 H291089 H291151 H272962 H273025 H273088 H273152 H290034 H290196 H290266 H290331 H290393 H290530 H291029 H291090 H291152 H272963 H273026 H273089 H273153 H290103 H290197 H290267 H290332 H290394 H290531 H291030 H291091 H291153 H272964 H273027 H273090 H273155 H290104 H290198 H290268 H290333 H290395 H290532 H291031 H291092 H291154 H272965 H273028 H273091 H273156 H290105 H290201 H290269 H290334 H290396 H290533 H291032 H291093 H291155 H272966 H273029 H273092 H273157 H290108 H290202 H290271 H290335 H290397 H290534 H291034 H291094 H291156 H272967 H273030 H273093 H273158 H290109 H290203 H290272 H290336 H290398 H290535 H291035 H291095 H291157 H272968 H273031 H273094 H273159 H290110 H290205 H290273 H290337 H290399 H290536 H291036 H291096 H291158 H272969 H273032 H273095 H273160 H290111 H290206 H290274 H290338 H290400 H290537 H291037 H291097 H291159 H272970 H273033 H273097 H273162 H290113 H290208 H290275 H290339 H290401 H290538 H291038 H291098 H291160 H272971 H273034 H273098 H273163 H290114 H290209 H290276 H290340 H290402 H290539 H291039 H291099 H291161 H272972 H273035 H273099 H273164 H290116 H290210 H290277 H290341 H290403 H290540 H291040 H291101 H291162 H272973 H273036 H273100 H273165 H290118 H290211 H290278 H290342 H290404 H290541 H291041 H291102 H291163 H272974 H273037 H273101 H273166 H290121 H290212 H290279 H290343 H290405 H290542 H291042 H291103 H291164 H272975 H273038 H273102 H273167 H290122 H290213 H290280 H290344 H290406 H290543 H291043 H291104 H291165 H272976 H273039 H273103 H273170 H290123 H290214 H290281 H290345 H290408 H290544 H291044 H291105 H291166 H291167 H291227 H293310 H293370 H293432 H293492 H293554 H293616 H293677 H293739 H293800 H293861 H293921 H291168 H291228 H293311 H293371 H293433 H293493 H293555 H293617 H293678 H293740 H293801 H293862 H293922 H291169 H291229 H293312 H293372 H293434 H293494 H293556 H293618 H293679 H293741 H293802 H293863 H293923 H291170 H291230 H293313 H293373 H293435 H293495 H293557 H293619 H293681 H293742 H293803 H293864 H293924 H291171 H291231 H293314 H293374 H293436 H293496 H293558 H293620 H293682 H293743 H293804 H293865 H293925 H291172 H291232 H293315 H293375 H293437 H293497 H293559 H293621 H293683 H293744 H293805 H293866 H293926 H291173 H291233 H293316 H293376 H293438 H293498 H293560 H293622 H293684 H293745 H293806 H293867 H293927 H291174 H291234 H293317 H293377 H293439 H293499 H293561 H293623 H293685 H293746 H293807 H293868 H293928 H291175 H291235 H293318 H293378 H293440 H293500 H293562 H293624 H293686 H293747 H293808 H293869 H293929 H291176 H291236 H293319 H293379 H293441 H293501 H293563 H293625 H293687 H293748 H293809 H293870 H293930 H291177 H291237 H293320 H293380 H293442 H293502 H293564 H293626 H293688 H293749 H293810 H293871 H293931 H291178 H291238 H293321 H293381 H293443 H293503 H293565 H293627 H293689 H293750 H293811 H293872 H293932 H291179 H291239 H293322 H293382 H293444 H293504 H293566 H293628 H293690 H293751 H293812 H293873 H293933 H291180 H291240 H293323 H293383 H293445 H293505 H293567 H293629 H293691 H293752 H293813 H293874 H293934 H291181 H291241 H293324 H293384 H293446 H293506 H293568 H293630 H293692 H293753 H293814 H293875 H293935 H291182 H291242 H293325 H293385 H293447 H293507 H293569 H293631 H293693 H293754 H293815 H293876 H293936 H291183 H291243 H293326 H293386 H293448 H293508 H293570 H293632 H293694 H293755 H293816 H293877 H293938 H291184 H291244 H293327 H293387 H293449 H293509 H293571 H293633 H293695 H293756 H293817 H293878 H293940 H291185 H291245 H293328 H293388 H293450 H293510 H293572 H293634 H293696 H293758 H293818 H293879 H293941 H291186 H291246 H293329 H293389 H293451 H293511 H293573 H293635 H293697 H293759 H293819 H293880 H293942 H291187 H291247 H293330 H293390 H293452 H293512 H293574 H293636 H293698 H293760 H293820 H293881 H293943 H291188 H291248 H293331 H293391 H293453 H293513 H293575 H293637 H293699 H293761 H293821 H293882 H293944 H291189 H291249 H293332 H293393 H293454 H293514 H293576 H293638 H293700 H293762 H293822 H293883 H293945 H291190 H291250 H293333 H293394 H293455 H293515 H293577 H293639 H293701 H293763 H293823 H293884 H293946 H291191 H291251 H293334 H293395 H293456 H293516 H293578 H293640 H293702 H293764 H293824 H293885 H293947 H291192 H291252 H293335 H293396 H293457 H293517 H293579 H293641 H293703 H293765 H293825 H293886 H293948 H291193 H291253 H293336 H293397 H293458 H293518 H293581 H293642 H293704 H293766 H293826 H293887 H293949 H291194 H291254 H293337 H293398 H293459 H293519 H293583 H293643 H293705 H293767 H293827 H293888 H293950 H291195 H291255 H293338 H293399 H293460 H293520 H293584 H293644 H293706 H293768 H293828 H293889 H293951 H291196 H291256 H293339 H293401 H293461 H293521 H293585 H293645 H293707 H293769 H293829 H293890 H293952 H291197 H291257 H293340 H293402 H293462 H293522 H293586 H293646 H293708 H293770 H293830 H293891 H293953 H291198 H291258 H293341 H293403 H293463 H293523 H293587 H293648 H293709 H293771 H293831 H293892 H293954 H291199 H291259 H293342 H293404 H293464 H293524 H293588 H293649 H293710 H293772 H293832 H293893 H293955 H291200 H291260 H293343 H293405 H293465 H293525 H293589 H293650 H293712 H293773 H293833 H293894 H293956 H291201 H291261 H293344 H293406 H293466 H293527 H293590 H293651 H293713 H293774 H293834 H293895 H293957 H291202 H291262 H293345 H293407 H293467 H293528 H293591 H293652 H293714 H293775 H293835 H293896 H293958 H291203 H291263 H293346 H293408 H293468 H293529 H293592 H293653 H293715 H293776 H293836 H293897 H293959 H291204 H291264 H293347 H293409 H293469 H293530 H293593 H293654 H293716 H293777 H293837 H293898 H293960 H291205 H291265 H293348 H293410 H293470 H293531 H293594 H293655 H293717 H293778 H293838 H293899 H293961 H291206 H291266 H293349 H293411 H293471 H293532 H293595 H293656 H293718 H293779 H293839 H293900 H293962 H291207 H291267 H293350 H293412 H293472 H293533 H293596 H293657 H293719 H293780 H293840 H293901 H293963 H291208 H291268 H293351 H293413 H293473 H293534 H293597 H293658 H293720 H293781 H293841 H293902 H293964 H291209 H291269 H293352 H293414 H293474 H293535 H293598 H293659 H293721 H293782 H293842 H293903 H293965 H291210 H291270 H293353 H293415 H293475 H293536 H293599 H293660 H293722 H293783 H293843 H293904 H293966 H291211 H291271 H293354 H293416 H293476 H293537 H293600 H293661 H293723 H293784 H293844 H293905 H293967 H291212 H291272 H293355 H293417 H293477 H293538 H293601 H293662 H293724 H293785 H293845 H293906 H293968 H291213 H291273 H293356 H293418 H293478 H293539 H293602 H293663 H293725 H293786 H293846 H293907 H293969 H291214 H292832 H293357 H293419 H293479 H293540 H293603 H293664 H293726 H293787 H293847 H293908 H293970 H291215 H293009 H293358 H293420 H293480 H293541 H293604 H293665 H293727 H293788 H293848 H293909 H293971 H291216 H293207 H293359 H293421 H293481 H293542 H293605 H293666 H293728 H293789 H293849 H293910 H293972 H291217 H293300 H293360 H293422 H293482 H293543 H293606 H293667 H293729 H293790 H293850 H293911 H293974 H291218 H293301 H293361 H293423 H293483 H293544 H293607 H293668 H293730 H293791 H293851 H293912 H293975 H291219 H293302 H293362 H293424 H293484 H293545 H293608 H293669 H293731 H293792 H293852 H293913 H293976 H291220 H293303 H293363 H293425 H293485 H293546 H293609 H293670 H293732 H293793 H293854 H293914 H293977 H291221 H293304 H293364 H293426 H293486 H293548 H293610 H293671 H293733 H293794 H293855 H293915 H293978 H291222 H293305 H293365 H293427 H293487 H293549 H293611 H293672 H293734 H293795 H293856 H293916 H293979 H291223 H293306 H293366 H293428 H293488 H293550 H293612 H293673 H293735 H293796 H293857 H293917 H293980 H291224 H293307 H293367 H293429 H293489 H293551 H293613 H293674 H293736 H293797 H293858 H293918 H293981 H291225 H293308 H293368 H293430 H293490 H293552 H293614 H293675 H293737 H293798 H293859 H293919 H293982 H291226 H293309 H293369 H293431 H293491 H293553 H293615 H293676 H293738 H293799 H293860 H293920 H293983 H293984 H294045 H294107 H294169 H294229 H294291 H294352 H294413 H294477 H294537 H294604 H294666 H294730 H293985 H294046 H294108 H294170 H294230 H294292 H294353 H294414 H294478 H294538 H294605 H294667 H294731 H293986 H294047 H294109 H294171 H294231 H294293 H294354 H294415 H294479 H294539 H294606 H294668 H294732 H293987 H294048 H294110 H294172 H294232 H294294 H294355 H294416 H294480 H294540 H294607 H294669 H294733 H293988 H294049 H294111 H294173 H294233 H294295 H294356 H294417 H294481 H294541 H294608 H294670 H294734 H293989 H294050 H294112 H294174 H294234 H294296 H294357 H294418 H294482 H294542 H294610 H294671 H294735 H293990 H294051 H294113 H294175 H294235 H294297 H294358 H294419 H294483 H294543 H294611 H294672 H294736 H293991 H294052 H294114 H294176 H294236 H294298 H294359 H294420 H294484 H294544 H294612 H294673 H294737 H293992 H294053 H294116 H294177 H294237 H294299 H294360 H294421 H294485 H294545 H294613 H294674 H294738 H293993 H294054 H294117 H294178 H294238 H294300 H294361 H294422 H294486 H294546 H294614 H294675 H294739 H293994 H294055 H294118 H294179 H294239 H294301 H294362 H294423 H294487 H294547 H294615 H294676 H294740 H293995 H294056 H294119 H294180 H294240 H294302 H294363 H294424 H294488 H294548 H294616 H294677 H294741 H293996 H294057 H294120 H294181 H294241 H294303 H294364 H294425 H294489 H294549 H294617 H294678 H294742 H293997 H294058 H294121 H294182 H294242 H294304 H294365 H294426 H294490 H294550 H294618 H294679 H294743 H293998 H294059 H294122 H294183 H294243 H294305 H294366 H294427 H294491 H294551 H294619 H294680 H294744 H293999 H294060 H294123 H294184 H294244 H294306 H294367 H294428 H294492 H294552 H294620 H294681 H294745 H294000 H294061 H294124 H294185 H294245 H294307 H294368 H294429 H294493 H294553 H294621 H294682 H294746 H294001 H294062 H294125 H294186 H294246 H294308 H294369 H294430 H294494 H294554 H294622 H294683 H294747 H294002 H294063 H294126 H294187 H294247 H294309 H294370 H294431 H294495 H294555 H294623 H294684 H294748 H294003 H294064 H294127 H294188 H294248 H294310 H294371 H294432 H294496 H294556 H294624 H294685 H294749 H294004 H294065 H294128 H294189 H294249 H294311 H294372 H294433 H294497 H294557 H294625 H294686 H294750 H294005 H294066 H294129 H294190 H294250 H294312 H294373 H294434 H294498 H294564 H294626 H294687 H294751 H294006 H294067 H294130 H294191 H294252 H294313 H294374 H294435 H294499 H294565 H294627 H294688 H294752 H294007 H294068 H294131 H294192 H294253 H294314 H294375 H294436 H294500 H294567 H294628 H294689 H294753 H294008 H294069 H294132 H294193 H294254 H294315 H294376 H294437 H294501 H294568 H294629 H294690 H294754 H294009 H294070 H294133 H294194 H294255 H294316 H294377 H294438 H294502 H294569 H294630 H294691 H294755 H294010 H294071 H294134 H294195 H294256 H294317 H294378 H294439 H294503 H294570 H294631 H294692 H294756 H294011 H294072 H294135 H294196 H294257 H294318 H294379 H294440 H294504 H294571 H294632 H294693 H294757 H294012 H294073 H294136 H294197 H294258 H294319 H294380 H294442 H294505 H294572 H294633 H294694 H294758 H294013 H294074 H294137 H294198 H294259 H294320 H294381 H294443 H294506 H294573 H294634 H294695 H294759 H294014 H294075 H294139 H294199 H294260 H294321 H294382 H294444 H294507 H294574 H294635 H294696 H294760 H294015 H294076 H294140 H294200 H294261 H294322 H294383 H294445 H294508 H294575 H294636 H294697 H294761 H294016 H294077 H294141 H294201 H294262 H294323 H294384 H294446 H294509 H294576 H294637 H294698 H294762 H294017 H294078 H294142 H294202 H294263 H294324 H294385 H294447 H294510 H294577 H294638 H294700 H294763 H294018 H294079 H294143 H294203 H294264 H294325 H294386 H294448 H294511 H294578 H294639 H294701 H294764 H294020 H294080 H294144 H294204 H294265 H294326 H294387 H294449 H294512 H294579 H294640 H294702 H294765 H294021 H294081 H294145 H294205 H294266 H294327 H294388 H294450 H294513 H294580 H294641 H294704 H294766 H294022 H294083 H294146 H294206 H294267 H294328 H294389 H294451 H294514 H294581 H294642 H294705 H294767 H294023 H294084 H294147 H294207 H294268 H294329 H294390 H294452 H294515 H294582 H294643 H294706 H294768 H294024 H294085 H294148 H294208 H294269 H294330 H294391 H294454 H294516 H294583 H294644 H294707 H294769 H294025 H294086 H294149 H294209 H294270 H294331 H294392 H294457 H294517 H294584 H294645 H294708 H294770 H294026 H294087 H294150 H294210 H294271 H294332 H294393 H294458 H294518 H294585 H294646 H294709 H294771 H294027 H294088 H294151 H294211 H294272 H294333 H294394 H294459 H294519 H294586 H294647 H294710 H294772 H294028 H294089 H294152 H294212 H294273 H294334 H294395 H294460 H294520 H294587 H294648 H294711 H294773 H294029 H294090 H294153 H294213 H294274 H294335 H294396 H294461 H294521 H294588 H294649 H294712 H294774 H294030 H294091 H294154 H294214 H294275 H294336 H294397 H294462 H294522 H294589 H294650 H294713 H294775 H294031 H294092 H294155 H294215 H294276 H294337 H294398 H294463 H294523 H294590 H294651 H294714 H294776 H294032 H294093 H294156 H294216 H294277 H294338 H294399 H294464 H294524 H294591 H294652 H294715 H294777 H294033 H294094 H294157 H294217 H294278 H294339 H294401 H294465 H294525 H294592 H294653 H294716 H294778 H294034 H294095 H294158 H294218 H294279 H294340 H294402 H294466 H294526 H294593 H294654 H294717 H294779 H294035 H294097 H294159 H294219 H294280 H294341 H294403 H294467 H294527 H294594 H294655 H294718 H294780 H294036 H294098 H294160 H294220 H294281 H294342 H294404 H294468 H294528 H294595 H294656 H294719 H294781 H294037 H294099 H294161 H294221 H294283 H294343 H294405 H294469 H294529 H294596 H294657 H294720 H294782 H294038 H294100 H294162 H294222 H294284 H294345 H294406 H294470 H294530 H294597 H294658 H294721 H294783 H294039 H294101 H294163 H294223 H294285 H294346 H294407 H294471 H294531 H294598 H294659 H294723 H294784 H294040 H294102 H294164 H294224 H294286 H294347 H294408 H294472 H294532 H294599 H294660 H294724 H294785 H294041 H294103 H294165 H294225 H294287 H294348 H294409 H294473 H294533 H294600 H294661 H294726 H294786 H294042 H294104 H294166 H294226 H294288 H294349 H294410 H294474 H294534 H294601 H294663 H294727 H294787 H294043 H294105 H294167 H294227 H294289 H294350 H294411 H294475 H294535 H294602 H294664 H294728 H294788 H294044 H294106 H294168 H294228 H294290 H294351 H294412 H294476 H294536 H294603 H294665 H294729 H294789 H294790 H294854 H294917 H294984 H295049 H295120 H295203 H294791 H294855 H294918 H294986 H295050 H295121 H295206 H294792 H294856 H294919 H294987 H295051 H295122 H295207 H294793 H294857 H294920 H294988 H295053 H295125 H295209 H294794 H294858 H294921 H294989 H295054 H295126 H295211 H294795 H294859 H294922 H294990 H295056 H295127 H295212 H294796 H294860 H294923 H294991 H295057 H295128 H295213 H294797 H294861 H294924 H294992 H295058 H295129 H295216 H294798 H294862 H294925 H294993 H295059 H295130 H295217 H294799 H294863 H294926 H294995 H295060 H295131 H295224 H294800 H294864 H294927 H294996 H295061 H295132 H295225 H294801 H294865 H294928 H294997 H295062 H295133 H295226 H294802 H294866 H294929 H294998 H295063 H295134 H295227 H294803 H294867 H294930 H294999 H295064 H295135 H295231 H294804 H294868 H294931 H295000 H295065 H295136 Hcx3582 H294806 H294869 H294932 H295001 H295066 H295138 Hcx4019 H294807 H294870 H294933 H295002 H295067 H295139 Hcx4138 H294808 H294871 H294934 H295003 H295068 H295140 Hcx4344 H294809 H294872 H294935 H295004 H295069 H295141 Hcx4400 H294810 H294874 H294936 H295006 H295070 H295142 Hcx4566 H294811 H294875 H294937 H295008 H295071 H295143 Hcx4662 H294812 H294876 H294938 H295009 H295072 H295144 Hcx4703 H294813 H294877 H294940 H295010 H295073 H295145 Hcx4849 H294814 H294878 H294941 H295011 H295074 H295146 Hcx4852 H294815 H294879 H294942 H295012 H295075 H295147 Hcx4973 H294816 H294880 H294943 H295013 H295076 H295148 Hcx4994 H294817 H294881 H294945 H295014 H295077 H295149 Hcx5055 H294818 H294882 H294946 H295015 H295078 H295150 Hcx5088 H294820 H294883 H294947 H295016 H295080 H295151 Hcx5113 H294821 H294884 H294949 H295017 H295081 H295152 H294822 H294885 H294950 H295018 H295082 H295155 H294823 H294886 H294951 H295019 H295083 H295156 H294824 H294887 H294952 H295020 H295084 H295157 H294825 H294888 H294953 H295021 H295085 H295158 H294826 H294889 H294954 H295022 H295087 H295159 H294827 H294890 H294955 H295023 H295089 H295160 H294828 H294891 H294956 H295024 H295090 H295161 H294829 H294892 H294957 H295025 H295093 H295162 H294830 H294893 H294958 H295026 H295094 H295169 H294831 H294894 H294959 H295027 H295095 H295170 H294832 H294895 H294960 H295028 H295096 H295171 H294833 H294896 H294961 H295029 H295097 H295172 H294834 H294897 H294962 H295030 H295099 H295173 H294835 H294898 H294963 H295031 H295100 H295174 H294836 H294899 H294965 H295032 H295101 H295176 H294837 H294900 H294966 H295033 H295102 H295177 H294838 H294901 H294967 H295034 H295103 H295178 H294839 H294902 H294969 H295035 H295105 H295179 H294840 H294903 H294970 H295036 H295106 H295182 H294841 H294904 H294971 H295037 H295107 H295183 H294842 H294905 H294972 H295038 H295108 H295184 H294843 H294906 H294974 H295039 H295109 H295185 H294844 H294907 H294975 H295040 H295110 H295186 H294845 H294908 H294976 H295041 H295111 H295188 H294846 H294910 H294977 H295042 H295112 H295193 H294847 H294911 H294978 H295044 H295114 H295195 H294848 H294912 H294979 H295045 H295116 H295196 H294850 H294914 H294980 H295046 H295117 H295199 H294851 H294915 H294981 H295047 H295118 H295201 H294853 H294916 H294982 H295048 H295119 H295202 ***12-20-12, AMENDED INFO ***: H140971, H140942, H295370, H295371, H295271, H295266 |
Recalling Firm/ Manufacturer |
Varian Medical Systems, Inc. Oncology Systems 911 Hansen Way Palo Alto CA 94304-1028
|
For Additional Information Contact |
Michael Pignataro 650-424-6471
|
Manufacturer Reason for Recall |
Varian has received reports in which a user has remotely rotated the gantry into contact with the couch or with the patient, in both the manual mode and the automate mode.
|
FDA Determined Cause 2 |
Device Design |
Action |
The firm, Varian Medical Systems, sent an "URGENT MEDICAL DEVICE CORRECTION/URGENT FIELD SAFETY NOTICE" letter dated September 9, 2011 to all affected customers via Federal Express with attached instructions. The letter provides the description of problem; references, details, recommended user actions, and varian actions.
***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.***
In addition, the letter also states "THIS LETTER IS NOT A SUBSTITUTE FOR THE CAREFUL REVIEW OF ALL USER DOCUMENTATION ASSOCIATED WITH YOUR PARTICULAR EQUIPMENT AND SOFTWARE."
If you have any questions, contact Varian Oncology Help Desk at USA and Canada: 1.888.VARIANS (888.827.4265); Europe +41 41 749 8844 and/or Internet: Oncology Systems customer site-www.myvarian.com and Varian Medical Systems public site-www.varian.com. |
Quantity in Commerce |
5823 devices |
Distribution |
Worldwide Distribution. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IYE and Original Applicant = VARIAN MEDICAL SYSTEMS 510(K)s with Product Code = IYE and Original Applicant = VARIAN MEDICAL SYSTEMS, INC.
|
|
|
|