Class 2 Device Recall IMMULITE/IMMULITE Turbo 1000 CKMB

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall IMMULITE/IMMULITE Turbo 1000 CKMB see related information
Date Initiated by Firm January 25, 2012
Date Posted March 21, 2012
Recall Status1 Terminated 3
Recall Number Z-1272-2012
Recall Event ID 61029
510(K)Number K022118  
Product Classification Fluorometer method, cpk or isoenzymes - Product Code JHX
Product IMMULITE¿/IMMULITE¿ Turbo 1000 CK-MB LSKCP --- Catalog number: LSKCP1 (100 tests), LSKCP5 (500 tests) --- Test code: RCK Color: Orange.

For in vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers --- for the quantitative measurement of creatine kinase isoenzyme MB (CK-MB) in heparinized plasma or serum, as an aid in patient management and the assessment of prognosis of myocardial infarction.
Code Information kit lot 316, 318
Recalling Firm/
Manufacturer
Siemens Healthcare Diagnostics
511 Benedict Ave
Tarrytown NY 10591-5005
For Additional Information Contact Patricia Lazarro
914-524-2955
Manufacturer Reason
for Recall
Siemens Healthcare Diagnostics has observed that the median for the reference range with current kit lots of the CK-MB assay differs from the guidelines published in the Instructions for Use (IFU). This difference applies to all IMMULITE platforms.
FDA Determined
Cause 2
Nonconforming Material/Component
Action Siemens sent an "URGENT DEVICE RECALL" letter dated January 2012 to all affected customers. The letter identifies the product , problem, and actions to be taken by the customers. Siemens recommends that customers reestablish their own medians and reference ranges, in line with the recommendations made in the IFU for these products. A Field Correction Effectiveness Check form was included for customers to complete and return to the firm. Customers with questions or who need additional information should contact their Siemens Technical Solutions Center.
Distribution Worldwide Distribution.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = JHX and Original Applicant = DIAGNOSTIC PRODUCTS CORP.
-
-