| | Class 2 Device Recall Sarns Modular Perfusion System 9000 |  |
| Date Initiated by Firm | June 18, 2012 |
| Date Posted | July 19, 2012 |
| Recall Status1 |
Terminated 3 on April 12, 2013 |
| Recall Number | Z-2052-2012 |
| Recall Event ID |
61479 |
| 510(K)Number | K871131 |
| Product Classification |
Console, heart-lung machine, cardiopulmonary bypass - Product Code DTQ
|
| Product | Sarns Modular Perfusion System 9000, 100V with color screen.
The Sarns Perfusion System 9000 is indicated for use as a heart-lung system for pumping blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment. |
| Code Information |
Catalog Number 164280 and serial number 1019-5379 |
Recalling Firm/ Manufacturer |
Terumo Cardiovascular Systems Corporation 6200 Jackson Road Ann Arbor MI 48103-9586
|
| For Additional Information Contact | 734-741-6173 |
Manufacturer Reason for Recall | Terumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malfunctions were expressed in various failure modes (over sensitivity, false alarms and were found to have various root causes. The malfunctions occurred in systems that were excluded from prior corrective actions or in syst |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Terumo CVS sent an "URGENT MEDICAL DEVICE RECALL" letter dated March 30, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers are asked to complete and return a Customer Response Form via fax to 1-800-292-6551. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions regarding this recall. |
| Quantity in Commerce | 31 units |
| Distribution | Worldwide Distribution-USA (nationwide) and the countries of ARGENTINA, BANGLADESH, BELGIUM, Bolivia, BRAZIL, BRISBANE, CANADA, CHILE, China, COLOMBIA, COLOMBIA, Costa Rica, DENMARK, Dominican Republic, Ecuador, Egypt, FRANCE, GERMANY, Guatemala, HOLLAND, Honduras, Hong Kong, India, INDIA, Indonesia, Israel, ITALY, Jordan, KOREA, Kuwait, Lebanon, LEICESTERSHIRE, Malaysia, Mexico, NETHERLANDS, New Zealand, Nicaragua, NORWAY, Pakistan, Panama, PERU, Philippines, Russia, Saudi Arabia, SINGAPORE , SOUTH AFRICA, South Korea, SPAIN, Sri Lanka, SWEDEN, SWITZERLAND, Taiwan, THAILAND, Turkey, UNITED ARAB EMIRATES (UAE), Uruguay, Venezuela, Vietnam, WEST GERMANY and Yemen. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|