|
Class 2 Device Recall GE Discovery NM/CT670; GE Discovery NM 630 |
 |
Date Initiated by Firm |
October 12, 2012 |
Date Posted |
October 23, 2012 |
Recall Status1 |
Terminated 3 on November 14, 2013 |
Recall Number |
Z-0121-2013 |
Recall Event ID |
63424 |
510(K)Number |
K093514 K111445
|
Product Classification |
System, tomography, computed, emission - Product Code KPS
|
Product |
GE Discovery NM/CT670; GE Discovery NM 630
Product Usage: a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce: NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. |
Code Information |
Mfg Lot or Serial # 21058, 21002, 21013, 21017, 21020, 21029, 21039, 21041, 21049, 21066, 21075, 21079, 21080, 21081, 21082, 21119, 21140, 21144, 21145, 21147, 21150, 96BH8, 21001, 21008, 21034, 21050, 21056, 21104, 21132, 22003, 22020, 22038, 210088, 21042, 21048, 21061, 21077, 21085, 21094, 21115, 21122, 21123, 21137, 21143, 22051
21005, 21072, 21028, 21052, 21076, 21083, 21092, 21101, 21113, 21118, 21131, 21138, 21024, 22001, 22011, 22023, 21019, 21055, 21154, 83BH6 /000, 21007, 21047, 21009, 21021, 21022, 21033, 21035, 21037, 21063, 21064, 21067, 21070, 21073, 21086, 21095, 21121, 21128, 21134, 21141, 21146, 21152, 21153, 22014, 22016, 21003, 21006, 21010, 21027, 21036, 21043, 21046, 21051, 21084, 21108, 21116, 21120, 21124, 21133, 22015, 22028, 21093, 21158, 22021, 21031, 21129, 00ME1, 21004, 21014, 21053, 22005, 21012, 21059, 21151, 22022, 22031, 22035, 21089, 21100, 21136, 22044, 22071, 81BH0 /000, 82BH8 /000, 92BH7 /000, 93BH5 /000, 94BH3, 21016, 21040, 21060, 21103, 21125, 22037, 21038, 22008, 21068, 21126, 22009, 21026, 21011, 21018, 21025, 21071, 21090, 21109, 21110, 21130, 21142, 21160, 21096, 21107, 21091, 21099, 21000, 21023, 21030, 21032, 21044, 21045, 21054, 21062, 21065, 21069, 21074, 21105, 21111, 21114, 21127, 21149, 22002, 22004, 22007, 22010, 22012, 22017, 22018, 22024, 22025, 22026, 22027, 22029, 22030, 22032, 22033, 22034, 22036, 22039, 22040, 22041, 22042, 22043, 22045, 22046, 22047, 22048, 22049, 22050, 22052, 22053, 22054, 22056, 22057, 22059, 22064, 22067, |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
262-513-4122
|
Manufacturer Reason for Recall |
GE Healthcare has recently become aware of a potential safety issue associated with the
table pallet of your Discovery NM/CT 670 and/or Discovery NM630 system which may impact
operation safety.
At scan setup, during gantry rotation towards scanning position, a potential contact between the table-pallet-edge and the detector was identified. Such occurrence may damage the pallet front-edge and
|
FDA Determined Cause 2 |
Software design |
Action |
GE Healthcare, LLC sent an " Urgent Medical Device Correction" letter dated October 12, 2012 to all affected customers. The letter was addressed to Managers of Nuclear Medicine, Hospital Administrators / Risk Mangers, and Mangers of Radiology / Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details and Product Correction. For questions or concerns contact your local GE Healthcare Service Representative. You may also contact GE Healthcare Call Center: United States or Canada: 800-437-1171 and Other countries: 262-896-2890. |
Quantity in Commerce |
210 |
Distribution |
Worldwide distribution: USA (Nationwide) in the states of: AL, CA, FL, GA, IL, IN, KS, LA, MD, MA, MN, MS, MO, NJ, NY, NC, OH, OK, PA, SC, TN, TX, VA, WI and WV. AUSTRALIA, BELGIUM, CANADA, CHINA, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GUADELOUPE, HONG KONG, INDIA, IRELAND, ISRAEL,ITALY, JAPAN, REPUBLIC OF KOREA, NETHERLANDS,NORWAY, POLAND, RUSSIAN FEDERATION, SPAIN, SWITZERLAND, TAIWAN,THAILAND, TURKEY, UNITED KINGDOM. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = KPS and Original Applicant = GE HEALTHCARE 510(K)s with Product Code = KPS and Original Applicant = GE MEDICAL SYSTEMS F.I. HAIFA
|
|
|
|