| Class 2 Device Recall Beckman Coulter AU5800 Clinical Chemistry Analyzer | |
Date Initiated by Firm | October 26, 2012 |
Date Posted | August 02, 2013 |
Recall Status1 |
Terminated 3 on October 18, 2013 |
Recall Number | Z-1873-2013 |
Recall Event ID |
63605 |
510(K)Number | K112412 |
Product Classification |
Analyzer, chemistry (photometric, discrete), for clinical - Product Code JJE
|
Product | Beckman Coulter AU5800 Clinical Chemistry Analyzer, All Serial Numbers.
An automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC), material and other accessories. |
Code Information |
Models, Catalogue Numbers: AU5811-01, A94906; AU5811-02, A94907; AU5811-03, A94908; AU5811-04, A94909; AU5811-06, A94910; AU5821-01, A94911; AU5821-02, A94912; AU5821-03, A94913; AU5821-04, A94914; AU5821-06, A94915; AU5831-01, A94916; AU5831-02, A94917; AU5831-03, A94918; AU5831-04 A94919; AU5831-06 A94920; AU5831-01 A94921; AU5831-02 A94922; AU5831-03 A94923; AU5831-04 A94924; AU5831-06 A94925. |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 250 S Kraemer Blvd Brea CA 92821-6232
|
Manufacturer Reason for Recall | Beckman Coulter is recalling the AU8500 Clinical Chemistry Analyzer because there is no labeling on the diluted detergent tank to show the proper disposition of the on/off valve. |
FDA Determined Cause 2 | Software design |
Action | A recall notification letter was forwarded to all their customers on 11/1/12 who purchased the AU5800 Clinical Chemistry Analyzer. The customers were informed of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Reponse form within 10 days. Customers with questions are instructed to call (800) 854-3633 or contact their local BCI Representative. |
Quantity in Commerce | 338 units (11 units in the US) |
Distribution | Worldwide Distribution-USA (nationwide) and Puerto Rico and the countries of Germany, China, Austria, Brazil, Italy, Argentina, Korea, Vietnam, Hungary, Ireland, France, Spain, Turkey, Switzerland, Israel, Australia, India, United Kingdom, Netherlands, Malaysia, Russian Federation, Japan, Taiwan, and Canada |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|