• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall QuickDraw Venous Cannula

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall QuickDraw Venous Cannulasee related information
Date Initiated by FirmMay 14, 2013
Date PostedJune 13, 2013
Recall Status1 Terminated 3 on October 14, 2014
Recall NumberZ-1542-2013
Recall Event ID 65173
510(K)NumberK981995 
Product Classification Catheter, cannula and tubing, vascular, cardiopulmonary bypass - Product Code DWS
ProductQuickDraw Venous Cannula with Insertion Kit. ThruPort Systmes. Models QD22 (22 Fr) and QD25 (25 Fr). The product is packaged sterile and non-pyrogenic in a sealed, peel-type pouch. Two pouches (cannula and accessories) go into one shelf carton. One shelf carton goes in one shipper carton. Use of the QuickDraw venous cannula is indicated for patients undergoing cardiopulmonary bypass. The QuickDraw venous cannula serves to drain non-oxygenated blood from the venae cavae or right atrium during cardiopulmonary bypass. The product is packaged sterile and nona pyrogenic in a sealed, peel-type pouch.
Code Information Model QD22: Lot No. 59049758, 59054488, 59067502, 59074170, 59081626, 59083736, 59091652, 59094887, 59114755, 59117451, 59134508, 59149651, 59154310, 59178579, 59201243, 59212118, 59222657, 59233636, 59233637, 59246684, 59246701, 59255899, 59255900, 59269439, 59284780, 59287725, 59296462, 59299605, 59315339, 59320575, 59334652, 59356129, 59360072, 59365198, 59365199, 59374009, 59384309, 59390552, 59394166, 59394167, 59421250, 59448841  Model QD25: Lot No. 59046082,59046847,59052393,59052395,59056270,59059029,59060798,59062681,59069772,59073141,59077234,59079573,59085168,59085169,59097289,59101637,59106005,59110720,59122491,59126281,59131923,59140133,59142266,59143453,59145027,59158524,59158579,59163164,59184252,59191107,59201246,59207044,59209886,59214360,59214361,59216991,59222658,59222659,59233651,59246685,59246703,59255901,59268625,59284781,59287728,59296463,59299610,59313057,59320582,59320583,59334656,59344191,59346468,59352608,59356135,59360084,59374014,59374015,59414791,59425355,59437699,59437700,59448853
Recalling Firm/
Manufacturer
Edwards Lifesciences, LLC
12050 Lone Peak Pkwy
Draper UT 84020-9414
For Additional Information ContactSuzanne Carpenter
801-565-6195
Manufacturer Reason
for Recall
Edwards is recalling the QuickDraw Venous Cannula due to complaints of cannula separation during withdrawal of the device when the Percutaneous insertion method is used.
FDA Determined
Cause 2
Process design
ActionThe firm, Edwards Lifesciences, sent an "URGENT FIELD SAFETY NOTICE PRODUCT RECALL - ACTION REQUIRED" notice dated May 14, 2013 to its customers. The letter described the product, problem, and actions to be taken. The customers were instructed to review our entire inventory for any QuickDraw Venus Cannula that has not expired; quarantine affected product from your inventory; discontinue use and return unused devices; complete and return the acknowledgment form via fax to Edwards Customer Service on 800-422-9329 within three days of receipt of this Field Safey Notice; contact Customer Service at 800-424-3278 to obtain a Returned Goods Authorization number and replacement product; return product to Edwards Lifesciences, Attn: Santosh Bhagat, 12050 Lone Peak Drive, Draper, UT 84020 Attention: RECALL, RGA#xxxxxx. If you have any questions that have not been answered by this notice, please call Edwards Customer Service at 800-424-3278 from the hours of 6:00AM - 4:30PM PST.
Quantity in Commerce22,616
DistributionWorldwide distribution: US Nationwide and country of Canada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = DWS
-
-