|
Class 2 Device Recall IsoGel Air |
 |
Date Initiated by Firm |
July 12, 2013 |
Date Posted |
August 09, 2013 |
Recall Status1 |
Terminated 3 on September 11, 2014 |
Recall Number |
Z-1946-2013 |
Recall Event ID |
65337 |
Product Classification |
Bed, flotation therapy, powered - Product Code IOQ
|
Product |
IsoGel AIR 84, Support Surface.
Intended to assist in improving human patient outcomes.
|
Code Information |
Model Numbers: 2860-000-001 and 2860-000-002 and serial numbers: 121228600100355, 121228600100356, 121228600100357, 121228600200315, 121228600200316 121228600200322, 121228600200323, 121228600200324, 121228600200325, 121228600200326 121228600200332, 121228600200333, 121228600200334, 121228600200335, 121228600200336 121228600200342, 121228600200343, 121228600200344, 121228600200345, 121228600200346 121228600200352, 121228600200353, 121228600200367, 121228600200368, 121228600200369 121228600200375, 121228600200376, 121228600200377, 121228600200378, 121228600200379 121228600200386, 121228600200393, 121228600200394, 121228600200395, 121228600200396 121228600200402, 121228600200403, 121228600200404, 121228600200405, 011328600100406 011328600100412, 011328600100413, 011328600100414, 011328600100415, 011328600100416 011328600100422, 011328600100423, 011328600100424, 011328600100425, 011328600100426 011328600100432, 011328600100433, 011328600100434, 011328600100435, 011328600100436 011328600100442, 011328600100443, 011328600100444, 011328600100445, 011328600100446 011328600100452, 011328600100453, 011328600100454, 011328600100455, 011328600100456 011328600100462, 011328600100463, 011328600100464, 011328600100465, 011328600100466 011328600100474, 011328600100475, 011328600100476, 011328600100477, 011328600100478 011328600100492, 011328600100493, 011328600100494, 011328600100495, 011328600100497 011328600100611, 011328600100612, 011328600100615, 011328600100617, 011328600100618 011328600100625, 011328600100626, 011328600100629, 011328600100630, 011328600100631 011328600100637, 011328600100638, 011328600100639, 011328600100640, 011328600100641 011328600100647, 011328600100648, 011328600100649, 011328600100650, 011328600100651 011328600100657, 011328600100658, 011328600100659, 011328600100660, 011328600100661 011328600100667, 011328600100668, 011328600100669, 011328600100670, 011328600100671 011328600100677, 011328600100678, 011328600100679, 011328600100680, 011328600100681 011328600100687, 011328600100688, 011328600100689, 011328600100690, 011328600100691 011328600100697, 011328600100698, 011328600100699, 011328600100700, 011328600100701 011328600100707, 011328600100708, 011328600100709, 011328600100710, 011328600100711 011328600100717, 011328600100718, 011328600100719, 011328600100720, 011328600100746 011328600100752, 011328600100753, 011328600100754, 011328600100755, 011328600100756 011328600100762, 011328600100763, 011328600100764, 011328600100765, 011328600100766 011328600100772, 011328600100773, 011328600100774, 011328600100775, 011328600100777 011328600100783, 011328600100784, 011328600100785, 011328600100786, 011328600100787 011328600100793, 011328600100794, 011328600100795, 011328600100798, 011328600100799 011328600100805, 011328600100806, 011328600100807, 011328600100808, 011328600100809 011328600100815, 011328600100816, 011328600100817, 011328600100818, 011328600100819 011328600100825, 011328600100826, 011328600100827, 011328600100828, 011328600100829 011328600100835, 011328600100836, 011328600100837, 011328600100838, 011328600100839 011328600100845, 011328600100846, 011328600100847, 011328600100848, 011328600100849 011328600100856, 011328600100857, 011328600100858, 011328600100859, 011328600100861 011328600100867, 011328600100868, 011328600100869, 011328600100870, 011328600100871 011328600100879, 011328600100880, 011328600100881, 011328600100885, 011328600100886 011328600100893, 011328600100894, 011328600100895, 011328600100896, 011328600100897 011328600200511, 011328600200512, 011328600200513, 011328600200514, 011328600200515 011328600200526, 011328600200527, 011328600200528, 011328600200529, 011328600200530 011328600200536, 011328600200537, 011328600200538, 011328600200539, 011328600200540 011328600200546, 011328600200547, 011328600200548, 011328600200549, 011328600200550 011328600200556, 011328600200557, 011328600200558, 011328600200559, 011328600200560 011328600200566, 011328600200567, 011328600200568, 011328600200569, 011328600200570 011328600200576, 011328600200577, 011328600200578, 011328600200579, 011328600200580 011328600200586, 011328600200587, 011328600200588, 011328600200589, 011328600200590 011328600200596, 011328600200597, 011328600200598, 011328600200599, 011328600200600 011328600200721, 011328600200722, 011328600200723, 011328600200724, 011328600200725 011328600200731, 011328600200732, 011328600200733, 011328600200734, 011328600200735 011328600200741, 011328600200742, 011328600200743, 011328600200744, 011328600200745 011328600100499, 011328600100503, 011328600100504, 011328600100606, 011328600100607 011328600201034, 021328600201035, 021328600201036, 021328600201037, 021328600201038 021328600201044, 021328600201045, 021328600201046, 021328600201047, 021328600201048 021328600201078, 021328600201079, 021328600201080, 021328600201081, 021328600201082 021328600201161, 121228600100354, 011328600200860, 011328600200878, 011328600200882 011328600200900, 011328600200901, 011328600200902, 011328600200903, 011328600200904 021328600201153, 021328600201159, 021328600201160, 011328600100502, 011328600100912 011328600101111, 011328600101113, 021328600201162, 021328600201163, 021328600201164 121228600200388, 121228600200389, 121228600200390, 121228600200391, 121228600100359 121228600100365, 121228600100366, 121228600200309, 121228600200310, 121228600200311 121228600200317, 121228600200318, 121228600200319, 121228600200320, 121228600200321 121228600200327, 121228600200328, 121228600200329, 121228600200330, 121228600200331 121228600200337, 121228600200338, 121228600200339, 121228600200340, 121228600200341 121228600200347, 121228600200348, 121228600200349, 121228600200350, 121228600200351 121228600200370, 121228600200371, 121228600200372, 121228600200373, 121228600200374 121228600200381, 121228600200382, 121228600200383, 121228600200384, 121228600200385 121228600200397, 121228600200398, 121228600200399, 121228600200400, 121228600200401 011328600100407, 011328600100408, 011328600100409, 011328600100410, 011328600100411 011328600100417, 011328600100418, 011328600100419, 011328600100420, 011328600100421 011328600100427, 011328600100428, 011328600100429, 011328600100430, 011328600100431 011328600100437, 011328600100438, 011328600100439, 011328600100440, 011328600100441 011328600100447, 011328600100448, 011328600100449, 011328600100450, 011328600100451 011328600100457, 011328600100458, 011328600100459, 011328600100460, 011328600100461 011328600100467, 011328600100470, 011328600100471, 011328600100472, 011328600100473 011328600100479, 011328600100487, 011328600100488, 011328600100489, 011328600100490 011328600100500, 011328600100501, 011328600100608, 011328600100609, 011328600100610 011328600100619, 011328600100620, 011328600100621, 011328600100623, 011328600100624 011328600100632, 011328600100633, 011328600100634, 011328600100635, 011328600100636 011328600100642, 011328600100643, 011328600100644, 011328600100645, 011328600100646 011328600100652, 011328600100653, 011328600100654, 011328600100655, 011328600100656 011328600100662, 011328600100663, 011328600100664, 011328600100665, 011328600100666 011328600100672, 011328600100673, 011328600100674, 011328600100675, 011328600100676 011328600100682, 011328600100683, 011328600100684, 011328600100685, 011328600100686 011328600100692, 011328600100693, 011328600100694, 011328600100695, 011328600100696 011328600100702, 011328600100703, 011328600100704, 011328600100705, 011328600100706 011328600100712, 011328600100713, 011328600100714, 011328600100715, 011328600100716 011328600100747, 011328600100748, 011328600100749, 011328600100750, 011328600100751 011328600100757, 011328600100758, 011328600100759, 011328600100760, 011328600100761 011328600100767, 011328600100768, 011328600100769, 011328600100770, 011328600100771 011328600100778, 011328600100779, 011328600100780, 011328600100781, 011328600100782 011328600100788, 011328600100789, 011328600100790, 011328600100791, 011328600100792 011328600100800, 011328600100801, 011328600100802, 011328600100803, 011328600100804 011328600100810, 011328600100811, 011328600100812, 011328600100813, 011328600100814 011328600100820, 011328600100821, 011328600100822, 011328600100823, 011328600100824 011328600100830, 011328600100831, 011328600100832, 011328600100833, 011328600100834 011328600100840, 011328600100841, 011328600100842, 011328600100843, 011328600100844 011328600100850, 011328600100851, 011328600100852, 011328600100854, 011328600100855 011328600100862, 011328600100863, 011328600100864, 011328600100865, 011328600100866 011328600100872, 011328600100873, 011328600100874, 011328600100875, 011328600100876 011328600100888, 011328600100889, 011328600100890, 011328600100891, 011328600100892 011328600200506, 011328600200507, 011328600200508, 011328600200509, 011328600200510 011328600200516, 011328600200517, 011328600200518, 011328600200519, 011328600200520 011328600200531, 011328600200532, 011328600200533, 011328600200534, 011328600200535 011328600200541, 011328600200542, 011328600200543, 011328600200544, 011328600200545 011328600200551, 011328600200552, 011328600200553, 011328600200554, 011328600200555 011328600200561, 011328600200562, 011328600200563, 011328600200564, 011328600200565 011328600200571, 011328600200572, 011328600200573, 011328600200574, 011328600200575 011328600200581, 011328600200582, 011328600200583, 011328600200584, 011328600200585 011328600200591, 011328600200592, 011328600200593, 011328600200594, 011328600200595 011328600200601, 011328600200602, 011328600200603, 011328600200604, 011328600200605 011328600200726, 011328600200727, 011328600200728, 011328600200729, 011328600200730 011328600200736, 011328600200737, 011328600200738, 011328600200739, 011328600200740 121228600100358, 011328600100468, 011328600100469, 011328600100496, 011328600100498 011328600100613, 011328600100614, 011328600100616, 011328600100627, 011328600100628 021328600201039, 021328600201040, 021328600201041, 021328600201042, 021328600201043 021328600201049, 021328600201050, 021328600201051, 021328600201058, 021328600201077 011328600200521, 011328600200522, 011328600200523, 011328600200524, 011328600200525 011328600200883, 011328600200884, 011328600200887, 011328600200898, 011328600200899 011328600200905, 011328600200930, 021328600201150, 021328600201151, 021328600201152 011328600100924, 011328600100925, 011328600100931, 011328600100932, 011328600100933 021328600201165, 021328600201166, 021328600201167, 121228600200380, 121228600200387 121228600100360, 121228600100361, 121228600100362, 121228600100363, 121228600100364 121228600200312, 121228600200313, 121228600200314, 121228600200354, and 121228600200392. |
Recalling Firm/ Manufacturer |
Stryker Medical Division of Stryker Corporation 3800 E Centre Ave Portage MI 49002-5826
|
For Additional Information Contact |
Dennis Waters 269-389-8129
|
Manufacturer Reason for Recall |
Complaints were received from the field indicating that some Air Pumps were flashing and alarming when connected to the IsoGel AIR" surface upon turning the pump on and the setting it to the IsoGel AIR Mode. Upon further investigation, it was determined that the flashing/alarm being experienced was due to the air pressure being out of the expected range; indicating a blockage within the LAL tubi
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
On 7/12/13, an URGENT MEDICAL DEVICE RECALL NOTIFICATION was mailed to consignees. The notification addressed the affected product and issue, potential hazard, actions needed including faxing (269-488-8691) or emailing (ProductFieldAction@stryker.com) the attached reply form to confirm receipt of notification, and risk mitigation. Please contact the firm at 269-389-8129 if you have questions or concerns regarding this recall. Business hours are M-F 8am-5pm(ET). |
Quantity in Commerce |
586 units |
Distribution |
Nationwide Distribution-USA including the states of OH, ME, IL, CO , FL, TX, MI, GA, and VA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|