|
Class 2 Device Recall TRAM module |
 |
Date Initiated by Firm |
April 03, 2013 |
Date Posted |
August 28, 2013 |
Recall Status1 |
Terminated 3 on July 25, 2016 |
Recall Number |
Z-2117-2013 |
Recall Event ID |
66016 |
510(K)Number |
K053121
|
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product |
GE Healthcare, TRAM MODULE, High-acuity patient monitoring.
Product Usage: The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. |
Code Information |
Mfg Lot or Serial # RT412081449GA RT412081483GA RT412091489GA RT412091490GA RT412091492GA RT412091510GA RT412091511GA RT412091517GA RT412091522GA SBB12281535GA RT412101534GA SBB12281525GA SBB12281528GA SBB12281561GA SBB12301640GA SBB12361656GA SBB12361658GA SBB12361659GA SBB12361660GA SBB12361661GA SBB12361662GA SBB12361663GA SBB12301635GA RT411220584GA RT412081484GA RT412091488GA RT412091501GA RT412101576GA RT412101599GA RT412101601GA RT412101604GA RT412101605GA RT412101606GA RT412101609GA RT412101611GA RT412101621GA RT412101624GA RT412101626GA RT412101628GA RT412101629GA RT412101632GA RT412101646GA RT412101655GA RT412101657GA RT412101661GA RT412101662GA RT412101663GA RTW11346657GA RTW11346667GA RTW11366669GA RTW11366672GA RTW11366673GA RTW11366674GA RTW11366691GA RTW11366695GA RTW11366697GA RTW11366700GA RTW11366701GA RTW11366702GA RTW11366704GA RTW11366712GA RTW11366715GAMKE RTW11366721GA RTW11366722GA RTW11366724GA RTW11366725GA RTW11366726GA RTW11366731GA RTW11366736GA RTW11366737GA RTW11366738GA RTW11366739GA RTW11366742GA RTW11366747GA RTW11366748GA RTW11366751GA RTW11366752GA SBB12281564GA SBB12281575GA SBB12301617GA SBB12301618GA SBB12301620GA SBB12301626GA SBB12301627GA SBB12301628GA SBB12301630GA SBB12301631GA SBB12131373GA SBB12301629GA SBB12281536GA SBB12281539GA SBB12281542GA SBB12281549GA SBB12281563GA SBB12301623GA SBB12301624GA SBB12301632GA SBB12301633GA SBB12301634GA SBB12301636GA SBB12301637GA SBB12301638GA SBB12301639GA SBB12301641GA SBB12301642GA SBB12301643GA SBB12301644GA SBB12301645GA SBB12301646GA RT412081429GA RT412081448GA RT412091507GA RT412101533GA RT412101536GA RTW11366717GA RTW11366732GA RTW11376817GA RTW11376819GA RTW11376821GA SBB12281533GA SBB12281537GA SBB12281538GA SBB12281540GA SBB12281545GA SBB12281548GA SBB12281551GA SBB12281552GA SBB12281553GA SBB12281554GA SBB12281556GA SBB12281557GA SBB12281558GA SBB12281562GA SBB12281565GA SBB12281566GA SBB12281574GA SBB12281576GA SBB12301621GA SBB12301622GA SBB12301625GA RTW11366716GA SBB12281547GA SBB12281559GA SBB12281531GA RT412101598GA SBB12261509GA SBB12281546GA RT412101597GA RT412101600GA RT412101603GA RT412101615GA RT412101633GA RT412101637GA SBB12281526GA SBB12281529GA RTW11366684GA RTW11366687GA RTW11366688GA RTW11366694GA RTW11366696GA RTW11366720GA SBB12281522GA SBB12281523GA SBB12281527GA SBB12281532GA RT412101660GA SBB12261508GA SBB12261510GA SBB12261511GA SBB12261512GA SBB12261514GA RT412081435GA RT412091498GA RT412091512GA RT412091515GA RT412101563GA RT412101588GA RT412101590GA RTW11245985GA SBB12111305GA SBB12121312GA SBB12121318GA SBB12121322GA SBB12131356GA SBB12131365GA SBB12131379GA SBB12131386GA SBB12131396GA SBB12131398GA RTW11316250GA RTW11316392GA RTW11346656GA RTW11366685GA RTW11366692GA SBB12131402GA SBB12131408GA SBB12131393GA SBB12131391GA SBB12131409GA RTW11346632GA RTW11346640GA RTW11346642GA RTW11366777GA RTW11376799GA SBB12131392GA SBB12131407GA RT412101619GA RT412101640GA RT412101643GA SBB12131353GA SBB12131370GA SBB12131375GA SBB12131383GA SBB12131385GA SBB12131388GA SBB12131374GA SBB12131359GA SBB12131389GA RT412091503GA RT412101672GA RTW11366761GA RTW11446934GA SBB12131361GA SBB12131362GA SBB12131338GA RT412091493GA RT412091495GA SBB12020996GA SBB12031051GA SBB12041074GA SBB12041082GA SBB12041084GA SBB12041092GA SBB12041115GA SBB12041125GA SBB12051139GA SBB12051146GA SBB12051148GA SBB12051152GA SBB12061189GA SBB12061196GA SBB12061198GA SBB12061210GA SBB12061214GA SBB12061222GA SBB12071234GA SBB12071249GA SBB12071259GA SBB12071263GA SBB12071271GA SBB12111287GA SBB12111295GA SBB12111298GA SBB12111303GA SBB12111306GA SBB12111307GA SBB12121308GA SBB12121310GA SBB12121311GA SBB12121313GA SBB12121314GA SBB12121315GA SBB12121316GA SBB12121317GA SBB12121319GA SBB12121320GA SBB12121321GA SBB12131340GA SBB12131341GA SBB12131342GA SBB12131346GA SBB12131348GA SBB12131349GA SBB12131351GA SBB12131352GA SBB12131354GA SBB12131357GA SBB12131358GA SBB12131360GA SBB12131366GA SBB12131371GA SBB12131372GA SBB12131377GA SBB12131378GA SBB12131382GA SBB12131384GA SBB12131387GA SBB12131390GA SBB12131397GA SBB12131399GA SBB12131403GA SBB12131405GA RT412071428GA RT412081434GA RT412081439GA RT412081442GA RT412081445GA RT412081459GA RT412081464GA RT412081471GA RT412081480GA RT412091497GA RT412101577GA RT412081437GA RT412081473GA RT412081474GA RT412101568GA RT412101575GA RT412101579GA RT412101584GA RT412101589GA RT412101593GA SBB12131394GA RTW11346665GA RTW11346668GA RT412101581GA RT412081444GA RT412081476GA RT412081477GA RT412081479GA SBB12071244GA RT412081446GA RT412081447GA RT412091491GA SBB12071247GA SBB12071254GA RT412081451GA RT412081453GA RT412081454GA RT412081455GA RT412081458GA RT412081461GA RT412081467GA RT412081468GA RT412081478GA RT412081481GA SBB12061187GA SBB12061188GA SBB12061195GA SBB12071227GA SBB12071228GA SBB12071229GA SBB12071235GA SBB12071241GA SBB12071246GA SBB12071248GA SBB12071250GA SBB12071251GA SBB12071252GA SBB12071255GA SBB12071256GA RT412081433GA SBB12051151GA SBB12051155GA SBB12071231GA SBB12071232GA SBB12071236GA SBB12031037GA SBB12031038GA SBB12071240GA SBB12020993GA SBB12021011GA SBB12041068GA SBB12041071GA SBB12041087GA SBB12041110GA SBB12041114GA SBB12041119GA SBB12051149GA SBB12051150GA RT412081452GA SBB12031062GA SBB12041072GA SBB12041105GA SBB12041112GA SBB12051147GA SBB12051171GA SBB12051172GA SBB12051173GA SBB12051174GA SBB12051175GA SBB12051176GA SBB12051177GA SBB12051178GA SBB12051179GA SBB12051180GA SBB12051181GA SBB12051182GA SBB12051183GA SBB12051184GA SBB12051185GA SBB12061190GA SBB12061191GA SBB12061197GA SBB12061200GA SBB12061201GA SBB12061202GA SBB12061203GA SBB12061204GA SBB12061205GA SBB12061206GA SBB12061207GA SBB12061208GA SBB12061209GA RT412081430GA RT412081431GA RT412081432GA RT412081438GA RT412081443GA RT412081456GA RTW11346659GA RTW11346663GA RTW11346664GA SBB12020997GA SBB12021001GA SBB12031034GA SBB12031047GA SBB12031054GA SBB12031061GA SBB12041069GA SBB12041070GA SBB12041073GA SBB12041077GA SBB12041078GA SBB12041079GA SBB12041080GA SBB12041081GA SBB12041083GA SBB12041088GA SBB12041090GA SBB12041093GA SBB12041094GA SBB12041095GA SBB12041099GA SBB12041104GA SBB12041113GA SBB12041118GA SBB12041132GA SBB12051134GA SBB12051143GA SBB12051144GA SBB12051145GA SBB12051154GA SBB12051156GA SBB12041076GA SBB11440859GA SBB12041075GA SBB12051135GA SBB11410818GA SBB12041107GA SBB12041100GA RT412071425GA RT412071426GA RT412081466GA RT412071424GA RT412021415GA RT412071423GA RT412071427GA SBB12041102GA SBB12041127GA SBB12031035GA SBB12041122GA RT412011402GA SBB12041097GA SBB12041098GA SBB12041111GA RT412021412GA SBB12031056GA RTW11346605GA SBB12031040GA SBB12031042GA SBB12031044GA SBB12031045GA SBB12031049GA SBB12021002GA RT412021404GA RT412021405GA RT412021406GA RT412021407GA RT412021408GA RT412021409GA RT412021413GA RT412021414GA RT412021417GA RT412021418GA RT412021419GA SBB12031041GA RT412011392GA RT412011394GA RT412011395GA RT412011396GA RT411110344GA RT411110381GA RT411381067GA RT411381098GA RT412011393GA RT412011398GA RT412011399GA RT412011401GA SBB12020999GA SBB12021004GA SBB12031020GA SBB12020990GA SBB12020992GA SBB12020995GA SBB12021003GA RT411110382GA RT412011381GA RT412011384GA RT412011385GA RT412011386GA RT412011387GA RT412011389GA RT412011390GA RT412011391GA SBB10349892GA RT412011377GA RT412011379GA RT412011380GA RT411521366GA RT412011378GA RT411391202GA RT411401220GA RT411401222GA RT411401223GA RT411401224GA RT411401233GA RT411401236GA RT411401238GA RT411401239GA RT411401240GA RT411401241GA RT411411248GA RT411411253GA RT411411256GA RT411411262GA RT411411263GA RT411411268GA RT411411271GA RT411411272GA RT411411273GA RT411411274GA RT411411277GA RT411411281GA RT411411282GA RT411411285GA RT411411286GA RT411451293GA RT411451294GA RT411451298GA RT411511304GA RT411511305GA RT411511306GA RT411511308GA RT411511309GA RT411511313GA RT411511314GA RT411511315GA RT411511316GA RT411511317GA RT411511320GA RT411511321GA RT411511322GA RT411511323GA RT411511324GA RT411511326GA RT411521327GA RT411521328GA RT411521329GA RT411521330GA RT411521331GA RT411521332GA RT411521333GA RT411521334GA RT411521335GA RT411521336GA RT411521337GA RT411521338GA RT411521339GA RT411521340GA RT411521341GA RT411521342GA RT411521343GA RT411521344GA RT411521345GA RT411521346GA RT411521347GA RT411521348GA RT411521349GA RT411521350GA RT411521351GA RT411521352GA RT411521353GA RT411521354GA RT411521355GA RT411521356GA RT411521357GA RT411521358GA RT411521359GA RT411521360GA RT411521361GA RT411521362GA RT411521363GA RT411521364GA RT411521365GA RT411521367GA RT411521368GA RT411521369GA RT411521370GA RT411521371GA RT411521372GA RT411521373GA RT411521374GA RT411521375GA RT411521376GA RT411411254GA RT411411283GA RT411511307GA RT411511318GA RT411511319GA RT411511325GA RTW11346626GA SBB11490907GA SBB11140631GA SBB11400796GA SBB11490910GA SBB11490925GA SBB11510947GA SBB11510953GA SBB11510955GA SBB11510956GA SBB11510957GA SBB11510958GA SBB11510959GA SBB11510948GA SBB11490893GA SBB11490927GA SBB11490936GA SBB11490937GA SBB11500938GA SBB11500939GA SBB11510940GA SBB11510942GA SBB11510943GA SBB11510944GA SBB11510945GA SBB11510946GA SBB11510949GA SBB11510950GA SBB11510952GA RTW11316345GA RTW11346590GA RTW11346594GA SBB11490922GA SBB11490933GA SBB11510941GA RT411401204GA RT411401225GA RT411401235GA RT411401242GA RT411411249GA RT411411250GA RT411411252GA RT411411257GA RT411411258GA RT411411260GA RT411411265GA RT411451289GA RT411451290GA RT411451291GA RT411451292GA RT411451295GA RT411451296GA RT411451297GA RT411451299GA RT411451300GA RT411451301GA RT411451303GA RTW11346619GA SBB11490930GA SBB11490931GA SBB11490900GA SBB11490916GA SBB11490923GA SBB11490929GA SBB11490935GA SBB11490894GA SBB11490899GA SBB11490903GA SBB11490904GA SBB11490928GA SBB11490932GA SBB11490934GA SBB11490872GA SBB11490908GA RT411411278GA RT411411280GA RTW11346586GA SBB11490873GA SBB11490876GA SBB11450871GA SBB11410831GA SBB11440857GA SBB11440862GA SBB11440869GA SBB11490874GA SBB11490875GA SBB11170660GA SBB11170684GA SBB11170761GA SBB11390764GA SBB11390765GA SBB11390766GA SBB11390767GA SBB11390768GA SBB11390769GA SBB11390770GA SBB11390771GA SBB11390772GA SBB11390773GA SBB11390774GA SBB11390775GA SBB11390776GA SBB11390777GA SBB11390778GA RTW11346631GA SBB11110586GA SBB11110591GA SBB11110596GA SBB11170682GA RT411401231GA RT411401243GA RT411401244GA RT411411251GA RT411411264GA RT411411266GA RT411411267GA RT411411270GA RT411411284GA RT411401208GA SBB11410818GA SBB11440859GA SBB11410828GA SBB11410846GA SBB11410847GA RT411391175GA SBB11440867GA SBB11440868GA SBB11440870GA SBB11410830GA SBB11410845GA SBB11410853GA SBB11410813GA SBB11410844GA SBB11410841GA RT411401206GA SBB11410843GA SBB11410835GA RT411391157GA RT411391160GA RT411391162GA RT411210506GA RT411381085GA RT411381103GA RT411391145GA RT411391148GA RT411391149GA RT411391166GA RT411391170GA RT411391173GA RT411391185GA RT411391189GA RT411391197GA RT411391198GA RT411391199GA RT411391200GA SBB11400797GA SBB11400803GA SBB11400805GA SBB11410823GA SBB11050526GA SBB11050562GA SBB11060499GA SBB11400804GA SBB11390779GA SBB11390781GA SBB11390783GA SBB11390784GA SBB11390785GA RT411110339GAMKE RT411110341GA RT411110346GA RT411110352GA RT411110358GA RT411110361GA RT411110374GA RT411140396GA RT411170432GAMKE RT411170460GA RT411180477GA RT411180479GA RT411180480GA RT411180484GA RT411180485GAMKE RT411180488GAMKE RT411180489GA RT411210501GA RT411210503GA RT411210504GA RT411210505GA RT411210508GA RT411210509GA RT411210511GA RT411210514GA RT411210516GA RT411210517GA RT411210518GA RT411210520GA RT411210521GA RT411210524GA RT411210526GA RT411210529GA RT411210530GA RT411210531GA RT411210534GA RT411210535GA RT411210537GA RT411210539GA RT411210540GA RT411220546GA RT411220547GA RT411220548GA RT411220550GA RT411220551GA RT411220554GA RT411220566GA RT411220567GA RT411220568GA RT411220569GA RT411220571GA RT411220572GA RT411220573GA RT411220574GA RT411220575GA RT411220577GAMKE RT411220578GA RT411220579GA RT411220580GA RT411220583GA RT411240586GA RT411240587GA RT411240588GA RT411240589GA RT411240590GA RT411240591GA RT411240593GA RT411240594GA RT411240595GA RT411240598GA RT411240599GA RT411240601GA RT411240603GA RT411240604GA RT411240606GA RT411240607GA RT411240608GA RT411240610GA RT411240611GA RT411240614GA RT411240615GA RT411240643GA RT411240644GA RT411240645GA RT411240649GA RT411240651GA RT411240652GA RT411240654GA RT411240655GA RT411240658GA RT411240660GA RT411240665GA RT411250688GA RT411250689GA RT411260698GA RT411260703GA RT411260709GA RT411260711GA RT411260716GA RT411290728GA RT411290729GA RT411290730GA RT411290731GA RT411290732GA RT411290733GA RT411290734GA RT411290735GA RT411290736GA RT411290737GA RT411290738GA RT411290739GA RT411290740GA RT411290741GA RT411290742GA RT411290743GA RT411290744GA RT411290745GA RT411290746GA RT411290747GA RT411290748GA RT411290749GA RT411290750GA RT411290751GA RT411290752GA RT411290753GA RT411290754GA RT411290755GA RT411290756GA RT411290757GA RT411290758GA RT411290759GA RT411290760GA RT411290761GA RT411290762GA RT411290763GA RT411290764GA RT411290765GA RT411290766GA RT411290767GA RT411290768GA RT411290838GA RT411290839GA RT411290840GA RT411290841GA RT411290842GA RT411290843GA RT411290844GA RT411290845GA RT411290846GA RT411290847GA RT411290848GA RT411290849GA RT411290850GA RT411290851GA RT411290852GA RT411290853GA RT411290854GA RT411290855GA RT411290856GA RT411290857GA RT411290858GA RT411290859GA RT411290860GA RT411290861GA RT411290862GA RT411290863GA RT411290864GA RT411290865GA RT411290866GA RT411290868GA RT411290870GA RT411290871GA RT411290872GA RT411290873GA RT411290874GA RT411290875GA RT411290876GA RT411290877GA RT411290878GA RT411290879GA RT411290880GA RT411290882GA RT411290883GA RT411300884GA RT411300885GA RT411300886GA RT411300887GA RT411300888GA RT411300889GA RT411300890GA RT411300891GA RT411300893GA RT411300894GA RT411300895GA RT411300897GA RT411300898GA RT411300899GA RT411300900GA RT411300901GA RT411300902GA RT411300903GA RT411300904GA RT411300905GA RT411300906GA RT411300907GA RT411300908GA RT411300909GA RT411300910GA RT411300911GA RT411300912GA RT411300913GA RT411300914GA RT411300915GA RT411300916GA RT411300917GA RT411300918GA RT411300919GA RT411300920GA RT411300921GA RT411300922GA RT411300923GA RT411300924GA RT411300925GA RT411300926GA RT411300927GA RT411300930GA RT411300931GA RT411300932GA RT411300933GA RT411300934GA RT411300935GA RT411300936GA RT411300937GA RT411300938GA RT411300939GA RT411300940GA RT411300941GA RT411300942GA RT411300943GA RT411300944GA RT411300946GA RT411300947GA RT411300948GA RT411300949GA RT411300950GA RT411300952GA RT411300953GA RT411300954GA RT411300955GA RT411300956GA RT411300957GA RT411300958GA RT411300959GA RT411320960GA RT411320961GA RT411320962GA RT411320963GA RT411320964GA RT411320965GA RT411320966GA RT411320967GA RT411320968GA RT411320969GA RT411320970GA RT411320971GA RT411320972GA RT411320973GA RT411320974GA RT411320975GA RT411320976GA RT411320977GA RT411320978GA RT411320979GA RT411330980GA RT411330981GA RT411330983GA RT411330984GA RT411330985GA RT411330986GA RT411330987GA RT411330989GA RT411330990GA RT411330991GA RT411330992GA RT411330993GA RT411330994GA RT411330995GA RT411330996GA RT411330997GA RT411330998GA RT411330999GA RT411331000GA RT411331001GA RT411331002GA RT411331003GA RT411331004GA RT411331005GA RT411331006GA RT411331007GA RT411331008GA RT411331009GA RT411381010GA RT411381011GA RT411381012GA RT411381013GA RT411381014GA RT411381015GA RT411381016GA RT411381017GA RT411381018GA RT411381019GA RT411381020GA RT411381021GA RT411381022GA RT411381023GA RT411381024GA RT411381025GA RT411381026GA RT411381027GA RT411381028GA RT411381029GA RT411381030GA RT411381031GA RT411381032GA RT411381033GA RT411381034GA RT411381035GA RT411381036GA RT411381037GA RT411381038GA RT411381039GA RT411381040GA RT411381041GA RT411381042GA RT411381043GA RT411381044GA RT411381045GA RT411381046GA RT411381047GA RT411381048GA RT411381049GA RT411381050GA RT411381051GA RT411381052GA RT411381053GA RT411381054GA RT411381055GA RT411381056GA RT411381057GA RT411381058GA RT411381059GA RT411381060GA RT411381061GA RT411381062GA RT411381063GA RT411381064GA RT411381065GA RT411381066GA RT411381068GA RT411381069GA RT411381070GA RT411381071GA RT411381072GA RT411381073GA RT411381074GA RT411381075GA RT411381076GA RT411381077GA RT411381078GA RT411381079GA RT411381080GA RT411381081GA RT411381083GA RT411381084GA RT411381087GA RT411381088GA RT411381089GA RT411381091GA RT411381093GA RT411381094GA RT411381099GA RT411381100GA RT411381101GA RT411381102GA RT411381106GA RT411381107GA RT411381111GA RT411381112GA RT411381113GA RT411381114GA RT411381115GA RT411381116GA RT411381117GA RT411381082GA RT411381086GA RT411381090GA RT411381092GA RT411381095GA RT411381096GA RT411381097GA RT411381104GA RT411381105GA RT411381108GA RT411381109GA RT411381118GA RT411381119GA RT411381120GA RT411381121GA RT411381122GA RT411381123GA RT411381124GA RT411381125GA RT411381126GA RT411381127GA RT411381128GA RT411381129GA RT411381130GA RT411381132GA RT411381133GA RT411381134GA RT411381135GA RT411381136GA RT411381137GA RT411381138GA RT411381139GA RT411381140GA RT411381141GA RT411391142GA RT411391143GA RT411391144GA RT411391146GA RT411391152GA SBB11140626GA SBB11170670GA SBB11170753GA RT410149277GA RT411090301GA RT411110356GA RT411110357GA RT411110363GA RT411110365GA RT411110366GA RT411110369GA RT411110370GA RT411140402GA RT411140407GA RT411140408GA RT411140410GA RT411140411GA RT411140412GA RT411140416GA RT411140417GA RT411140419GA RT411140422GA RT411170431GA RT411170435GA RT411170439GA RT411170440GA RT411170449GA RT411170452GA RT411170454GA RT411170456GA RT411170459GA RT411180470GA RT411180473GA RT411180474GA RT411180476GA RT411210507GA RT411210513GA RT411210515GA RT411210519GA RT411210522GA RT411210523GA RT411210525GA RT411210528GA RT411210533GA RT411210536GA RT411210538GA RT411220549GA RT411220553GA RT411220555GA RT411220570GA RT411220576GA RT411220581GA RT411220582GA RT411220585GA RT411240592GA RT411240596GA RT411240597GA RT411240600GA RT411240602GA RT411240605GA RT411240609GA RT411240612GA RT411240613GA RT411240642GA RT411240646GA RT411240647GA RT411240648GA RT411240650GA RT411240653GA RT411240656GA RT411240657GA RT411240661GA RT411240663GA RT411240666GA RT411240667GA RT411250692GA RT411250693GA RT411250695GA RT411250696GA RT411260699GA RT411260700GA RT411260701GA RT411260704GA RT411260705GA RT411260706GA RT411260707GA RT411260708GA RT411260710GA RT411260712GA RT411260713GA RT411260714GA RT411260715GA RT411260717GA RT411260718GA RT411260719GA RT411260720GA RT411260721GA RT411260722GA RT411260726GA RT411260727GA SBB10410107GA SBB11170687GA RT411290881GA RT411300892GA RT411300896GA SBB11050563GA RT411290867GA RT411290867GA RT411290869GA RT411290869GA SBB11140607GA SBB11140608GA SBB11140614GA SBB11140630GA RT411300945GA RT411300951GA SBB11050512GA SBB11060504GA SBB11140617GA SBB11060501GA SBB11140604GA RT411240659GA SBB10299660GA SBB11170732GA SBB11140623GA SBB11140646GA SBB11170744GA SBB11170691GA RT411180486GA RT411330982GA RT411330988GA SBB11170737GA RTW11346581GA SBB11110577GA SBB11140606GA SBB11140613GA SBB11140633GA SBB11170681GA SBB11170683GA SBB11170747GA SBB11170751GA SBB11170735GA SBB11170748GA RTW11346593GA SBB10410118GA SBB11170739GA SBB11170692GA SBB11170701GA SBB11170749GA RT410149288GA RT411090297GA RT411210510GA RT411250690GA RT411250691GA RT411250694GA RT411250697GA RT411260702GA SBB11170745GA SBB11170759GA SBB11030399GA RT411110338GA RT411110340GA RT411110385GA SBB11140644GA SBB11140649GA SBB11170760GA SBB11170752GA SBB11140640GA SBB11140635GA SBB11170661GA SBB11170663GA SBB11170740GA SBB11050561GA RT411110333GA RT411110360GA RT411110371GA RT411110380GA RT411110384GA RT411140397GA RT411140398GA RT411140400GA RT411140401GA RT411140403GA RT411140404GA RT411140405GA RT411140409GA RT411140413GA RT411140414GA RT411140420GA RT411140423GA RT411140424GA RT411140426GA RT411140427GA RT411170430GA RT411170434GA RT411170437GA RT411170441GA RT411170445GA RT411170446GA RT411170453GA RT411170455GA RT411170458GA RT411210527GA RT411210532GA SBB11170750GA RT411180464GA SBB11170659GA RT411110334GA RT411110335GA RT411170442GA SBB11170658GA SBB11170731GA SBB10510316GA SBB11030383GA SBB11040410GA SBB11050435GA SBB11110573GA SBB11110575GA SBB11110592GA SBB11110593GA SBB11110599GA SBB11140627GA SBB11140641GA SBB11140647GA SBB11170655GA SBB11170667GA SBB11170678GA SBB11170679GA RT410149301GAMKE RT411110372GA RT411110376GA RT411110378GA RT411180461GA RT411180462GA RT411180471GA RT411180472GA RT411180481GA RT411180482GA RT411180487GAMKE RT411210502GA RT411210512GA RT411220552GA RTW11286170GAMKE RTW11286204GA RTW11316236GA SBB11170698GA RT411180467GA RT411180475GA RT411110342GA RT411110350GA RT411110351GA RT411140394GA RT411140399GA RT411140425GA SBB11110580GA SBB11110583GA RT411140428GA SBB11040406GA SBB11110571GA SBB11140652GA RT410149280GA RTW11245941GA RTW11245942GA RT411140395GA RT411140418GA SBB11140651GA SBB10309763GA SBB11140615GA RTW11205694GA SBB11110574GA RT410149303GA RT411090290GA RT411100313GA RT411100315GA RT411100316GA RT411100323GA RT411100326GA RT411140406GA RT411140415GA RT411140421GA SBB11110582GA RT411080274GA RT411080281GA RT411090300GA RT411110377GA SBB11040423GA SBB11170662GA SBB11170668GA SBB11170673GA SBB11170676GA RT411080267GA RT411080272GA RT411090285GA RT411090295GA RT411090296GA RT411110375GA SBB11050430GA SBB11050449GA RT411090287GA SBB11050429GA SBB11110594GA SBB11170700GA RT411090288GA RT411090292GA RT411090309GA RT411170436GA SBB11050428GA SBB11050436GA SBB11050453GA SBB11110581GA SBB11110584GA SBB11060466GA SBB11060468GA SBB11060477GA SBB11060478GA SBB11060492GA SBB11060496GA RT411080279GA RT411090283GA RT411090286GA RT411090291GA RT411090294GA RT411090307GA SBB10299655GA SBB10299670GA SBB10299690GA SBB11050522GA SBB11060467GA SBB11060470GA SBB11060472GA SBB11060473GA SBB11050442GA SBB11060479GA SBB11050425GA SBB11110587GA SBB11110588GA SBB11140637GA SBB11170664GA SBB11110589GA RT411110355GA RT411090289GA RT411090299GA RT411090302GA RT411090303GA RT411090306GA RT411090308GA SBB11040419GA RT411110336GA SBB11140603GA SBB11140619GA SBB11140629GA SBB10329825GA SBB10530339GA RT411110353GA SBB11030362GA SBB10229583GA SBB10229589GA SBB10239603GA SBB10239610GA SBB10239612GA SBB10239616GA RTW11165633GA RT411090284GA RT411100327GA RT411110337GA RT411110343GA RT411110347GA RT411110362GA SBB10349901GA SBB11030374GA RT411080259GA RT411100311GA RT411100314GA RT411100317GA RT411100324GA RT411100332GA RTW11155530GA RT411080266GA RT411100322GA RT411100325GA RT411100331GA RT411110348GA RT411110349GA SBB10430200GA RT411090298GA RT411080257GA RT411080261GA RT411080268GA SBB10530322GA SBB10530329GA SBB10530355GA SBB11050510GA SBB11050523GA SBB10530337GA SBB10530346GA RTW11155511GA RTW11155526GA RTW11155527GA RTW11155550GA SBB10530342GA SBB11030376GA SBB10530338GA RT411100318GA SBB10510304GA SBB10510317GA SBB11030358GA SBB10530336GA SBB10530341GA SBB10530350GA RT411100320GA RT411100329GA RTW10465083GA RTW10465130GA RTW11045381GA RTW11045410GA RTW11055415GA RTW11055416GA RTW11055423GA RTW11125481GA SBB11030372GA SBB11030385GA SBB11030387GA SBB10350018GA SBB11060503GA RT410289453GA RT410289457GA RT410329546GA RT410329547GA RT410329548GA RT410329549GA RT410329551GA RT410329552GA RT410329553GA RT410329554GA RT410329555GA RT410329556GA RT410329557GA RT410329558GA RT410329559GA RT410329560GA RT410329561GA RT410329563GA RT410329564GA RT410379567GA RT410409579GA RT410409581GA RT410409586GA RT410419590GA RT410419591GA RT410419592GA RT410419593GA RT410419595GA RT410419596GA RT410419597GA RT410419598GA RT410419599GA RT410419601GA RT410419603GA RT410419604GA RT410419605GA RT410419606GA RT410419607GA RT410419608GA RT410419609GA RT410419610GA RT410419611GA RT410419612GA RT410419613GA RT410419614GA RT410419615GA RT410419616GA RT410429618GA RT410429622GA RT410429623GA RT410429625GA RT410429627GA RT410429629GA RT410429630GA RT410429634GA RT410429636GA RT410439668GA RT410439669GA RT410439672GA RT410439673GA RT410439674GA RT410439675GA RT410439676GA RT410439678GA RT410439680GA RT410439682GA RT410439684GA RT410439685GA RT410439686GA RT410439689GA RT410439690GA RT410439724GA RT410439726GA RT410439729GA RT410439730GA RT410439734GA RT410439737GA RT410439741GA RT410439742GA RT410439745GA RT410439746GA RT410439747GA RT410439748GA RT410439749GA RT410439750GA RT410439751GA RT410439754GA RT410439755GA RT410439757GA RT410439761GA RT410439763GA RT410439764GA RT410439766GA RT410439770GA RT410439780GA RT410439781GA RT410449783GA RT410449785GA RT410449786GA RT410449789GA RT410449913GA RT410449914GA RT410489915GA RT410489916GA RT410489917GA RT410489918GA RT410489919GA RT410489921GA RT410489927GA RT410489935GA RT410489953GA RT410489957GA RT410489959GA RT410489962GA RT410489964GA RT410489967GA RT410489970GA RT410489972GA RT411080260GA SBB11050450GA SBB11050517GA SBB11060494GA RT411110367GA RT411110379GA RT411110386GA RT411110387GA RT411110388GA RT411110389GA RT411110392GA RTW10415065GA SBB10530343GA SBB11030393GA SBB11030398GA SBB11030401GA RT410490044GA RT410500124GA RT411080228GA RT411080278GA RT411060151GA SBB10400038GA SBB10440266GA SBB10440279GA SBB10440289GA SBB11050452GA SBB11050515GA SBB11050519GA SBB11050570GA SBB11060500GA RT411100328GA SBB10440282GA SBB10440283GA SBB10440284GA SBB10440288GA SBB10440290GA SBB10440301GA RTW10475163GA RTW11045379GA SBB11050521GA SBB11060469GA SBB11060493GA RT410490066GA SBB11050460GA RT411110345GA RTW10495214GA SBB11050514GA SBB11050565GA SBB11050564GA RT411080233GA SBB11050527GA SBB11030394GA SBB11030395GA SBB11030400GA SBB11050441GA SBB11050446GA SBB11050454GA SBB11050456GA SBB10530326GA SBB10530327GA SBB10530345GA SBB10530348GA SBB10530349GA SBB11030363GA SBB11040417GA SBB11050437GA SBB11050439GA SBB11050457GA RT410449792GA RT410489981GA RT410500140GA RT411060143GA RT411060149GA RT411060161GA RT411070185GA RT411070186GA RT411070191GA RT411070193GA RT411070196GA RT411070198GA RT411070201GA RT411070202GA RT411080205GA RT411080210GA RT411080211GA RT411080213GA RT411080214GA RT411080215GA RT411080216GA RT411080217GA RT411080218GA RT411080219GA RT411080220GA RT411080221GA RT411080222GA RT411080223GA RT411080224GA RT411080225GA RT411080226GA RT411080227GA RT411080229GA RT411080230GA RT411080231GA RT411080232GA RT411080234GA RT411080235GA RT411080236GA RT411080237GA RT411080238GA RT411080239GA RT411080240GA RT411080241GA RT411080242GA RT411080243GA RT411080244GA RT411080245GA RT411080246GA RT411080247GA RT411080248GA RT411080249GA RT411080251GA RT411080252GA RT411080253GA RT411080254GA RT411080256GA RT411080258GA RT411080264GA RT411080269GA RT411080270GA RT411080271GA RT411080273GA RT411080275GA RT411080276GA RT411080277GA RT411080280GA RT410490050GA RT410490058GA RT411080263GA RT411070200GA RT410429619GA RT410489934GA RT411060157GA RT411060159GA RT411060164GA RT411070190GA SBB11050451GA RT410490070GA SBB10319779GA SBB10400064GA SBB10530334GA RT410490045GA RTW11035356GA RTW11045369GA RTW11045387GA RTW11045397GA RT410490057GA RT410500118GA RT410500119GA RT410500120GA RT410500123GA RT410500125GA RT410500127GA RT410500128GA RT410500129GA RT410500130GA RT410500131GA RT410500132GA RT410500133GA RT410500135GA RT410500141GA RT410500142GA SBB10510303GA SBB10510306GA SBB10510315GA RTW11045362GA SBB10510320GA RT410489984GA RTW11045372GA SBB10299663GA SBB10329829GA SBB10329845GA SBB10430235GA SBB10440242GA SBB10440254GA RTW11035335GA RTW11045365GA RTW11045374GA RTW11045371GA RT410490053GA RT410490054GA RT410490055GA RT410490061GA RT410490065GA RT410490077GA RT410490079GA RT410490081GA SBB10440269GA RT410409580GA RT410439667GA RT410439670GA RT410439671GA RT410439677GA RT410439679GA RT410439681GA RT410439683GA RT410439687GA RT410439688GA RTW10475168GA RTW11035322GA RTW11035337GA RTW11035338GA RTW11035339GA RTW11035345GA RTW11035346GA RTW11035350GA RTW11035353GA RTW11035355GA SBB10299656GA SBB10349913GA RT410490074GA SBB10259634GA SBB10309749GA SBB10319778GA SBB10299635GA RTW10495307GA SBB10319784GA RT410480007GA SBB10400075GA SBB10430180GA SBB10430199GA SBB10440286GA SBB10440287GA RT410259423GA RT410259425GA RT410289460GA RT410309514GA RT410490056GA SBB10219543GA SBB10430174GA SBB10430186GA RT410289481GA RT410289483GA RT410289486GA RT410329525GA RT410329542GA RTW10495207GA RTW10495286GA SBB10430213GA RT410419594GA RT410429620GA RT410449790GA RT410449793GA RT410449794GA RT410449798GA RT410449908GA RT410449909GA SBB10430220GA SBB10440267GA SBB10440277GA SBB10440294GA RT410490032GA RT410490035GA RT410490037GA RT410490038GA RT410490039GA RT410490049GA RTW10475149GA RTW10495295GA SBB10430192GA RT410490083GA SBB10299667GA SBB10430183GA SBB10430190GA SBB10430210GA SBB10430211GA RT410490034GA RT410490042GA SBB10299645GA SBB10430148GA SBB10430172GA SBB10430176GA SBB10430189GA SBB10400073GA SBB10400079GA SBB10400082GA SBB10430197GA SBB10430218GA SBB10430219GA SBB10430232GA SBB10430233GA SBB10430234GA SBB10430236GA SBB10430237GA SBB10430240GA SBB10440241GA SBB10440243GA SBB10440244GA SBB10440246GA SBB10440247GA SBB10440248GA SBB10440249GA SBB10440250GA SBB10440255GA SBB10440258GA SBB10440259GA SBB10440262GA SBB10440268GA SBB10440274GA SBB10440285GA SBB10440292GA SBB10440297GA SBB10440300GA RT410480011GA RT410489931GA RT410489974GA RT410489985GA SBB10309722GA SBB10309730GA SBB10309734GA SBB10309754GA SBB10430191GA SBB10430193GA SBB10430196GA SBB10430203GA SBB10430209GA SBB10430195GA SBB10430198GA SBB10430208GA SBB10400080GA SBB10410119GA SBB10350008GA SBB10400037GA SBB10410099GA SBB10410109GA SBB10410132GA SBB10410137GA SBB10430157GA SBB10430179GA SBB10430182GA SBB10430187GA SBB10430202GA SBB10329833GA SBB10349905GA SBB10410125GA SBB10410126GA SBB10420138GA RT410189351GA RT410259424GA RT410490046GA RTW10465131GA SBB10078937GA SBB10078939GA SBB10149402GA SBB10239608GA SBB10309714GA SBB10400033GA SBB10400036GA SBB10400039GA SBB10400040GA SBB10400046GA SBB10400048GA SBB10400050GA SBB10410097GA SBB10410120GA SBB10410121GA SBB10410128GA RT410229385GA RT410439732GA RT410439735GA RT410490030GA RT410490036GA SBB10329841GA SBB10350003GA RTW10495307GA RT410329532GA RT410429621GA RTW10475146GA SBB10309759GA SBB10319769GA SBB10319770GA SBB10319772GA SBB10319776GA SBB10319798GA SBB10440299GA RTW10475204GA RT410429635GA RT410439727GA RT410439740GA RT410449802GA RT410449804GA RT410449905GA RT410449906GA RT410449912GA SBB10319800GA SBB10319808GA SBB10319814GA RTW10415050GA RTW10415055GA RTW10415078GA SBB10430144GA SBB10430205GA SBB10440252GA SBB10440261GA RTW10465135GA RT406294095GR RT407035566GR SBB09298407GR SBB09318435GR SBB09318440GR SBB09318449GR SBB09378569GR SBB09398588GR SBB09498716GR SBB09508790GR SBB09508791GR SBB09508798GR SBB09528812GR SBB09528831GR SBB09528842GR SBB10058848GR SBB10129071GR SBB10510304GR SBB11030358GR SBB09508800GR SBB10129353GR SBB10430180GR RT412091504GA RT412091508GA SBB12111304GA SBB12121309GA RT406103579GR RT406083441GR RT406103591GR SBB10129145GR SBB11050514GR RT406294150GR RT410129247GR RT411391200GR RT407035564GR RT407075620GR RT410189356GR RT410309507GR RT410490047GR RT410490070GR RT411100320GR RT411100329GR RT410490022GR RT411240664GA RT410329534GR RT409318890GR RT406294180GR RT407367164GR RT408378248GR RT409048740GR RT406424630GR RT407035561GR RT407035570GR RT407065598GR RT408257945GR RT408257973GR RT408468575GR RT408478634GR SBB06513530GR RT407065612GR RT410419602GA RT410429628GA RT410439691GA RT410439758GA RT410439762GA RT410439771GA RT410449787GA RT410449907GA RT410489922GA RT410489923GA RT410489925GA RT410489926GA RT410489929GA RT410489954GA RT410489969GA RT410489971GA RT410489973GA RT410489975GA RT410489977GA RT410489978GA RT410489992GA RT410490026GA RT410490031GA RT410490033GA RT410490051GA RT410490072GA RT410490075GA RT410500121GA RT410500122GA RT410500126GA RT410500136GA RT411060144GA RT411060145GA RT411060146GA RT411060147GA RT411060148GA RT411060150GA RT411060152GA RT411060153GA RT411060154GA RT411060155GA RT411060156GA RT411060158GA RT411060160GA RT411060162GA RT411060163GA RT411060165GA RT411060166GA RT411060167GA RT411060168GA RT411060169GA RT411060170GA RT411060171GA RT411060172GA RT411060173GA RT411060174GA RT411060175GA RT411060176GA RT411060177GA RT411070178GA RT411070179GA RT411070180GA RT411070181GA RT411070182GA RT411070183GA RT411070184GA RT411070187GA RT411070189GA RT411070197GA RT411070199GA RT411080203GA RT411080204GA RT411080206GA RT411080207GA RT411080208GA RT411080209GA RT411080212GA SBB11060498GA SBB07114365GR SBB10420140GA SBB10430150GA SBB10440293GA RT407055590GR SBB06523708GR SBB08467797GR SBB10410131GAMKE SBB10430181GA RT410449801GA RT410449803GA RTW10475164GA RTW10475179GA SBB10410105GA RTW10415038GA RTW10415058GA RTW10415059GA RTW10415060GA RTW10415061GA RTW10415062GA RTW10415063GA RTW10415064GA RTW10415066GA RTW10415067GA RTW10415068GA RTW10415069GA RTW10465079GA RTW10465080GA RTW10465081GA RTW10465085GA RTW10465086GA RTW10465087GA RTW10465088GA RTW10465089GA RTW10465090GA RTW10465092GA RTW10465093GA RTW10465095GA RTW10465097GA RTW10465098GA RTW10465100GA RTW10465102GA RTW10465104GA RTW10465106GA RTW10465107GA RTW10475173GA RT410439739GA RTW10384771GA RTW10394847GA RTW10394849GA RTW10394857GA RTW10394863GA RTW10394865GA RTW10394866GA RTW10394867GA RTW10394873GA RTW10394875GA RTW10394877GA RTW10394879GA RTW10394880GA RTW10394905GA RTW10404960GA RTW10404963GA RTW10404964GA RTW10404965GA RTW10404966GA RTW10404979GA RTW10404980GA RTW10404985GA RTW10414993GA RTW10414997GA RTW10414998GA RTW10415001GA RTW10415002GA RTW10415004GA RTW10415006GA RTW10415010GA RTW10415023GA RTW10415027GA RTW10415031GA RTW10415034GA RTW10415036GA RTW10415044GA RTW10415045GA RTW10415053GA RTW10415054GA RTW10415077GA RTW10475157GA RTW10475174GA SBB10410106GA SBB10410108GA SBB10410115GA RT410439733GA RT410439743GA RTW10394851GA RTW10394853GA RTW10394855GA RTW10394860GA RTW10394861GA RTW10394870GA RTW10394907GA RTW10404959GA RTW10404962GA RTW10404972GA RTW10404976GA RTW10404977GA RTW10404984GA RTW10404987GA RTW10404988GA RTW10414995GA RTW10415008GA RTW10415026GA RTW10415056GA RTW10415076GA SBB10410113GA SBB10410133GA SBB10309760GA SBB10400045GA SBB10430145GA RTW10415046GA RTW10415073GA SBB10400035GA SBB10400044GA RT410419600GA RT410439728GA RT410439731GA RT410439744GA RT410449910GA SBB10430141GA SBB10430147GA SBB10430159GA SBB10430166GA SBB10440264GA RT410439738GA RT410409585GA RT410409588GA SBB10400034GA SBB10400052GA SBB10430158GA SBB10430161GA SBB10430194GA RT410409582GA RT410409583GA RT410409584GA RT410409587GA RT410429617GA RT410429624GA RT410429626GA RT410429631GA RT410429632GA RT410429633GA RT410439725GA SBB10430152GA RT410449788GA SBB10430163GA SBB10430164GA RTW10415000GA SBB10430143GA SBB10430168GA SBB10430207GA SBB10440245GA SBB10440275GA SBB10440298GA RT406485023GR RT406515425GR RT407065607GR RT410449784GA SBB10259626GA SBB10259629GA SBB10329843GA SBB10410093GA SBB10410096GA SBB10410100GA SBB10410101GAMKE SBB10410102GA SBB10410104GA SBB10410112GA SBB10430155GA SBB10309696GA SBB10309709GA SBB10430167GA SBB10430173GA SBB10430177GA RTW10415003GA RTW10415005GA RTW10415018GA RTW10415021GA RTW10415022GA RTW10415074GA SBB10430160GA SBB10430165GA SBB10430169GA RT410439753GA RTW10394874GA RTW10394903GA RTW10394909GA RTW10394910GA RTW10415048GA RT410439752GA RT410439756GA SBB10430153GA SBB10410129GA SBB10430142GA RTW10384770GA RTW10384772GA RTW10384773GA RTW10384774GA RTW10384775GA RTW10384776GA RTW10384777GA RTW10384778GA RTW10384780GA RTW10384783GA RTW10384784GA RTW10384790GA RTW10384797GA RTW10384799GA RTW10384800GA RTW10384802GA RTW10384803GA RTW10384804GA RTW10384805GA RTW10384806GA RTW10384807GA RTW10384808GA RTW10384809GA RTW10384811GA RTW10394852GA RTW10394858GA RTW10394859GA RTW10394862GA RTW10394864GA RTW10394868GA RTW10394869GA RTW10394871GA RTW10394872GA RTW10394878GA RTW10394881GA RTW10394904GA RTW10394911GA RTW10394913GA RTW10394919GA RTW10394920GA RTW10394921GA RTW10394922GA RTW10394923GA RTW10394929GA RTW10415011GA RTW10415028GA RTW10415029GA RTW10415035GA RTW10415052GA RTW10415057GA SBB10430149GA SBB10430151GA SBB10430156GA SBB10430170GA RTW10394902GA SBB10430171GA SBB10430175GA RTW10394916GA SBB10410130GA SBB10400051GA SBB10410116GA SBB10078925GA SBB10299673GA SBB10299677GA SBB10299683GA SBB10299686GA SBB10299691GA SBB10309697GA SBB10309700GA SBB10309713GA SBB10309718GA SBB10319816GA SBB10350024GA SBB10410122GA SBB10299694GA SBB10349914GA SBB10410095GA SBB10299671GA RTW10394908GA SBB10239621GA SBB10078899GA SBB10149385GA SBB10299680GA SBB10299682GA SBB10309710GA SBB10309715GA SBB10400047GA SBB10309723GA SBB10078864GA SBB10309719GA SBB10309752GA RTW10394848GA SBB10149389GA SBB12131350GA SBB12111285GA SBB12131344GA SBB12131345GA SBB12031032GA SBB12051137GA RT412011400GA RT412011403GA RTW11346556GA RTW11346603GA RTW11346609GA SBB12041086GA SBB12041117GA SBB12041128GA SBB12051133GA SBB12051136GA SBB12051141GA SBB11410827GA RTW11346598GA SBB11410829GA SBB11410839GA SBB11170685GA SBB10440280GA SBB11050458GA RTW11055430GA RTW10224304GA SBB10440278GA RTW11045390GA RTW10164214GA RTW10164213GA SBB10319811GA SBB10329840GA SBB12031053GA SBB12031012GA SBB10430178GA RT411090282GA RT411090293GA SBB12281560GA SBB12131369GA RT412081440GA SBB12031048GA SBB11170665GA SBB11140616GA RTW11155538GA RT410490041GA SBB10430154GA SBB10430184GA SBB10319771GA SBB10319773GA SBB12281530GA SBB10430146GA RT410229383GA RTW11366778GA RTW11396873GA RTW11366698GA RTW11366714GA RTW11366670GA RTW11366675GA RTW11366677GA RTW11366683GA RTW11376811GA RTW11396883GA RT412101667GA RTW11346618GA RTW11346620GA RTW11346662GA RTW11316385GA RTW11346622GA RTW11346666GA RTW11346615GA RTW11376820GA RTW11346643GA RTW11346661GA RTW11346660GA RTW11346624GA RTW11376828GA RTW11376837GA RTW11386842GA RTW11346627GA RTW11346613GA RTW11376823GA RTW11346610GA RTW11346585GA RTW11055426GA RTW11346592GA RTW11235855GA RTW11235904GA RTW11235801GA RTW11256033GA RTW11256034GA RTW11256035GA RTW11256036GA RTW11256037GA RTW11256038GA RTW11256040GA RTW11256041GA RTW11256042GA RTW11256043GA RTW11256044GA RTW11256045GA RTW11256046GA RTW11256047GA RTW11256049GA RTW11256050GA RTW11256052GA RTW11256063GA RTW11256064GA RTW11256065GA RTW11256066GA RTW11256067GA RTW11256068GA RTW11256069GA RTW11256070GA RTW11256071GA RTW11256072GA RTW11326503GA RTW11346578GA RTW11346591GA RTW11245932GA RTW11245963GA RTW11245964GA RTW11245965GA RTW11245966GA RTW11245967GA RTW11245968GA RTW11245969GA RTW11245970GA RTW11245971GA RTW11245972GA RTW11245973GA RTW11245974GA RTW11245975GA RTW11245976GA RTW11245977GA RTW11316294GA RTW11326420GA RTW11326434GA RTW11326493GA RTW11326494GA RTW11326500GA RTW11326501GA RTW11326502GA RTW11336538GA RTW11346546GA RTW11346547GA RTW11346549GA RTW11346550GA RTW11346551GA RTW11346552GA RTW11346553GA RTW11346554GA RTW11346555GA RTW11346557GA RTW11346558GA RTW11346559GA RTW11346560GA RTW11346561GA RTW11346562GA RTW11346563GA RTW11346564GA RTW11346565GA RTW11346566GA RTW11346567GA RTW11346568GA RTW11346569GA RTW11346570GA RTW11346571GA RTW11346572GA RTW11346574GA RT411100310GA RT411100312GA RT411100319GA RT411110354GA RT411110359GA RT411110368GA RT411170429GA RT411170438GA RT411170444GA RT411170447GA RT411170448GA RT411170451GA RT411170457GA RT411180466GA RT411180468GA RT411180469GA RT411180478GA RT411180483GA RT411180490GA RT411180500GA RTW11125454GA RTW11125501GA RTW11286089GA RTW11286130GA RTW11286198GA RTW11316299GA RTW11316308GA RTW11316321GA RTW11316322GA RTW11316327GA RTW11316328GA RTW11316329GA RTW11316330GA RTW11316332GA RTW11316333GA RTW11316334GA RTW11316335GA RTW11316361GA RTW11316362GA RTW11316364GA RTW11316369GA RTW11316379GA RTW11316380GA RTW11316384GA RTW11316387GA RTW11316388GA RTW11316389GA RTW11316390GA RTW11326403GA RTW11326413GA RTW11326415GA RTW11326416GA RTW11326417GA RTW11326424GA RTW11326425GA RTW11326427GA RTW11326430GA RTW11326445GA RTW11326448GA RTW11326449GA RTW11326450GA RTW11326453GA RTW11326455GA RTW11326456GA RTW11326457GA RTW11326460GA RTW11326461GA RTW11326462GA RTW11326468GA RTW11326469GA RTW11326470GA RTW11326471GA RTW11326472GA RTW11326473GA RTW11326476GA RTW11326477GA RTW11326480GA RTW11326482GA RTW11326483GA RTW11326484GA RTW11326485GA RTW11326487GA RTW11326488GA RTW11326489GA RTW11326490GA RTW11326492GA RTW11326495GA RTW11326497GA RTW11326498GA RTW11326499GA RTW11326504GA RTW11326505GA RTW11326506GA RTW11326507GA RTW11326508GA RTW11326509GA RTW11326511GA RTW11326512GA RTW11326513GA RTW11326514GA RTW11326518GA RTW11326519GA RTW11326520GA RTW11326523GA RTW11326524GA RTW11326525GA RTW11326526GA RTW11326527GA RTW11326528GA RTW11326529GA RTW11326531GA RTW11326532GA RTW11326537GA RTW11336539GA RTW11336540GA RTW11336541GA RTW11336542GA RTW11336543GA RTW11336544GA RTW11336545GA RTW11316373GA RTW11316377GA RTW11316378GA RTW11316381GA RTW11316382GA RTW11316386GA RTW11316391GA RTW11316393GA RTW11316394GA RTW11316395GA RTW11326396GA RTW11326397GA RTW11326398GA RTW11326399GA RTW11326400GA RTW11326401GA RTW11326402GA RTW11326404GA RTW11326405GA RTW11326406GA RTW11326407GA RTW11326408GA RTW11326409GA RTW11326410GA RTW11326411GA RTW11326412GA RTW11326414GA RTW11326418GA RTW11326419GA RTW11326421GA RTW11326422GA RTW11326423GA RTW11326428GA RTW11326429GA RTW11326431GA RTW11326433GA RTW11326435GA RTW11326436GA RTW11326437GA RTW11326441GA RTW11326442GA RTW11326443GA RTW11326444GA RTW11326446GA RTW11326447GA RTW11326451GA RTW11326452GA RTW11326454GA RTW11326458GA RTW11326459GA RTW11125502GA RTW11245979GA RTW11286157GA RTW11286160GA RTW11286166GA RTW11316227GA RTW11316228GA RTW11316229GA RTW11316230GA RTW11316231GA RTW11316232GA RTW11316233GA RTW11316234GA RTW11316235GA RTW11316237GA RTW11316238GA RTW11316239GA RTW11316240GA RTW11316241GA RTW11316242GA RTW11316243GA RTW11316244GA RTW11316245GA RTW11316246GA RTW11316247GA RTW11316249GA RTW11316281GA RTW11316282GA RTW11316283GA RTW11316284GA RTW11316285GA RTW11316286GA RTW11316287GA RTW11316288GA RTW11316289GA RTW11316290GA RTW11316291GA RTW11316292GA RTW11316293GA RTW11316295GA RTW11316296GA RTW11316297GA RTW11316298GA RTW11316300GA RTW11316301GA RTW11316302GA RTW11316303GA RTW11316304GA RTW11316305GA RTW11316306GA RTW11316307GA RTW11316309GA RTW11316310GA RTW11316311GA RTW11316312GA RTW11316313GA RTW11316314GA RTW11316315GA RTW11316316GA RTW11316317GA RTW11316318GA RTW11316319GA RTW11316320GA RTW11316323GA RTW11316324GA RTW11316325GA RTW11316326GA RTW11316336GA RTW11316337GA RTW11316338GA RTW11316339GA RTW11316341GA RTW11316342GA RTW11316343GA RTW11316344GA RTW11316346GA RTW11316347GA RTW11316348GA RTW11316349GA RTW11316350GA RTW11316351GA RTW11316352GA RTW11316353GA RTW11316354GA RTW11316355GA RTW11316356GA RTW11316357GA RTW11316358GA RTW11316359GA RTW11316360GA RTW11316365GA RTW11316366GA RTW11316367GA RTW11316368GA RTW11316370GA RTW11316371GA RTW11316372GA RTW11316374GA RTW11316375GA RTW11316376GA RTW11225775GA RTW11225780GA RTW11225792GA RTW11245989GA RTW11256019GA RTW11286087GA RTW11286088GA RTW11286090GA RTW11286091GA RTW11286092GA RTW11286093GA RTW11286094GA RTW11286095GA RTW11286096GA RTW11286097GA RTW11286099GA RTW11286101GA RTW11286102GA RTW11286141GA RTW11286143GA RTW11286144GA RTW11286146GA RTW11286159GA RTW11286168GA RTW11286186GA RTW11286187GA RTW11286196GA RTW11286200GA RTW11286202GA RTW11306207GA RTW11306208GA RTW11306209GA RTW11306210GA RTW11306211GA RTW11306212GA RTW11306213GA RTW11306214GA RTW11306215GA RTW11306216GA RTW11306217GA RTW11306218GA RTW11306219GA RTW11306220GA RTW11306221GA RTW11306222GA RTW11306223GA RTW11306224GA RTW11306225GA RTW11306226GA RTW11316248GA RTW11256008GA RTW11256015GA RTW11256073GA RTW11256074GA RTW11256075GA RTW11256077GA RTW11256079GA RTW11256081GA RTW11256085GA RTW11256086GA RTW11286106GA RTW11286128GA RTW11286129GA RTW11286133GA RTW11286134GA RTW11286138GA RTW11286147GA RTW11286148GA RTW11286149GA RTW11286150GA RTW11286151GA RTW11286152GA RTW11286153GA RTW11286155GA RTW11286156GA RTW11286158GA RTW11286161GA RTW11286162GA RTW11286163GA RTW11286169GA RTW11286171GA RTW11286172GA RTW11286173GA RTW11286174GA RTW11286175GA RTW11286176GA RTW11286177GA RTW11286178GA RTW11286179GA RTW11286180GA RTW11286181GA RTW11286182GA RTW11286183GA RTW11286184GA RTW11286185GA RTW11286189GA RTW11286190GA RTW11286194GA RTW11286195GA RTW11286199GA RTW11245936GA RTW11256001GA RTW11256002GA RTW11256005GA RTW11256007GA RTW11256010GA RTW11256013GA RTW11256014GA RTW11256016GA RTW11256017GA RTW11256018GA RTW11256020GA RTW11256021GA RTW11256022GA RTW11256023GA RTW11256024GA RTW11256025GA RTW11256026GA RTW11256027GA RTW11256028GA RTW11256029GA RTW11256030GA RTW11256031GA RTW11256032GA RTW11256076GA RTW11256078GA RTW11256080GA RTW11256082GA RTW11256083GA RTW11256084GA RTW11286098GA RTW11286100GA RTW11286105GA RTW11286127GA RTW11286131GA RTW11286132GA RTW11286135GA RTW11286136GA RTW11286137GA RTW11286139GA RTW11286142GA RTW11286188GA RTW11286191GA RTW11286192GA RTW11286193GA RTW11286197GA RTW11286201GA RTW11286203GA RTW11286205GA RTW11286206GA RTW11235850GA RTW11235895GA RTW11235903GA RTW11235916GA RTW11245920GA RTW11245921GA RTW11245922GA RTW11245926GA RTW11245931GA RTW11245933GA RTW11245938GA RTW11245944GA RTW11245945GA RTW11245946GA RTW11245947GA RTW11245948GA RTW11245949GA RTW11245950GA RTW11245951GA RTW11245952GA RTW11245953GA RTW11245954GA RTW11245955GA RTW11245956GA RTW11245958GA RTW11245960GA RTW11245962GA RTW11245978GA RTW11245980GA RTW11245981GA RTW11245982GA RTW11245983GA RTW11245984GA RTW11245986GA RTW11245987GA RTW11245988GA RTW11245990GA RTW11245991GA RTW11245992GA RTW11255993GA RTW11255994GA RTW11255995GA RTW11255996GA RTW11255998GA RTW11255999GA RTW11256000GA RTW11256003GA RTW11256009GA RTW11256011GA RTW11256012GA RTW11215750GA RTW10224261GA RTW10224296GA RTW10274512GA RTW11125478GA RTW11175635GA RTW11215747GA RTW11215748GA RTW11225760GA RTW11235796GA RTW11235797GA RTW11235798GA RTW11235799GA RTW11235837GA RTW11235838GA RTW11235839GA RTW11235840GA RTW11235841GA RTW11235842GA RTW11235843GA RTW11235844GA RTW11235845GA RTW11235846GA RTW11235847GA RTW11235848GA RTW11235849GA RTW11235852GA RTW11235853GA RTW11235854GA RTW11235856GA RTW11235857GA RTW11235858GA RTW11235860GA RTW11235861GA RTW11235862GA RTW11235863GA RTW11235864GA RTW11235865GA RTW11235866GA RTW11235867GA RTW11235868GA RTW11235869GA RTW11235870GA RTW11235871GA RTW11235873GA RTW11235874GA RTW11235875GA RTW11235876GA RTW11235877GA RTW11235878GA RTW11235879GA RTW11235880GA RTW11235881GA RTW11235882GA RTW11235883GA RTW11235884GA RTW11235885GA RTW11235886GA RTW11235887GA RTW11235888GA RTW11235889GA RTW11235890GA RTW11235891GA RTW11235892GA RTW11235893GA RTW11235894GA RTW11235896GA RTW11235897GA RTW11235898GA RTW11235899GA RTW11235900GA RTW11235901GA RTW11235902GA RTW11235905GA RTW11235906GA RTW11235907GA RTW11235908GA RTW11235909GA RTW11235910GA RTW11235911GA RTW11235912GA RTW11235913GA RTW11235914GA RTW11235915GA RTW11235917GA RTW11235918GA RTW11245919GA RTW11245923GA RTW11245924GA RTW11245925GA RTW11245927GA RTW11245928GA RTW11245929GA RTW11245930GA RTW11245934GA RTW11245935GA RTW11245937GA RTW11245939GA RTW11245940GA RTW11245943GA RTW11245957GA RTW11245959GA RTW11245961GA RTW10414996GA RTW11035328GA RTW10475160GA RTW11155506GA RTW11155523GA RTW11155563GA RTW11155565GA RTW11155566GA RTW11155567GA RTW11155579GA RTW11155581GA RTW11155583GA RTW11155584GA RTW11155585GA RTW11155587GA RTW11155588GA RTW11155589GA RTW11155590GA RTW11155591GA RTW11155592GA RTW11155593GA RTW11155594GA RTW11155597GA RTW11155599GA RTW11155600GA RTW11155604GA RTW11155605GA RTW11155606GA RTW11155607GA RTW11155608GA RTW11165614GA RTW11165620GA RTW11165624GA RTW11165631GA RTW11175644GA RTW11175647GA RTW11175651GA RTW11175657GA RTW11175658GA RTW11205661GA RTW11205662GA RTW11205663GA RTW11205666GA RTW11205669GA RTW11205671GA RTW11205683GA RTW11205685GA RTW11205686GA RTW11205690GA RTW11205691GA RTW11205692GA RTW11205693GA RTW11205695GA RTW11205697GA RTW11205698GA RTW11205701GA RTW11205703GA RTW11205704GA RTW11205705GA RTW11205708GA RTW11205710GA RTW11205711GA RTW11205713GA RTW11205715GA RTW11205717GA RTW11215718GA RTW11215724GA RTW11215727GA RTW11215728GA RTW11215729GA RTW11215730GA RTW11215733GA RTW11215734GA RTW11215735GA RTW11215738GA RTW11215740GA RTW11215741GA RTW11215742GA RTW11215743GA RTW11215746GA RTW11215752GA RTW11215753GA RTW11215754GA RTW11215755GA RTW11215756GA RTW11215757GA RTW11225778GA RTW11225779GA RTW11225781GA RTW11225784GA RTW11225786GA RTW11225788GA RTW11225789GA RTW11235794GA RTW11235795GA RTW11235800GA RTW11235806GA RTW11235807GA RTW11235808GA RTW11235811GA RTW11235833GA RTW11235851GA RTW11165621GA RTW11155531GA RTW11155532GA RTW11155541GA RTW11155543GA RTW11155546GA RTW11155551GA RTW11155564GA RTW11155573GA RTW11155575GA RTW11155580GA RTW11155582GA RTW11155586GA RTW11155595GA RTW11155596GA RTW11155598GA RTW11155601GA RTW11155602GA RTW11165632GA RTW11175636GA RTW11175638GA RTW11175639GA RTW11175640GA RTW11175641GA RTW11175643GA RTW11175645GA RTW11175646GA RTW11175648GA RTW11175650GA RTW11175656GA RTW11205660GA RTW11205664GA RTW11205665GA RTW11205667GA RTW11205668GA RTW11205670GA RTW11205672GA RTW11205673GA RTW11205674GA RTW11205675GA RTW11205676GA RTW11205677GA RTW11205678GA RTW11205679GA RTW11205680GA RTW11205681GA RTW11205682GA RTW11205684GA RTW11205687GA RTW11205688GA RTW11205689GA RTW11205696GA RTW11205699GA RTW11205700GA RTW11205702GA RTW11205706GA RTW11205707GA RTW11205709GA RTW11205714GA RTW11205716GA RTW11215719GA RTW11215720GA RTW11215723GA RTW11215725GA RTW11215726GA RTW11215731GA RTW11215732GA RTW11215737GA RTW11215739GA RTW11215745GA RTW11215749GA RTW11215751GA RTW11225758GA RTW11225759GA RTW11225761GA RTW11225762GA RTW11225763GA RTW11225764GA RTW11225765GA RTW11225766GA RTW11225767GA RTW11225768GA RTW11225769GA RTW11225770GA RTW11225771GA RTW11225772GA RTW11225773GA RTW11225776GA RTW11225777GA RTW11225782GA RTW11225787GA RTW11225790GA RTW11235803GA RTW11235804GA RTW11235805GA RTW11235809GA RTW11235810GA RTW11235812GA RTW11235834GA RTW11235835GA RTW11235836GA RTW11125456GA RTW11155544GA RTW11165611GA RTW11165617GA RTW11165618GA RTW11175637GA RTW11175642GA RTW11175649GA RTW11175652GA RTW11175653GA RTW11175654GA RTW11175655GA RTW11175659GA RTW11165609GA RTW11165610GA RTW11165612GA RTW11165616GA RTW11165619GA RTW11165622GA RTW11165623GA RTW11165625GA RTW11165626GA RTW11165629GA RTW11165634GA RTW11165615GA RTW11155529GA RT411110383GA RTW10465103GA RTW10465109GA RTW10465120GA RTW10475182GA RTW10475188GA RTW10495215GA RTW11045370GA RTW11045376GA RTW11045409GA RTW11045413GA RTW11055417GA RTW11055421GA RTW11055425GA RTW11055427GA RTW11055429GA RTW11055433GA RTW11055436GA RTW11115441GA RTW11115442GA RTW11115443GA RTW11115445GA RTW11115446GA RTW11115447GA RTW11115448GA RTW11115449GA RTW11115450GA RTW11115451GA RTW11115452GA RTW11125453GA RTW11125455GA RTW11125457GA RTW11125459GA RTW11125461GA RTW11125462GA RTW11125463GA RTW11125464GA RTW11125468GA RTW11125469GA RTW11125471GA RTW11125472GA RTW11125473GA RTW11125474GA RTW11125475GA RTW11125476GA RTW11125477GA RTW11155505GA RTW11155507GA RTW11155508GA RTW11155509GA RTW11155510GA RTW11155513GA RTW11155514GA RTW11155519GA RTW11155520GA RTW11155521GA RTW11155524GA RTW11155534GA RTW11155536GA RTW11155539GA RTW11155540GA RTW11155542GA RTW11155545GA RTW11155549GA RTW11155552GA RTW11155553GA RTW11155554GA RTW11155556GA RTW11155557GA RTW11155559GA RTW11155560GA RTW11155561GA RTW11155562GA RTW11155568GA RTW11155569GA RTW11155570GA RTW11155571GA RTW11155572GA RTW11155574GA RTW11155576GA RTW11155577GA SBB11030379GA SBB11140605GA SBB11140639GA RTW10495210GA RTW10495211GA RTW10495212GA RTW10495221GA RTW10495296GA RTW10495306GA RTW11115444GA RTW11155504GA RTW11155512GA RTW11155515GA RTW11155516GA RTW11155518GA RTW11155522GA RTW11155525GA RTW11155528GA RTW11155533GA RTW11155535GA RTW11155537GA RTW11155548GA RTW11155555GA RTW11155558GA RTW10465111GA RTW10465118GA RTW10465127GA RTW10465128GA RTW10475148GA RTW11055435GA RTW11055440GA RTW10475155GA RTW10475156GA RTW10475159GA RTW10475165GA RTW10475166GA RTW10475167GA RTW10475169GA RTW10475170GA RTW10475171GA RTW10475175GA RTW10475176GA RTW10475177GA RTW10475178GA RTW10475183GA RTW10475186GA RTW10475187GA RTW10475189GA RTW10475190GA RTW10475191GA RTW10475192GA RTW10475193GA RTW10475194GA RTW10475195GA RTW10475196GA RTW10475197GA RTW10475199GA RTW10475200GA RTW10475201GA RTW10475202GA RTW10475203GA RTW11055424GA RT410489928GA RT410490071GA RT410490085GA RT410500134GA RTW10465122GA RTW10495213GA RTW10415017GA RTW10394906GA RTW10394912GA RTW10394914GA RTW10394915GA RTW10394918GA RTW10404961GA RTW10404967GA RTW10404968GA RTW10404969GA RTW10404970GA RTW10404971GA RTW10404973GA RTW10404975GA RTW10404978GA RTW10404981GA RTW10404982GA RTW10404983GA RTW10404986GA RTW10404989GA RTW10414994GA RTW10414999GA RTW10415019GA RTW10415032GA RTW10415037GA RTW10415039GA RTW10415040GA RTW10415041GA RTW10415043GA RTW10415051GA RTW10415070GA RTW10415071GA RTW10415072GA RTW10415075GA RTW10465082GA RTW10465084GA RTW10465091GA RTW10465099GA RTW10465101GA RTW10465105GA RTW10465108GA RTW10465110GA RTW10465112GA RTW10465113GA RTW10465115GA RTW10465117GA RTW10465121GA RTW10465124GA RTW10465125GA RTW10465126GA RTW10465129GA RTW10465132GA RTW10465133GA RTW10465137GA RTW10475140GA RTW10475143GA RTW10475144GA RTW10475150GA RTW10475151GA RTW10475153GA RTW10475158GA SBB11410838GA SBB10530335GA SBB10440272GA SBB10400069GA SBB12301619GA SBB12281541GA SBB12281555GA SBB12261518GA SBB12261519GA SBB12261520GA SBB12261517GA SBB12261516GA SBB12261513GA SBB12131364GA RT412081475GA RT412081485GA RT412091500GA RT412101572GA RT412101617GA RT412101618GA RT412101641GA SBB12131395GA SBB12131343GA SBB12031060GA SBB12131339GA SBB12131367GA SBB12131380GA SBB12131336GA SBB12051138GA SBB12071237GA SBB12061192GA SBB12061193GA SBB12061194GA SBB12061199GA SBB12061211GA SBB12061212GA SBB12061215GA SBB12061216GA SBB12061217GA SBB12061218GA SBB12061219GA SBB12061220GA SBB12061221GA SBB12071223GA SBB12071224GA SBB12071225GA SBB12071226GA SBB12071230GA SBB12071238GA SBB12071239GA SBB12071242GA SBB12071243GA SBB12071253GA SBB12071257GA SBB12071258GA SBB12071260GA SBB12071261GA SBB12071262GA SBB12071264GA SBB12071265GA SBB12071267GA SBB12071268GA SBB12071269GA SBB12071270GA SBB12071272GA SBB12071273GA SBB12071274GA SBB12071275GA SBB12071276GA SBB12111283GA SBB12111284GA SBB12111286GA SBB12111288GA SBB12111289GA SBB12111290GA SBB12111291GA SBB12111292GA SBB12111293GA SBB12111294GA SBB12111296GA SBB12111297GA SBB12111299GA SBB12111300GA SBB12111301GA SBB12111302GA SBB12121323GA SBB12121324GA SBB12121325GA SBB12121326GA SBB12121327GA SBB12121328GA SBB12121329GA SBB12121330GA SBB12121332GA SBB12131333GA SBB12131334GA SBB12131335GA SBB12131337GA SBB12031033GA SBB12071233GA SBB12041126GA SBB12041123GA SBB12041067GA SBB12041091GA SBB12051140GA SBB12051142GA SBB12031059GA SBB12031036GA SBB12031014GA SBB12031052GA SBB12031046GA SBB12031019GA RT412011388GA RT412011397GA RT412021411GA RT412021421GA SBB12021007GA SBB12031043GA SBB12021008GA SBB12021010GA SBB11410807GA SBB12010983GA SBB11170669GA SBB11520961GA SBB12010963GA SBB12010964GA SBB12010966GA SBB12010967GA SBB12010968GA SBB12010969GA SBB12010970GA SBB12010971GA SBB12010972GA SBB12010973GA SBB12010974GA SBB12010975GA SBB12010976GA SBB12010977GA SBB12010978GA SBB12010979GA SBB12010980GA SBB12010981GA SBB11510951GA SBB11510954GA SBB11440858GA SBB11490887GA SBB11490888GA SBB11490889GA SBB11490890GA SBB11490891GA SBB11490892GA SBB11490885GA RT411411247GA RT411411255GA SBB11490882GA SBB11490883GA SBB11410837GA SBB11410821GA RT411411276GA SBB11050448GA SBB11440860GA SBB10410114GA SBB11440856GA SBB11440861GA SBB11440864GA SBB11440866GA SBB11410826GA RT411391147GA RT411391155GA RT411391158GA RT411391159GA RT411391161GA RT411391164GA RT411391165GA RT411391168GA RT411391174GA RT411391176GA RT411391177GA RT411391178GA RT411391180GA RT411391181GA RT411391182GA RT411391183GA RT411391184GA RT411391188GA RT411391191GA RT411391192GA RT411391193GA RT411391194GA RT411391195GA RT411391196GA RT411391201GA RT411401203GA RT411401205GA RT411401209GA RT411401219GA RT411401221GA RT411401226GA RT411401227GA RT411401228GA RT411401229GA RT411401230GA RT411401232GA RT411401234GA RT411401237GA RT411401245GA RT411401246GA SBB11410851GA SBB11410852GA SBB11410832GA SBB11410848GA SBB11410836GA SBB11410842GA SBB11410840GA SBB11410806GA SBB11410850GA SBB11410810GA SBB11410815GA SBB11410819GA SBB11410834GA SBB11410820GA RT411381110GA RT411381131GA RT411391163GA RT411391169GA RT411391171GA SBB11400798GA SBB11410811GA SBB11410812GA SBB11410824GA RT411391151GA RT411391156GA RT411391167GA RT411391172GA RT411391179GA RT411391186GA RT411391187GA RT411391190GA SBB11400801GA SBB11170741GA RT411391150GA RT411391153GA RT411391154GA SBB11140634GA RT411260723GA RT411260724GA RT411260725GA SBB11170756GA SBB11170697GA SBB11170746GA RT411180465GA SBB11140609GA SBB11170704GA RT411110373GA SBB11040415GA SBB11140620GA SBB11170674GA SBB11170702GA SBB11170675GA SBB11110597GA SBB11050426GA SBB11060480GA SBB11110578GA SBB11110598GA SBB11030368GA SBB10530325GA SBB10309716GA SBB10299637GA SBB10530354GA SBB10530344GA SBB11050530GA SBB11050440GA SBB11050518GA RT410139272GA RT410149299GA SBB10510309GA SBB11060502GA SBB10410134GA SBB10410136GA SBB10430188GA SBB10430214GA SBB10430215GA SBB10430217GA SBB10430231GA SBB10430239GA SBB10440257GA SBB10440263GA SBB10440265GA SBB10440276GA SBB10510305GA SBB10510310GA SBB11050431GA SBB11040413GA SBB11040416GA SBB10440296GA SBB10510312GA SBB11040403GA SBB11040409GA SBB11040412GA SBB11040418GA SBB11040422GA SBB10440271GA RT410490076GA SBB10430216GA SBB10440273GA SBB10400084GA SBB10440291GA RT410490073GA SBB10400060GA RT410149290GA SBB11030369GA SBB10530330GA SBB10530340GA SBB10530353GA SBB10530324GA RT410480005GA RT410480008GA RT410480009GA RT410480010GA RT410480012GA RT410480013GA RT410480014GA RT410480015GA RT410480019GA RT410480020GA RT410489955GA RT410489961GA RT410489963GA RT410489968GA RT410489976GA RT410489979GA RT410489983GA RT410489987GA RT410489990GA RT410489993GA RT410490021GA RT410490023GA RT410490025GA RT410490027GA SBB10299642GA SBB10299653GA SBB10299687GA SBB10309698GA SBB10309764GA SBB10319792GA SBB10319794GA SBB10319796GA SBB10319804GA SBB10319805GA SBB10319819GA SBB10319820GA SBB10329821GA SBB10329835GA SBB10349902GA SBB10350016GA SBB10350023GA SBB10350026GA SBB10350031GA SBB10350032GA SBB10400049GA SBB10410094GA SBB10410103GA SBB10410111GA SBB10510302GA SBB10400057GA SBB10400058GA SBB10400059GA SBB10400061GA SBB10400063GA SBB10400065GA SBB10400066GA SBB10400067GA SBB10400068GA SBB10400070GA SBB10400071GA SBB10400076GA SBB10400077GA SBB10400083GA SBB10400085GA SBB10400086GA RT410289459GA RT410490022GA RT410490047GA SBB10430201GA SBB10400062GA SBB10400074GA RT410480018GA RT410489994GA SBB10430238GA RT410489982GA RT410489988GA SBB10309708GA SBB10430162GA SBB10309750GA SBB10319809GA SBB10349894GA SBB10350010GA SBB10410098GA SBB10430212GA SBB10309701GA SBB10319806GA SBB10410110GA SBB10440253GA RT410329550GA RT410379568GA RT410439759GA RT410439760GA RT410439767GA RT410439768GA RT410439769GA RT410439772GA RT410439773GA RT410439774GA RT410439775GA RT410439776GA RT410439777GA RT410439778GA RT410439779GA RT410449782GA RT410449791GA RT410449795GA RT410449796GA RT410449797GA RT410449799GA RT410449800GA RT410449911GA RT410480004GA RT410480006GA RT410480016GA RT410480017GA RT410489920GA RT410489924GA RT410489930GA RT410489932GA RT410489933GA RT410489956GA RT410489958GA RT410489960GA RT410489966GA RT410489980GA RT410489989GA SBB11050566GA SBB12031018GA SBB12020987GA SBB12020989GA RT411411279GA SBB11440863GA SBB11410817GA SBB11170657GA RT411180463GA SBB11050427GA SBB11060474GA SBB11060485GA SBB11060483GA SBB10440281GA SBB10530328GA SBB10309702GA SBB12051153GA SBB12041121GA RT411401207GA SBB11170690GA RTW10465123GA RTW10475162GA RTW10475181GA RTW11035320GA RTW11035325GA RTW11035340GA RTW11035341GA RTW11035342GA RTW11035344GA RTW11035349GA RT410489965GA RT410489991GA RTW10164193GA RTW10465134GA SBB10319817GA SBB10319818GA SBB12131376GA SBB12131347GA SBB12051157GA SBB11170680GA RT412011383GA SBB12010962GA SBB11140636GA SBB11440865GA SBB11170743GA SBB11410849GA SBB11170755GA SBB11140653GA SBB11170689GA SBB11170734GA SBB11140654GA SBB11170680GA SBB11170733GA SBB11060463GA SBB11030380GA SBB11110595GA SBB11140612GA SBB11170666GA SBB11170671GA SBB11060488GA SBB11050455GA SBB11050528GA SBB10078971GA SBB10329831GA SBB10530347GA RT410489986GA SBB10400078GA SBB10400081GA SBB10319807GA RTW10465136GA RTW10475154GA SBB12041130GA SBB12041131GA SBB12031015GA SBB12031050GA SBB12041108GA SBB12031065GA RT412011382GA SBB11490898GA SBB11490906GA SBB11490909GA SBB11490911GA SBB11490912GA SBB11490913GA SBB11490914GA SBB11490915GA SBB11490917GA SBB11490918GA SBB11490919GA SBB11490921GA SBB11490924GA SBB11490926GA SBB11060476GA SBB11140610GA SBB11140611GA SBB11140622GA SBB11410855GA SBB11170686GA SBB11050445GA SBB11040402GA SBB11140610GA SBB11140611GA SBB11140622GA SBB11170699GA SBB11170672GA SBB11060471GA SBB11060481GA SBB11060486GA SBB11060491GA SBB10400072GA SBB10530323GA SBB11050438GA SBB11060475GA SBB11060505GA RT411070188GA RT411070192GA RT411070194GA RT411070195GA SBB11050447GA SBB10510311GA SBB11040407GA SBB10530352GA SBB10430206GA SBB10440256GA RT410490028GA RT410490048GA SBB10309755GA SBB10319774GA SBB10410127GA SBB10420139GA SBB10430204GA RTW11366755GA RTW11366703GA RTW11346651GA RTW11346616GA RTW10475198GA RTW10495299GA SBB12071245GA SBB12071266GA SBB11490896GA SBB11490897GA SBB12041129GA SBB11410809GA SBB11400799GA RTW11346584GA SBB12010982GA RTW11346641GA SBB12031057GA SBB12031017GA SBB11140621GA RTW11346629GA RTW11286145GA RTW11346595GA RTW11055431GA RTW11045400GA RTW11035357GA RTW11045363GA RTW10304643GA RTW10475145GA SBB12061213GA SBB12041109GA SBB12041103GA SBB12031013GA SBB12031039GA SBB12031016GA SBB12010985GA SBB11390782GA SBB11170694GA SBB11170738GA SBB11170695GA SBB11140650GA SBB11170693GA SBB11140645GA SBB11170656GA SBB11110572GA SBB11170677GA SBB11050525GA SBB10530351GA SBB11030384GA SBB11110579GA SBB11050560GA SBB10510313GA SBB10510314GA RT410490024GA RT410490029GA RT410490040GA RT410490043GA SBB10410117GA SBB10410123GA SBB12131355GA SBB12131368GA RTW11366681GA RT412091519GA RTW11326426GA RTW11346604GA RTW11346607GA RTW11346611GA RTW11346612GA RTW11346625GA RTW11346628GA RTW11346635GA RTW11346638GA RTW11346653GA RTW11346601GA RTW11346600GA RTW11346608GA RTW11286154GA RTW11346596GA RTW11125498GA RTW11255997GA RTW11256004GA RTW11215721GA RT411170433GA RTW11165613GA RTW11165627GA RTW11165628GA RTW11045377GA RTW11055414GA RTW11055428GA RTW11035324GA RTW10495287GA RTW10495304GA RTW10465138GA RTW10475172GA RTW11346655GA RTW11366763GA RTW11366770GA SBB12261521GA RTW11346602GA RTW11346548GA RTW11346573GA RTW11346576GA RTW11346579GA RTW11346580GA RTW11346587GA RTW11346589GA RTW11346599GA RTW11346633GA RTW11346639GA RTW11346644GA RTW11346647GA RTW11346648GA SBB11170688GA SBB11060464GA RTW11215722GA RTW11215736GA RTW11215744GA RTW11225774GA RTW11225783GA RTW11225785GA RTW11225791GA RTW11235793GA RTW11235802GA RTW11125458GA RTW11125460GA RTW11125465GA RTW11125470GA RTW11125480GA RTW11125482GA RTW11125467GA RTW11125479GA RTW10465096GA RTW10465116GA RTW10465119GA RTW10475142GA RTW10475147GA RTW10475180GA RTW11035319GA RTW11035333GA RTW11035348GA RTW11045412GA RTW11055418GA RTW11055419GA RTW11055422GA RTW11055432GA RTW11055434GA RTW11125466GA RTW11045411GA RTW11055439GA RTW10465139GA RTW11045360GA RTW11045364GA RTW11045368GA RTW11045373GA RTW11045378GA RTW11045380GA RTW11045384GA RTW11045392GA RTW11045406GA RTW11045366GA RTW11045382GA RTW11045383GA RTW11045388GA RTW11045399GA RTW11045402GA RTW10475161GA RTW10495217GA RTW11035323GA RTW11035329GA RTW11035332GA RTW11035336GA RTW11035343GA RTW11035358GA RTW11035359GA RTW11045361GA RTW11045375GA RTW11045385GA RTW11045386GA RTW11045389GA RTW11045395GA RTW11045401GA RTW11045403GA RTW11045404GA RTW11045405GA RTW11045408GA RTW11035334GA RTW11035347GA RTW11045367GA RTW11045398GA RTW11035318GA RTW11035321GA RTW10495206GA RTW10495216GA RTW10495222GA RTW10495223GA RTW10495225GA RTW10495226GA RTW10495228GA RTW10495230GA RTW10495297GA RTW10495298GA RTW10495300GA RTW11035327GA RTW11035352GA RTW11035330GA RTW10495302GA RTW11035317GA RTW11035326GA RTW11035354GA RTW10495208GA RTW10495288GA RTW10465114GA RTW10475141GA RTW10475152GA RTW10384768GA RTW10384769GA RTW10384779GA RTW10384781GA RTW10384782GA RTW10384786GA RTW10384787GA RTW10384795GA RTW10394850GA RTW10394854GA RTW10394856GA RTW10394876GA RTW10415007GA RTW10415009GA RTW10415012GA RTW10415016GA RTW10415020GA RTW10415024GA RTW10415025GA RTW10415030GA RTW10415033GA RTW10415042GA RTW10415047GA SBB12281524GA SBB12020998GA SBB12031021GA RTW11346577GA RTW11346582GA RTW11346583GA RTW11346588GA RTW11346597GA RTW11346630GA RTW11346634GA RTW11346645GA RTW11346646GA RTW11346650GA RTW11346652GA RTW11346654GA SBB11170736GA SBB11140648GA RTW11235872GA RTW11286104GA RTW11316331GA RTW11316340GA RTW11316363GA RTW11316383GA RTW11326481GA RTW11326496GA RTW11326510GA RTW11326515GA RTW11326516GA RTW11326517GA RTW11326521GA RTW11326522GA RTW11326530GA RTW11326533GA RTW11326534GA RTW11326535GA SBB11140618GA SBB11030389GA SBB10349896GA SBB11140602GA SBB10410135GA RT411080250GA RT411080255GA RT411080265GA SBB12131404GA SBB11030373GA SBB11030382GA SBB11030386GA SBB12131381GA RT412081457GA SBB12041101GA SBB11410808GA SBB11410825GA SBB11410833GA RT411300928GA RT411300929GA RT411080262GA SBB10440270GA SBB10440295GA SBB12041106GA SBB12041120GA SBB12041124GA SBB12010984GA SBB11520960GA SBB12010965GA SBB10440260GA RT412081460GA RT412081462GA RT412081463GA RT412081465GA RT412081469GA RT412081470GA SBB11410854GA SBB11400802GA SBB11030370GA SBB10510307GA SBB11050432GA SBB11060482GA SBB11050459GA SBB11050511GA SBB11050524GA SBB11060465GA SBB11060487GA SBB11060490GA SBB11060495GA SBB11060497GA SBB12281543GA SBB12281544GA SBB12041096GA RT412021410GA RT412021420GA SBB12010986GA SBB11410816GA SBB11410822GA SBB11400800GA SBB10319810GA RT410490052GA RT411090304GA RT411090305GA RT411100321GA RT411100330GA RT411170443GA RT411170450GA SBB12281550GA SBB12281534GA RT412021416GA SBB12021006GA SBB12021000GA SBB12021005GA SBB12021009GA SBB12020988GA SBB12020991GA SBB12020994GA SBB11410814GA SBB11170696GA SBB11170742GA SBB11170754GA RT411110364GA SBB11030361GA RTW10495209GA SBB10319813GA SBB12041116GA SBB12041085GA SBB11390780GA SBB11170730GA RTW11035331GA RTW11366735GA RTW11366729GA RTW11376810GA RTW11406890GA RTW11376816GA RTW11286103GA SBB12041089GA SBB12131363GA SBB12131400GA SBB12131401GA SBB12131406GA RTW11346637GA SBB12031055GA SBB12031058GA SBB12031063GA SBB12031064GA SBB12031066GA RTW11346606GA SBB11490877GA SBB11490878GA SBB11490879GA SBB11490880GA SBB11490881GA SBB11490884GA SBB11490886GA SBB11490901GA SBB11490902GA SBB11490905GA RTW11346623GA RTW11346575GA RTW11346617GA RTW11346621GA RTW11256051GA RTW11286167GA RTW11256006GA SBB10299651GA SBB10400041GA SBB10400042GA SBB10400043GA SBB10430185GA SBB10510308GA SBB11030356GA SBB11030357GA SBB11030359GA SBB11030360GA SBB11030364GA SBB11030366GA SBB11030367GA SBB11030378GA SBB11030381GA SBB11040408GA SBB11040411GA SBB11040420GA SBB11040421GA SBB11040424GA SBB11050513GA SBB11050516GA SBB11050520GA SBB11060484GA SBB11060489GA SBB11110576GA SBB11110585GA SBB11110600GA SBB11140624GA SBB11140628GA SBB11140632GA SBB11140638GA SBB11140642GA SBB11140643GA RTW11155547GA SBB11030371GA SBB11030375GA SBB11030377GA SBB11030388GA SBB11030390GA SBB11030391GA SBB11030392GA SBB11030396GA SBB11030397GA SBB11040414GA RTW11155517GA RTW11055420GA RTW10495205GA RT411411259GA RT411411261GA RT411411269GA RT411411275GA RT411411287GA RT411411288GA RT411451302GA SBB11030365GA SBB10440251GA
|
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
800-558-7044
|
Manufacturer Reason for Recall |
GE Healthcare has recently become aware of a potential safety issue due to intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG connector on your DASH, PDM, and/or Tram Module.
ECG and Respiration signal noise caused by DASH, PDM, and/or Tram ECG cable connector movement may result in reduced ECG, Respiration and arrhythmia detection performance.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
GE Healthcare "Urgent Medical Device Correction" letter dated April 3, 2013 to all affected customers. The letter was addressed to Healthcare Administrator / Risk manger, Chief of Nursing and Director of Biomedical Engineering. The letter described the Safety Issue, Product, Safety Instructions, Affected Product Details, and Product Correction. For questions contact Technical Support at 1-800-558-7044. |
Quantity in Commerce |
4807 (2167 US, 2207 OUS, & 434 unknown. |
Distribution |
Worldwide Distribution - USA Nationwide including DC and PR except ND and the countries of: ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BANGLADESH, BELARUS, BELGIUM, BOSNIA-HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, CHANA, GREECE, HONDURAS, HUNGARY, ICELAND, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, NEPAL, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, PANAMA, PARAGUAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAK IA, SLOVENIA, SOUTH AFRICA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TUNSIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, BOLIVIA, HONG KONG, JAMAICA, PHILIPPINES, SOUTH KOREA, TRINIDAD TOBAGO, URUGUAY, VENEZUELA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MHX and Original Applicant = GE HEALTHCARE
|
|
|
|