Class 2 Device Recall Optima Mobile Xray System |
 |
Date Initiated by Firm |
November 07, 2013 |
Date Posted |
January 11, 2014 |
Recall Status1 |
Terminated 3 on May 06, 2014 |
Recall Number |
Z-0725-2014 |
Recall Event ID |
67136 |
510(K)Number |
K103476
|
Product Classification |
System, x-ray, mobile - Product Code IZL
|
Product |
GE Optima XR220amx and Optima XR200amx with Digital Upgrade Mobile Radiographic Systems. Model numbers 5555000-5, 5555000-6, and 5421698. |
Code Information |
Serial Numbers: 00000002796EN1, 00000002797EN9, 00000002798EN7, 00000002799EN5, 00000002800EN1, 00000002801EN9, 00000102726WK9, 00000102789WK3, 00000102937WK7, 00000130778WK3, 00000UA45526-8, 00001027058WK5, 00001027061WK9, 00001027146WK8, 00001027149WK2, 00001027150WK0, 00001027151WK8, 00001027153WK4, 00001027154WK2, 00001027167WK4, 00001027168WK2, 00001027169WK0, 00001027175WK7, 00001027193WK0, 00001027194WK8, 00001027195WK5, 00001027199WK7, 00001027200WK3, 00001027203WK7, 00001027204WK5, 00001027207WK8, 00001027214WK4, 00001027256WK5, 00001027259WK9, 00001027260WK7, 00001027262WK3, 00001027264WK9, 00001027265WK6, 00001027267WK2, 00001027268WK0, 00001027269WK8, 00001027276WK3, 00001027277WK1, 00001027278WK9, 00001027280WK5, 00001027293WK8, 00001027294WK6, 00001027295WK3, 00001027297WK9, 00001027298WK7, 00001027299WK5, 00001027300WK1, 00001027301WK9, 00001027302WK7, 00001027303WK5, 00001027314WK2, 00001027318WK3, 00001027322WK5, 00001027323WK3, 00001027324WK1, 00001027325WK8, 00001027326WK6, 00001027344WK9, 00001027346WK4, 00001027347WK2, 00001027349WK8, 00001027350WK6, 00001027351WK4, 00001027352WK2, 00001027353WK0, 00001027354WK8, 00001027356WK3, 00001027359WK7, 00001027362WK1, 00001027363WK9, 00001027364WK7, 00001027369WK6, 00001027371WK2, 00001027372WK0, 00001027373WK8, 00001027374WK6, 00001027376WK1, 00001027377WK9, 00001027380WK3, 00001027382WK9, 00001027383WK7, 00001027384WK5, 00001027386WK0, 00001027387WK8, 00001027388WK6, 00001027389WK4, 00001027390WK2, 00001027406WK6, 00001027407WK4, 00001027416WK5, 00001027417WK3, 00001027418WK1, 00001027421WK5, 00001027422WK3, 00001027423WK1, 00001027438WK9, 00001027445WK4, 00001027450WK4, 00001027451WK2, 00001027452WK0, 00001027464WK5, 00001027465WK2, 00001027466WK0, 00001027487WK6, 00001027488WK4, 00001027489WK2, 00001027490WK0, 00001027506WK3, 00001027507WK1, 00001027508WK9, 00001027539WK4, 00001027540WK2, 00001027541WK0, 00001027542WK8, 00001027543WK6, 00001027544WK4, 00001027551WK9, 00001027552WK7, 00001027553WK5, 00001027582WK4, 00001027583WK2, 00001027699WK6, 00001027700WK2, 00001027736WK6, 00001027737WK4, 00001027749WK9, 00001027762WK2, 00001027764WK8, 00001027765WK5, 00001027766WK3, 00001027768WK9, 00001027769WK7, 00001027772WK1, 00001027773WK9, 00001027774WK7, 00001027775WK4, 00001027778WK8, 00001027779WK6, 00001027781WK2, 00001027814WK1, 00001027819WK0, 00001027820WK8, 00001027821WK6, 00001027823WK2, 00001027824WK0, 00001027825WK7, 00001027826WK5, 00001027827WK3, 00001027828WK1, 00001027878WK6, 00001027880WK2, 00001027887WK7, 00001027888WK5, 00001027889WK3, 00001027890WK1, 00001027891WK9, 00001027892WK7, 00001027893WK5, 00001027894WK3, 00001027895WK0, 00001027896WK8, 00001027915WK6, 00001027932WK1, 00001027934WK7, 00001027935WK4, 00001027939WK6, 00001027992WK5, 00001027993WK3, 00001027994WK1, 00001027995WK8, 00001027996WK6, 00001028003WK0, 00001028021WK2, 00001028022WK0, 00001028023WK8, 00001028025WK3, 00001028035WK2, 00001028036WK0, 00001028042WK8, 00001028043WK6, 00001028044WK4, 00001028050WK1, 00001028051WK9, 00001028052WK7, 00001028053WK5, 00001028064WK2, 00001028065WK9, 00001028066WK7, 00001028077WK4, 00001028078WK2, 00001028079WK0, 00001028080WK8, 00001028091WK5, 00001028092WK3, 00001028093WK1, 00001028094WK9, 00001028119WK4, 00001028120WK2, 00001028205WK1, 00001028213WK5, 00001028214WK3, 00001028215WK0, 00001028216WK8, 00001028217WK6, 00001028230WK9, 00001028231WK7, 00001028232WK5, 00001028267WK1, 00001028269WK7, 00001028275WK4, 00001028276WK2, 00001028277WK0, 00001028279WK6, 00001028280WK4, 00001028281WK2, 00001028283WK8, 00001028284WK6, 00001028285WK3, 00001028286WK1, 00001028287WK9, 00001028288WK7, 00001028289WK5, 00001028290WK3, 00001028291WK1, 00001028292WK9, 00001028294WK5, 00001028331WK5, 00001028333WK1, 00001028334WK9, 00001028337WK2, 00001028338WK0, 00001028339WK8, 00001028340WK6, 00001028341WK4, 00001028342WK2, 00001028343WK0, 00001028344WK8, 00001028345WK5, 00001028346WK3, 00001028347WK1, 00001028351WK3, 00001028352WK1, 00001028353WK9, 00001028354WK7, 00001028363WK8, 00001028365WK3, 00001028369WK5, 00001028370WK3, 00001028396WK8, 00001028397WK6, 00001028398WK4, 00001028399WK2, 00001028464WK4, 00001028465WK1, 00001028467WK7, 00001028468WK5, 00001028469WK3, 00001028470WK1, 00001028471WK9, 00001028472WK7, 00001028476WK8, 00001028477WK6, 00001028478WK4, 00001028501WK3, 00001028502WK1, 00001028503WK9, 00001028505WK4, 00001028506WK2, 00001028517WK9, 00001028518WK7, 00001028519WK5, 00001028520WK3, 00001028522WK9, 00001028523WK7, 00001028524WK5, 00001028536WK9, 00001028537WK7, 00001028538WK5, 00001028539WK3, 00001028559WK1, 00001028560WK9, 00001028561WK7, 00001028562WK5, 00001028563WK3, 00001028574WK0, 00001028575WK7, 00001028576WK5, 00001028605WK2, 00001028617WK7, 00001028621WK9, 00001028680WK5, 00001028681WK3, 00001028682WK1, 00001028683WK9, 00001028685WK4, 00001028686WK2, 00001028687WK0, 00001028688WK8, 00001028689WK6, 00001028690WK4, 00001028691WK2, 00001028692WK0, 00001028693WK8, 00001028694WK6, 00001028695WK3, 00001028716WK7, 00001028717WK5, 00001028718WK3, 00001028719WK1, 00001028799WK3, 00001028800WK9, 00001028801WK7, 00001028802WK5, 00001028803WK3, 00001028850WK4, 00001028851WK2, 00001028852WK0, 00001028853WK8, 00001028854WK6, 00001028855WK3, 00001028856WK1, 00001028857WK9, 00001028858WK7, 00001028859WK5, 00001028889WK2, 00001028890WK0, 00001028891WK8, 00001028895WK9, 00001028896WK7, 00001028897WK5, 00001028898WK3, 00001028899WK1, 00001028900WK7, 00001028901WK5, 00001028902WK3, 00001028916WK3, 00001028917WK1, 00001028977WK5, 00001028979WK1, 00001028980WK9, 00001028982WK5, 00001028983WK3, 00001028984WK1, 00001028985WK8, 00001028986WK6, 00001028989WK0, 00001028990WK8, 00001028991WK6, 00001028992WK4, 00001028993WK2, 00001028994WK0, 00001028995WK7, 00001028996WK5, 00001028997WK3, 00001028998WK1, 00001029036WK9, 00001029037WK7, 00001029038WK5, 00001029039WK3, 00001029040WK1, 00001029041WK9, 00001029042WK7, 00001029043WK5, 00001029044WK3, 00001029046WK8, 00001029048WK4, 00001029049WK2, 00001029050WK0, 00001029051WK8, 00001029052WK6, 00001029053WK4, 00001029054WK2, 00001029055WK9, 00001029062WK5, 00001029066WK6, 00001029067WK4, 00001029068WK2, 00001029069WK0, 00001029070WK8, 00001029159WK9, 00001029160WK7, 00001029161WK5, 00001029162WK3, 00001029163WK1, 00001029212WK6, 00001029220WK9, 00001029366WK0, 00001029373WK6, 00001029374WK4, 00001029375WK1, 00001029376WK9, 00001029378WK5, 00001029442WK9, 00001029443WK7, 00001029444WK5, 00001029446WK0, 00001029460WK1, 00001029461WK9, 00001029462WK7, 00001029464WK3, 00001029465WK0, 00001029472WK6, 00001029473WK4, 00001029475WK9, 00001029476WK7, 00001029521WK0, 00001029523WK6, 00001029539WK2, 00001029540WK0, 00001029541WK8, 00001029542WK6, 00001029543WK4, 00001029544WK2, 00001029545WK9, 00001029546WK7, 00001029566WK5, 00001029567WK3, 00001029568WK1, 00001029569WK9, 00001029570WK7, 00001029571WK5, 00001029572WK3, 00001029573WK1, 00001029574WK9, 00001029575WK6, 00001029592WK1, 00001029593WK9, 00001029594WK7, 00001029595WK4, 00001029597WK0, 00001029598WK8, 00001029614WK3, 00001029615WK0, 00001029616WK8, 00001029620WK0, 00001029621WK8, 00001029622WK6, 00001029625WK9, 00001029626WK7, 00001029628WK3, 00001029630WK9, 00001029631WK7, 00001029632WK5, 00001029635WK8, 00001029636WK6, 00001029639WK0, 00001029640WK8, 00001029641WK6, 00001029642WK4, 00001029643WK2, 00001029649WK9, 00001029650WK7, 00001029651WK5, 00001029652WK3, 00001029653WK1, 00001029654WK9, 00001029655WK6, 00001029656WK4, 00001029657WK2, 00001029658WK0, 00001029674WK7, 00001029675WK4, 00001029676WK2, 00001029677WK0, 00001029678WK8, 00001029679WK6, 00001029681WK2, 00001029683WK8, 00001029684WK6, 00001029685WK3, 00001029693WK7, 00001029694WK5, 00001029695WK2, 00001029696WK0, 00001029697WK8, 00001029698WK6, 00001029699WK4, 00001029700WK0, 00001029701WK8, 00001029702WK6, 00001029724WK0, 00001029725WK7, 00001029726WK5, 00001029727WK3, 00001029728WK1, 00001029729WK9, 00001029733WK1, 00001029734WK9, 00001029735WK6, 00001029736WK4, 00001029737WK2, 00001029738WK0, 00001029739WK8, 00001029740WK6, 00001029741WK4, 00001029742WK2, 00001029743WK0, 00001029744WK8, 00001029745WK5, 00001029746WK3, 00001029747WK1, 00001029755WK4, 00001029756WK2, 00001029758WK8, 00001029759WK6, 00001029762WK0, 00001029763WK8, 00001029764WK6, 00001029766WK1, 00001029767WK9, 00001029768WK7, 00001029769WK5, 00001029770WK3, 00001029786WK9, 00001029787WK7, 00001029796WK8, 00001029797WK6, 00001029804WK0, 00001029813WK1, 00001029814WK9, 00001029820WK6, 00001029821WK4, 00001029825WK5, 00001029826WK3, 00001029827WK1, 00001029828WK9, 00001029861WK0, 00001029862WK8, 00001029863WK6, 00001029864WK4, 00001029865WK1, 00001029868WK5, 00001029869WK3, 00001029870WK1, 00001029885WK9, 00001029886WK7, 00001029887WK5, 00001029888WK3, 00001029889WK1, 00001029890WK9, 00001029900WK6, 00001029901WK4, 00001029902WK2, 00001029903WK0, 00001029931WK1, 00001029932WK9, 00001029933WK7, 00001029934WK5, 00001029935WK2, 00001029941WK0, 00001029948WK5, 00001029949WK3, 00001029950WK1, 00001029951WK9, 00001029952WK7, 00001029953WK5, 00001029954WK3, 00001029955WK0, 00001029956WK8, 00001029959WK2, 00001029960WK0, 00001029961WK8, 00001029962WK6, 00001029963WK4, 00001029969WK1, 00001029970WK9, 00001029973WK3, 00001029974WK1, 00001030002WK8, 00001030003WK6, 00001030004WK4, 00001030008WK5, 00001030009WK3, 00001030010WK1, 00001030011WK9, 00001030012WK7, 00001030025WK9, 00001030026WK7, 00001030027WK5, 00001030028WK3, 00001030029WK1, 00001030031WK7, 00001030032WK5, 00001030033WK3, 00001030035WK8, 00001030045WK7, 00001030046WK5, 00001030049WK9, 00001030050WK7, 00001030052WK3, 00001030054WK9, 00001030057WK2, 00001030065WK5, 00001030075WK4, 00001030082WK0, 00001030206WK5, 00001030210WK7, 00001030211WK5, 00001030212WK3, 00001030216WK4, 00001030217WK2, 00001030218WK0, 00001030219WK8, 00001030235WK4, 00001030236WK2, 00001030239WK6, 00001030240WK4, 00001030241WK2, 00001030242WK0, 00001030243WK8, 00001030269WK3, 00001030270WK1, 00001030271WK9, 00001030272WK7, 00001030273WK5, 00001030284WK2, 00001030285WK9, 00001030286WK7, 00001030287WK5, 00001030288WK3, 00001030289WK1, 00001030290WK9, 00001030291WK7, 00001030292WK5, 00001030293WK3, 00001030366WK7, 00001030367WK5, 00001030368WK3, 00001030369WK1, 00001030371WK7, 00001030406WK1, 00001030407WK9, 00001030408WK7, 00001030432WK7, 00001030433WK5, 00001030437WK6, 00001030438WK4, 00001030510WK0, 00001030511WK8, 00001030512WK6, 00001030513WK4, 00001030514WK2, 00001030531WK6, 00001030532WK4, 00001030533WK2, 00001030534WK0, 00001030535WK7, 00001030536WK5, 00001030537WK3, 00001030538WK1, 00001030539WK9, 00001030540WK7, 00001030557WK1, 00001030558WK9, 00001030559WK7, 00001030560WK5, 00001030561WK3, 00001030562WK1, 00001030585WK2, 00001030586WK0, 00001030587WK8, 00001030588WK6, 00001030589WK4, 00001030590WK2, 00001030591WK0, 00001030592WK8, 00001030593WK6, 00001030594WK4, 00001030595WK1, 00001030596WK9, 00001030597WK7, 00001030598WK5, 00001030602WK5, 00001030603WK3, 00001030604WK1, 00001030605WK8, 00001030606WK6, 00001030607WK4, 00001030617WK3, 00001030618WK1, 00001030619WK9, 00001030627WK2, 00001030628WK0, 00001030636WK3, 00001030637WK1, 00001030638WK9, 00001030654WK6, 00001030655WK3, 00001030656WK1, 00001030657WK9, 00001030660WK3, 00001030661WK1, 00001030669WK4, 00001030671WK0, 00001030672WK8, 00001030673WK6, 00001030674WK4, 00001030675WK1, 00001030736WK1, 00001030737WK9, 00001030738WK7, 00001030743WK7, 00001030744WK5, 00001030759WK3, 00001030760WK1, 00001030761WK9, 00001030762WK7, 00001030763WK5, 00001030775WK9, 00001030776WK7, 00001030777WK5, 00001030778WK3, 00001030779WK1, 00001030780WK9, 00001030781WK7, 00001030782WK5, 00001030791WK6, 00001030792WK4, 00001030793WK2, 00001030794WK0, 00001030795WK7, 00001030796WK5, 00001030797WK3, 00001030799WK9, 00001030800WK5, 00001030801WK3, 00001030802WK1, 00001030803WK9, 00001030804WK7, 00001030805WK4, 00001030806WK2, 00001030807WK0, 00001030808WK8, 00001030809WK6, 00001030810WK4, 00001030811WK2, 00001030812WK0, 00001030813WK8, 00001030814WK6, 00001030815WK3, 00001030816WK1, 00001030839WK3, 00001030840WK1, 00001030841WK9, 00001030842WK7, 00001030843WK5, 00001030844WK3, 00001030845WK0, 00001030846WK8, 00001030847WK6, 00001030849WK2, 00001030850WK0, 00001030854WK2, 00001030856WK7, 00001030899WK7, 00001030900WK3, 00001030901WK1, 00001030903WK7, 00001030904WK5, 00001030905WK2, 00001030907WK8, 00001030908WK6, 00001030909WK4, 00001030915WK1, 00001030916WK9, 00001030917WK7, 00001030918WK5, 00001030919WK3, 00001030920WK1, 00001030921WK9, 00001030922WK7, 00001030925WK0, 00001030926WK8, 00001030928WK4, 00001030929WK2, 00001030930WK0, 00001030950WK8, 00001030951WK6, 00001030952WK4, 00001030953WK2, 00001030954WK0, 00001030955WK7, 00001030956WK5, 00001030957WK3, 00001030958WK1, 00001030959WK9, 00001030960WK7, 00001030962WK3, 00001030963WK1, 00001030964WK9, 00001030993WK8, 00001030994WK6, 00001030995WK3, 00001031006WK8, 00001031007WK6, 00001031008WK4, 00001031009WK2, 00001031010WK0, 00001031011WK8, 00001031012WK6, 00001031013WK4, 00001031014WK2, 00001031017WK5, 00001031018WK3, 00001031019WK1, 00001031020WK9, 00001031021WK7, 00001031028WK2, 00001031029WK0, 00001031031WK6, 00001031032WK4, 00001031033WK2, 00001031053WK0, 00001031054WK8, 00001031055WK5, 00001031058WK9, 00001031059WK7, 00001031060WK5, 00001031061WK3, 00001031062WK1, 00001031063WK9, 00001031064WK7, 00001031108WK2, 00001031109WK0, 00001031110WK8, 00001031111WK6, 00001031112WK4, 00001031113WK2, 00001031116WK5, 00001031117WK3, 00001031118WK1, 00001031119WK9, 00001031120WK7, 00001031121WK5, 00001031122WK3, 00001031123WK1, 00001031124WK9, 00001031125WK6, 00001031153WK8, 00001031154WK6, 00001031155WK3, 00001031156WK1, 00001031157WK9, 00001031158WK7, 00001031166WK0, 00001031167WK8, 00001031168WK6, 00001031169WK4, 00001031170WK2, 00001031172WK8, 00001031173WK6, 00001031174WK4, 00001031175WK1, 00001031176WK9, 00001031177WK7, 00001031178WK5, 00001031179WK3, 00001031181WK9, 00001031206WK4, 00001031207WK2, 00001031208WK0, 00001031213WK0, 00001031214WK8, 00001031215WK5, 00001031215WK5, 00001031216WK3, 00001031219WK7, 00001031220WK5, 00001031221WK3, 00001031222WK1, 00001031226WK2, 00001031227WK0, 00001031228WK8, 00001031229WK6, 00001031235WK3, 00001031238WK7, 00001031239WK5, 00001031246WK0, 00001031247WK8, 00001031248WK6, 00001031249WK4, 00001031250WK2, 00001031251WK0, 00001031281WK7, 00001031282WK5, 00001031283WK3, 00001031284WK1, 00001031286WK6, 00001031287WK4, 00001031332WK8, 00001031334WK4, 00001031335WK1, 00001031336WK9, 00001031337WK7, 00001031338WK5, 00001031339WK3, 00001031341WK9, 00001031342WK7, 00001031343WK5, 00001031350WK0, 00001031351WK8, 00001031352WK6, 00001031353WK4, 00001031361WK7, 00001031362WK5, 00001031363WK3, 00001031386WK4, 00001031387WK2, 00001031389WK8, 00001031390WK6, 00001031391WK4, 00001031392WK2, 00001031401WK1, 00001031402WK9, 00001031403WK7, 00001031408WK6, 00001031413WK6, 00001031414WK4, 00001031415WK1, 00001031416WK9, 00001031457WK3, 00001031458WK1, 00001031459WK9, 00001031460WK7, 00001031461WK5, 00001031462WK3, 00001031464WK9, 00001031465WK6, 00001031466WK4, 00001031467WK2, 00001031468WK0, 00001031479WK7, 00001031480WK5, 00001031481WK3, 00001031483WK9, 00001031484WK7, 00001031490WK4, 00001031491WK2, 00001031492WK0, 00001031493WK8, 00001031495WK3, 00001031496WK1, 00001031506WK7, 00001031507WK5, 00001031508WK3, 00001031510WK9, 00001031511WK7, 00001031512WK5, 00001031515WK8, 00001031516WK6, 00001031517WK4, 00001031518WK2, 00001031519WK0, 00001031520WK8, 00001031521WK6, 00001031522WK4, 00001031523WK2, 00001031524WK0, 00001031525WK7, 00001031537WK2, 00001031538WK0, 00001031539WK8, 00001031566WK1, 00001031567WK9, 00001031568WK7, 00001031569WK5, 00001031570WK3, 00001031571WK1, 00001031572WK9, 00001031573WK7, 00001031574WK5, 00001031581WK0, 00001031649WK5, 00001031650WK3, 00001031651WK1, 00001031652WK9, 00001031653WK7, 00001031654WK5, 00001031655WK2, 00001031656WK0, 00001031657WK8, 00001031658WK6, 00001031659WK4, 00001031660WK2, 00001031661WK0, 00001031662WK8, 00001031665WK1, 00001031666WK9, 00001031667WK7, 00001031669WK3, 00001031670WK1, 00001031671WK9, 00001031672WK7, 00001031673WK5, 00001031674WK3, 00001031675WK0, 00001031676WK8, 00001031677WK6, 00001031699WK0, 00001031700WK6, 00001031702WK2, 00001031703WK0, 00001031704WK8, 00001031705WK5, 00001031706WK3, 00001031711WK3, 00001031712WK1, 00001031728WK7, 00001031729WK5, 00001031730WK3, 00001031733WK7, 00001031734WK5, 00001031735WK2, 00001031738WK6, 00001031739WK4, 00001031740WK2, 00001031749WK3, 00001031750WK1, 00001031751WK9, 00001031752WK7, 00001031810WK3, 00010299771WK7. |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
Call Center 800-437-1171
|
Manufacturer Reason for Recall |
There exists a software issue associated with patient selection from the worklist on the Optima XR200 with Digital Upgrade, and Optima XR220amx Mobile X-ray Systems. Intermittently when a large number of patient exams are retrieved from the HIS/RIS system, the exams when displayed on the mobile x-ray system are not properly displayed. If the user continues to attempt to select one of the exams w
|
FDA Determined Cause 2 |
Software design |
Action |
GE sent an Urgent Medical Device Correction letter, dated 11/7/2013 to consignees. The letter described the issue, identified devices affected and stated that users should pay close attention to the patient exam/entry selected from the worklist prior to selecting the Start Exam button to ensure that the system did not update and select an incorrect exam. The user may also confirm during the exam that the appropriate patient information was selected via the display on the upper left portion. GE will fix the units with a software fix at no cost. |
Quantity in Commerce |
1027 |
Distribution |
Worldwide Distribution-USA (nationwide) including Washington D.C., and Guam and the countries of ALBANIA, AUSTRALIA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CHILE, CHINA , COLOMBIA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, IRAQ, IRELAND, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MEXICO, NEW ZEALAND, NIGERIA, NORWAY, POLAND, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, THAILAND, TRINIDAD AND TOBAGO, TURKEY, UNITED ARAB EMIRATES, and UNITED KINGDOM. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IZL and Original Applicant = GE MEDICAL SYSTEMS, LLC
|
|
|
|