 Class 2 Device Recall Apollo" Revision/Constrained Knee Systemsee related information
Date Initiated by FirmFebruary 25, 2014
Date PostedMarch 20, 2014
Recall Status1 Terminated 3 on March 24, 2015
Recall NumberZ-1261-2014
Recall Event ID 67503
510(K)NumberK972501 
Product Classification Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
ProductApollo" Revision/Constrained Knee System: SZ0-16MM AK CCK TIB INSER SZ1-11MM AK CCK TIB INSER SZ1-16MM AK CCK TIB INSER SZ1-22MM AK CCK TIB INSER SZ1-30MM AK CCK TIB INSER SZ2-11MM AK CCK TIB INSER SZ2-13MM AK CCK TIB INSER SZ2-16MM AK CCK TIB INSER SZ2-22MM AK CCK TIB INSER SZ2-30MM AK CCK TIB INSER SZ3-16MM AK CCK TIB INSER SZ3-22MM AK CCK TIB INSER SZ3-25MM AK CCK TIB INSER SZ4-13MM AK CCK TIB INSER SZ4-16MM AK CCK TIB INSER SZ5-11MM AK CCK TIB INSER
Code Information Item Number 682000016 682001011 682001016 682001022 682001030 682002011 682002013 682002016 682002022 682002030 682003016 682003022 682003025 682004013 682004016 682005011 
FEI Number 1000220733
Recalling Firm/
Manufacturer
Zimmer, Inc.
1800 W Center St
Warsaw IN 46580-2304
For Additional Information Contact
877-946-2761
Manufacturer Reason
for Recall
A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.
FDA Determined
Cause 2
Storage
ActionBeginning 2/25/2014 all distributors were notified via electronic mail. Hospital risk managers and distributors, surgeons with affected inventory will also be notified via courier or direct mail. The Urgent Lot Specific Recall Notice identified the issue and responsibilities. Distributors were asked to locate and remove the affected product in their territory, as well as identifying hospitals and surgeons who have previously implanted the affected product. Distributors were asked to return on-hand affected products to Zimmer and ensure all of their products are accounted for using the form provided in the letter and email it back to corporatequality.postmarket@zimmer.com. Hospital risk managers and surgeons were asked to locate affected product and return it to their Zimmer sales representative. Questions or concerns should be addressed to customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST Communications outside of the United States will occur approximately two weeks after the United States communications. 100 % effective checks will be done.
Quantity in Commerce3638 units
DistributionWorldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = JWH
-
-