| Class 2 Device Recall AGFA Digital Radiography XRay system DXD 100 |  |
Date Initiated by Firm | June 06, 2014 |
Date Posted | July 21, 2014 |
Recall Status1 |
Terminated 3 on June 27, 2016 |
Recall Number | Z-2081-2014 |
Recall Event ID |
68567 |
510(K)Number | K103597 |
Product Classification |
System, x-ray, mobile - Product Code IZL
|
Product | AGFA Digital Radiography X-Ray system DX-D 100
Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be used with MUSICA2 image processing to create radiographic images of the skeleton including skull, spinal column and extremities: chest, abdomen, and other body parts. Agfa's DX-D 100 is not indicated for use in mammography. |
Code Information |
Serial # A5410000053, A5410000054, A5410000067, A5410000068, A5410000069, A5410000070, A5410000072, A5410000073, A5411000315, A5411000331, A5411000332, A5411000141, A5411000144, A5411000038, A5411000069, A5411000133, A5411000175, A5411000119, A5411000123, A5411000333, A5411000334, A5410000038, A5411000019, A5411000047, A5411000048, A5411000053, A5410000041, A5410000042, A5410000046, A5410000016, A5410000039, A5410000021, A5410000023, A5411000074, A5411000080, A5411000024, A5411000184, A5411000193, A5411000196, A5410000100, A5410000114, A5410000115, A5410000116, A5410000117, A5411000342, A5411000343, A5411000344, A5411000366, A5411000368, A5411000237, A5411000027, A5411000020, A5411000021, A5411000132, A5411000070, A5411000078, A5411000082, A5411000154, A5411000156, A54110000107, A5410000082, A5410000066, A5411000236, A5411000155, A5411000151, A5411000148, A5411000149, A5411000134, A5411000120, A5411000125, A5411000147, A5410000079, A5411000185, A5411000191, A5411000192, A5411000194, A5411000239, A5411000052, A5411000030, A5411000314, A5411000348, A5411000268, A5411000347, A5411000349, A5411000140, A5411000142, A5411000097, A5410000022, A5411000017, A5410000027, A5411000153, A5411000233, A5411000234, A5411000235, A5411000318, A5410000044, A5411000037, A5410000043, A5411000181, A5411000158, A5411000152, A5411000263, A5411000309, A5411000199, A5411000209, A5411000210, A5411000232, A5411000346, A5411000124, A5411000174, A5411000177, A5411000230, A5411000173, A5411000316, A5411000145, A5411000188, A5411000197, A5411000202, A5411000203, A5411000206, A5411000208, A5411000335, A5411000336, A5411000159, A5411000041, A5411000050, A5411000088, A5411000247, A5411000079, A5411000040, A5411000240, A5411000085, A5411000277, A5411000278, A5411000279, A5411000280, A5411000281, A5411000282, A5411000245, A5411000064, A5411000065, A5411000187, A5411000198, A5411000207, A5411000211, A5411000212, A5411000231, A5411000039, A5411000032, A5411000083, A5411000122, A5411000108, A5411000110, A5411000241, A5411000246, A5411000114, A5411000033, A5411000139, A5411000312, A5411000182, A5411000190, A5411000218, A5411000249, A5411000084, A5411000098, A5411000238, A5411000116, A5410000042, A5411000023, A5411000337, A5411000338, A5411000340, A5411000314, A5411000117, A5411000121, A5411000262, A5411000270, A5411000272, A5411000273, A5411000293, A5411000260, A5411000264, A5411000269, A5411000271, A5411000291, A5411000292, A5411000261, A5411000265, A5411000266, A5411000267, A5411000363, A5411000034, A5411000025, A5411000081, A5411000146, A5411000092, A5411000093, A5411000026, A5411000068, A5411000131, A5411000060, A5411000018, A5411000100, A5411000313, A5411000284, A5411000076, A5411000135, A5411000180, A5411000285, A5411000311, A5411000286, A5411000062, A5411000136, A5411000178, A5411000179, A5411000283 |
Recalling Firm/ Manufacturer |
AGFA Corp. 10 S Academy St Greenville SC 29601-2632
|
For Additional Information Contact | Debbie Huff 864-421-1754 |
Manufacturer Reason for Recall | When liquid comes in contact with the DX-D 100 touch screen, the device may incorrectly recognize this as user input altering device settings. |
FDA Determined Cause 2 | Use error |
Action | Agfa Healthcare sent an "URGENT FIELD SAFETY NOTICE" dated June 6, 2014, to the US and Canadian customers. The letter described the safety alert and mitigation. Acknowledgment, via FAX-Back or email, that the information was received and understood was requested from the customers.
Although medical staff are required to disinfect hands on a frequent basis, users must ensure that their hands are dry before using the touch screen of the DX-0100 because liquid residue may activate the action buttons on the touch screen.
Customers were instructed:
"Do not operate touch screen monitor with wet hands!
"Do not let liquids come in contact with the touch screen while the DXD 100 is powered on!
"Always double check your parameter settings prior to exposing the patient. We would like to remind you of this specific warning statement in the DX-0100 User Manual.
Please complete the feedback form as soon as possible and return it to us by June 16, 2014.
Should the above information not apply to your facility or should the device have been transferred to another organization, please be so kind as to indicate this on the attached feedback form and pass this Urgent Field Safety Notice to the organization where the device has been transferred.
We thank you for your careful attention to this issue and your continued support.
If you have any questions about this matter, please contact your local Agfa Healthcare organization: 1-877-777-2432, prompt 1 and reference PR1406020002 |
Quantity in Commerce | 216 units |
Distribution | USA (nationwide) including the states of AL, CA, FL, GA, IL, IN, IA, KY, LA, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, SC, TN, TX, VA, WV, WI, WY and Internationally to Canada. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = IZL
|
|
|
|