 Class 2 Device Recall ENDOTAK RELIANCE SGsee related information
Date Initiated by FirmJune 19, 2014
Date PostedAugust 05, 2014
Recall Status1 Terminated 3 on December 03, 2014
Recall NumberZ-2147-2014
Recall Event ID 68700
PMA NumberP919973S115 
Product Classification Implantable cardioverter defibrillator (non-CRT) - Product Code LWS
ProductBoston Scientific ENDOTAK RELIANCE SG, transvenous defibrillation lead, Models 0180 & 0292, Sterile EO. Product Usage: The ENDOTAK RELIANCE leads provide pacing and rate-sensing and deliver cardioversion and defibrillation shocks for automatic implantable cardioverter defibrillator (AICD) systems.
Code Information Model 0180 s/n 310735 & 310892. Model 0292 s/n 130450, & 309339. 
Recalling Firm/
Manufacturer
Boston Scientific CRM Corp
4100 Hamline Ave N
Saint Paul MN 55112-5700
For Additional Information ContactLocal Sales Representative
800-227-3422
Manufacturer Reason
for Recall
A review of manufacturing test records for Boston Scientific ENDOTAK RELIANCE implantable leads revealed a suspected test data recording error. Specifically, some test results were recorded as "failed" without any other indication of failure. There are no reported injuries from the devices.
FDA Determined
Cause 2
Software Manufacturing/Software Deployment
ActionSales representatives hand delivered a Boston Scientific "Medical Device Retrieval" Letter dated June 19, 2014 or June 20, 2014 when they were retrieving the devices from the hospital shelves. The letter was addressed to Hospital Administrator. The letter described the problem, product being recalled and the retrieval of the device by Boston scientific sales representative. For further questions they can contact their local sales representative or Boston Scientific International Technical Services. Physicians were contacted via telephone starting on July 6, 2014. A Boston Scientific letter dated July 8, 2014 followed the telephone conversation. For further information physicians can contact their local Boston Scientific representative or Technical Services at 1-800-227-3422.
Quantity in Commerce21
DistributionWorldwide Distribution -US (nationwide) in the states of OH, MN and country of France