|
Class 2 Device Recall Innova IGS 540 |
 |
Date Initiated by Firm |
September 04, 2014 |
Date Posted |
December 16, 2014 |
Recall Status1 |
Terminated 3 on January 21, 2015 |
Recall Number |
Z-0451-2015 |
Recall Event ID |
69286 |
510(K)Number |
K122457
|
Product Classification |
System, x-ray, angiographic - Product Code IZI
|
Product |
Innova IGS 540
Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. |
Code Information |
Mfg Lot or Serial # System ID 00000642076BU4 303695IR1 00000642193BU7 0004040462 00000646910BU0 314768SM10 00000647792BU1 904UFHNSP 00000646492BU9 XV386041GE8 00000646048BU9 S4160099 00000644611BU6 H2009VAS44 00000646493BU7 H0008VAS06 00000645027BU4 XV106184CX5 00000646491BU1 4184240 |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
800-437-1171
|
Manufacturer Reason for Recall |
Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
|
FDA Determined Cause 2 |
Device Design |
Action |
Consignees were sent on 9/4/2014 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref#12218 dated September 4, 2014. The letter was addressed to Director of Biomedical Engineering, Director of Radiology and Chief of Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. Customers are to perform a system functional check as instructed in the letter. If the check fails, they should contact their local GE Healthcare Service Representative. GE Healthcare will correct all systems at no cost to the customer, and a representative will contact customers to arrange for the correction. For questions they can call US 800-437-1171. For other countries they can call their local GE Healthcare Service Representative. |
Quantity in Commerce |
10 (4 US, 6 OUS). |
Distribution |
Worldwide Distribution -- US, including the states of CA, CO, FL, MO, PA; and, the countries of BELGIUM, CHINA, COLOMBIA, FRANCE, GREECE, MEXICO, and RUSSIA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IZI and Original Applicant = GE HEALTHCARE
|
|
|
|