| | Class 2 Device Recall Laksell Gamma Knife Perfexion |  |
| Date Initiated by Firm | October 06, 2014 |
| Date Posted | October 30, 2014 |
| Recall Status1 |
Terminated 3 on September 15, 2021 |
| Recall Number | Z-0152-2015 |
| Recall Event ID |
69425 |
| 510(K)Number | K120811 |
| Product Classification |
System, radiation therapy, radionuclide - Product Code IWB
|
| Product | Laksell Gamma Knife Perfexion |
| Code Information |
Serial No. 6052, 6144, 6088, 6122, 6158, 6005, 6028, 6097, 6153, 6064, 6081, 6103, 6120, 6143, 6145, 6213, 6215, 6076, 6070, 6136, 6002, 6090, 6131, 6063, 6160, 6077, 6018, 6031, 6040, 6105, 6121, 6013, 6026, 6038, 6050, 6089, 6042, 6055, 6058, 6071, 6080, 6087, 6091, 6096, 9098, 6106, 6108, 6109, 6116, 6118, 6125, 6132, 6133, 6142, 6148, 6149, 6161, 6162, 6175, 6181, 6183, 6198, 6205, 6206, 6163, 6169, 6033, 6041, 6046, 6068, 6113, 6019, 6067, 6110, 6104, 6043, 6044, 6062, 6085, 6086, 6101, 6111, 6123, 6126, 6157, 6177, 6178, 6024, 6069, 6094, 6007, 6075, 6182, 6065, 6159, 6199, 6003, 6017, 6027, 6039, 6112, 6146, 6194, 6004, 6006, 6008, 6009, 6010, 6011, 6012, 6014, 6015, 6016, 6020, 6021, 6022, 6023, 6025, 6029, 6030, 6032, 6034, 6035, 6036, 6037, 6045, 6047, 6048, 6053, 6054, 6056, 6057, 6060, 6061, 6066, 6072, 6073, 6074, 6078, 6079, 6082, 6083, 6084, 6092, 6093, 6095, 6099, 6100, 6102, 6107, 6114, 6115, 6117, 6119, 6124, 6127, 6128, 6129, 6130, 6134, 6135, 6139, 6141, 6150, 6151, 6165, 6167, 6171, 6172, 6173, 6176, 6179, 6180, 6184, 6185, 6191, 6192, 6200, 6203, 6207, 6211 |
| FEI Number |
1037831
|
Recalling Firm/ Manufacturer |
Elekta, Inc. 400 Perimeter Center Ter Ne Ste 50 Atlanta GA 30346-1227
|
| For Additional Information Contact | Linda Wetsel 770-300-9725 |
Manufacturer Reason for Recall | Elekta has become aware that the latches may be locked before they have been fully turned resulting in a poorly locked frame. |
FDA Determined Cause 2 | Device Design |
| Action | Elekta sent an" Important Field Safety Notice" distributed on October 6, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
This Important Field Safety Notice describes the problem and possible work arounds for customers to follow until a permanent solution can be developed. They are:
" Always assure that the plastic lever is only operated when at a right angle with the frame adapter.
" Do not force the plastic lever in place if it meets significant resistance when turned.
" Ensure that the plastic lever is completely flush with the frame adapter and that no angulations are present.
Corrective Action #2 - Elekta will check and replace as necessary affected frame adapters in the field. This will be released at the end of October 2014, and Elekta will have six months to complete the checks and replacements for those products in the field.
If you have any questions please contact your local Elekta office, or call (770) 300-9725. |
| Quantity in Commerce | 181 |
| Distribution | Worldwide Distribution - USA (nationwide) including the states of AK, AZ, AR, CA, CO, CT, FL, GA, HI, IL, KY, LA, MA, MI, MN, MS, MO, MT, NJ, NM, NY, NC, OH, OR, PA, RI, SC, TX, TN, UT, VA, WA and WV, and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Mexico, Morocco, Netherland, Norway, Poland, Portugal, Russia, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = IWB
|
|
|
|