Class 2 Device Recall Bausch & Lomb Stellaris PC w/ Laser, BL14304 |
 |
Date Initiated by Firm |
November 04, 2014 |
Date Posted |
December 23, 2014 |
Recall Status1 |
Terminated 3 on March 23, 2017 |
Recall Number |
Z-0862-2015 |
Recall Event ID |
69717 |
510(K)Number |
K133486
|
Product Classification |
Laser, ophthalmic - Product Code HQF
|
Product |
Bausch & Lomb Stellaris PC w/ Laser, BL14304 The responsible firm name is Bausch & Lomb Incorporated Rochester, NY |
Code Information |
Serial numbers: SPC01790, SPC01791, SPC01792, SPC01793, SPC01794, SPC01795, SPC01796, SPC01797, SPC01798, SPC01799, SPC01800, SPC01801, SPC01802, SPC01803, SPC01806, SPC01807, SPC01808, SPC01809, SPC01810, SPC01811, SPC01899, SPC01900, SPC01901, SPC01902, SPC01903, SPC01909, SPC01910, SPC01911, SPC01920, SPC01921, SPC01922, SPC01923, SPC01927, SPC01928, SPC01929, SPC01930, SPC01931, SPC01932, SPC01939, SPC01943, SPC01944, SPC01945, SPC01946, SPC01947, SPC01948, SPC01950, SPC01951, SPC01952, SPC01953, SPC01954, SPC01955, SPC01956, SPC01957, SPC01958, SPC01961, SPC01963, SPC01968, SPC01969, SPC01975, SPC01980, SPC01981, SPC01982, SPC01983, SPC01995, SPC01996, SPC01997, SPC02004, SPC02005, SPC02006, SPC02007, SPC02008, SPC02009, SPC02010, SPC02011, SPC02012, SPC02013, SPC02014, SPC02020, SPC02029, SPC02030, SPC02037, SPC02043, SPC02051, SPC02052, SPC02053, SPC02056, SPC02057, SPC02059, SPC02061, SPC02067, SPC02068, SPC02073, SPC02074, SPC02075, SPC02076, SPC02077, SPC02099, SPC02108, SPC02109, SPC02110, SPC02111, SPC02112, SPC02113, SPC02114, SPC02115 and SPC02120 |
Recalling Firm/ Manufacturer |
Bausch & Lomb Inc 3365 Tree Court Industrial Blvd Saint Louis MO 63122-6615
|
For Additional Information Contact |
Glenn M. Mattei 585-338-6839
|
Manufacturer Reason for Recall |
A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.
|
FDA Determined Cause 2 |
Software Manufacturing/Software Deployment |
Action |
Bausch & Lomb contacted customers by phone on November 4, 2014. Customers were informed of the conditions required for the device to malfunction. Customers were also notified that a software update to address the situation will be installed in their system within the next few weeks, and that the firm's service team will be scheduling a visit to each facility to install this software update. Customers were also told that in the meantime they may continue to use the device. For further questions please call (585) 338-6839. |
Quantity in Commerce |
106 |
Distribution |
Worldwide Distribution - USA Distribution including to the states of AL, CA, CT, FL, GA, HI, IL, IN, KS, MI, MN, MO, MS, NH, NY, OH, OK, PA, TX, WI and WV., and to the countries of Austria, Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Ireland, Italy, Jordan, Malaysia, Morocco, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, Switzerland, Turkey and the United Arab Emirates. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = HQF and Original Applicant = BAUSCH & LOMB
|
|
|
|