• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Getinge 46Series Medical WasherDisinfector

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Getinge 46Series Medical WasherDisinfectorsee related information
Date Initiated by FirmNovember 26, 2014
Date PostedDecember 23, 2014
Recall Status1 Terminated 3 on May 12, 2017
Recall NumberZ-0852-2015
Recall Event ID 69823
Product Classification Disinfector, medical devices - Product Code MEC
ProductGETINGE 46-SERIES Medical Washer-Disinfector, Models 46-4 and 46-5, with PACS 300/350 Control Series with dryer --- Class II, exempt from 510(k) requirements. For use in healthcare facilities and laboratories where reusable items such as instruments, utensils, anesthesia sets, and glassware are handled for decontamination purposes.
Code Information Serial Numbers: SEV0810077, SEV0537108, SEV0633084, W50030454, SEV0724143, SEV0724146, SEV0609110, W50044787, SEV0533068, W50042461, W50023731, W50023732, W50038744, W50028760, SEV0802081, W50015468, W50042473, W50045852, W50048201, W50041156, SEV0911139, SEV0832021, W50031347, SEV0538174, SEV0910190, SEV0734181, W50033518, SEV0802083, W50043877, W50015494, W50047828, SEV0633086, SEV0647025, SEV0650003, SEV0650004, SEV0534069, SEV0617130, W50011492, SEV0538177, W50010176, W50027826, SEV0743108, W50012356, SEV0614021, W50039525, SEV0619121, W50021052, W50025573, W50024233, W50010768, SEV0827061, SEV0738121, SEV0804079, SEV0806117, SEV0820063, SEV0820064, W50041988, SEV0727015, SEV0739202, SEV0906031, SEV0814068, SEV0805187, W50050381, W50050415, SEV0808424, SEV0626127, SEV0645102, SEV0710062, SEV0802086, SEV0815031, SEV0815032, W50027223, W50027224, SEV0826158, W50027541, W50043237, SEV0906032, W50012254, W50038229, W50014011, W50013832, W50012253, W50036280, W50036309, W50048577, SEV0537107, SEV0842075, SEV0750143, W50016365, SEV0808427, SEV0724139, SEV0724145, W50038722, SEV0908129, SEV0908130, SEV0649007, W50023502, SEV0842076, SEV0808429, SEV0837132, SEV0827051, W50027618, SEV0705085, SEV0705084, SEV0705083, W50026375, W50023804, W50014025, SEV0742115, W50028434, W50040964, W50019155, W50010908, W50013646, W50036143, W50045596, W50029312, W50029313, SEV0806118, W50028808, W50045860, W50014993, SEV0636055, SEV0738120, W50050491, W50028245, W50035227, W50048232, SEV0626230, W50023210, W50030137, SEV0808430, SEV0724144, SEV0534064, WAA053021, W50041067, W50030350, SEV0540080, W50045855, W50032192, W50029835, W50031507, SEV0739208, SEV0802085, SEV0622093, SEV0650002, SEV0619122, W50016086, W50039500, SEV0543058, SEV0910191, SEV0837135, W50018212, SEV0537103, WAA052845, SEV0605066, W50034170, SEV0524337, W50011160, W50011491, W50032801, W50015395, W50042391, SEV0547204, W50045890, W50047809, W50013644, W50013645, W50030121, W50021208, SEV0626228, SEV0851034, SEV0827047, SEV0552041, W50037851, SEV0833107, SEV0833108, SEV0833109, SEV0833110, SEV0833111, W50036056, W50036081, SEV0837131, W50015490, SEV0716176, SEV0716177, W50043774, W50023888, SEV0738132, SEV0827050, SEV0724140, W50015641, W50045601, W50024979, SEV0614035, W50016223, W50025076, SEV0815030, SEV0820065, W50016368, SEV0647023, SEV0909043, W50037190, SEV0642143, W50017770, SEV0537110, SEV0543054, SEV0538178, W50023488, SEV0614036, SEV0614037, SEV0533067, W50029860, W50029036, SEV0814067, SEV0851035, SEV0851036, W50013647, SEV0738001, SEV0739204, SEV0810081, SEV0533069, W50016222, SEV0906030, W50014012, W50019465, W50038218, SEV0521190, W50038730, SEV0826159, W50012251, SEV0738133, SEV0619119, SEV0649010, SEV0619120, SEV0614022, W50050577, W50041166, W50048183, W50017921, SEV0739205, SEV0749123, W50042459, SEV0524338, SEV0820049, SEV0640030, W50030144, W50030181, W50028247, W50014027, W50032581, W50013666, W50042946, W50045884, W50050574, W50013831, SEV0618050, SEV0618051, SEV0727034, SEV0534074, SEV0716188, W50010079, W50033538, W50047210, W50019156, W50032618, W50015495, W50036207, SEV0733024, W50021211, SEV0827056, W50043040, W50028248, W50028468, SEV0605063, SEV0815025, SEV0824064, SEV0808426, W50036147, W50017918, W50018215, W50030353, W50036140, WAA053036, SEV0724142, SEV0533065, W50035801, SEV0907057, W50045822, SEV0836007, W50028320, W50010906, W50028809, W50023209, W50048247, W50027514, SEV0739206, SEV0739209, SEV0626231, SEV0738131, W50038686, W50011489, W50011490, W50046986, SEV0633083, W50021142, SEV0907054, W50021143, SEV0716048, W50023897, SEV0907055, W50011159, SEV0647022, SEV0851037, SEV0907056, W50021141, SEV0826157, SEV0848144, SEV0621158, W50047960, SEV0633088, SEV0705086, SEV0727019, SEV0626121, SEV0626229, W50010177, SEV0738122, SEV0738123, SEV0521184, W50016376, SEV0633087, W50019467, W50015466, W50017920, W50033104, W50025852, W50016366, SEV0632078, SEV0543059, W50048551, W50035559, SEV0521191, SEV0515157, SEV0515159, SEV0548037, SEV0644133, SEV0727032, SEV0820048, SEV0844215, SEV0906028, SEV0906029, W50010905, W50015493, W50037099, W50037204, W50039512, W50042471, W50042478, W50050571,  W50050573, W50050578, WAA053175, SEV0618048, SEV0545170, SEV0724147, SEV0632079, SEV0827060, SEV0743107, SEV0820057, SEV0826160, W50029023, SEV0524334, W50011161, W50022983, W50045665, W50048002, W50012252, W50018216, W50018928,  W50019154, SEV0734183, W50050575, SEV0820050, W50027512, SEV0716185, SEV0820058, W50041125, SEV0605064, SEV0615074, SEV0649008, SEV0622094, W50016364, SEV0737120, W50038421, SEV0824063, W50017919, SEV0815033, W50033839, SEV0626126, SEV0808428, W50015496, SEV0533071, W50023733, SEV0716184, SEV0742118, SEV0727033, SEV0649009, SEV0810078, W50039166, SEV0618049, W50038256, SEV0536194, SEV0716183, SEV0613001, W50042222, W50043233, SEV0636042, SEV0636043, W50039286, W50017771, SEV0804077, SEV0633085, W50021356, W50021357, SEV0532062, W50020918, W50030145, W50038244, W50036145, SEV0626120, SEV0716175, W50023501, SEV0733032, SEV0734182, SEV0738002, W50010907, W50037207, SEV0708018, W50048219, W50014030, SEV0626234, SEV0738130, W50042491, SEV0543056, SEV0536195, SEV0543055, W50013311, W50028993, SEV0739203, SEV0613098, SEV0739207, SEV0647024, W50035975, SEV0708019, SEV0702017, SEV0805191, SEV0521192, SEV0910192, W50041173, SEV0842074, W50021214, SEV0723112, W50037115, SEV0833100, SEV0833101, SEV0833102, SEV0833103, SEV0833104, W50017914, W50024126, W50029644, SEV0637017, SEV0605065, SEV0724138, W50037201, SEV0536197, W50045549, W50048239, W50011158, SEV0834133, SEV0708020, SEV0710061, SEV0849100, W50017917, W50030122, W50023198, W50018213, W50018214, SEV0533070, SEV0612048, SEV0632080, SEV0727018, SEV0537104, SEV0524333, W50024980, SEV0808423, SEV0734180, W50024130, W50024131, SEV0626125, W50035799, W50015488, SEV0827046, SEV0613100, SEV0615042, SEV0637018, SEV0543057, SEV0810080, W50021145, SEV0548036, SEV0606022, W50028246, W50028112, W50028113, W50027222, W50033877, SEV0908131, W50029035, SEV0827045, W50024981, SEV0521186, SEV0538173, SEV0827057, W50013312, SEV0626232, SEV0613099, W50016367, SEV0626233, W50018929, SEV0911140, SEV0911141, W50015398, SEV0521187, W50041062, SEV0524327, WAA052236, SEV0910189, W50021210, W50030183, SEV0804078, SEV0808425, SEV0626119, SEV0724141, W50047838, W50015491, SEV0643166, SEV0802082 --- Canadian Units: SEV0538175, SEV0613101, W50030182, W50030412.
Recalling Firm/
Manufacturer
Getinge Disinfection Ab
Ljungadalsgatan 11
Vaxjo Sweden
For Additional Information ContactMs. Ann G. Wheeler
585-272-5036
Manufacturer Reason
for Recall
Getinge Disinfection AB received complaints regarding sediment residuals in the manifold of the washer and disinfector of Getinge 46-Series (CM302) with PACS 300/350 control. After performing detailed root cause analysis, the firm has identified the problem as insufficient water flow throughout the manifold due to formation of air pockets where sedimentation may occur.
FDA Determined
Cause 2
Device Design
ActionAn Urgent Field Safety Notice was sent to consignees in North America via USPS Certified Mail with signature receipt on 11/25/14 informing them of the correction. The actions that will be carried out will fully address the problem of sedimentation i.e. stop any future sedimentation. You will be contacted by a Getinge representative to schedule a visit. No additional precautions need to be taken by the end user until the Getinge representative arrives. For questions, please contact Getinge USA at 1-800-320-3532.
Quantity in Commerce551 total: USA: 547 units, Canada: 4 units
DistributionDistributed US (nationwide) and the country of Canada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-