| Class 2 Device Recall PKG, 5MM INSERT, TENACULUM, 33CM, |  |
Date Initiated by Firm | November 17, 2014 |
Date Posted | December 12, 2014 |
Recall Status1 |
Terminated 3 on July 17, 2015 |
Recall Number | Z-0726-2015 |
Recall Event ID |
69838 |
Product Classification |
Endoscope and/or accessories - Product Code KOG
|
Product | PKG, 5MM INSERT, TENACULUM, 33CM, P/N 0250080698
Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. |
Code Information |
44485 341728 541773 543282 742876 745131 845497 847458 940778 942549 942646 943423 943739 944093 944094 945232 946004 1040629 1040994 1041614 1042529 1043032 1044517 1046583 1140148 1141926 1142615 1144486 1145071 1145795 1146118 1147099 1240111 1240545 1241176 1242120 1242928 1244073 1244707 1246143 1340125 1343307 1343553 1345287 1345526 1345630 1346245 1346662 1440422 1440461 1441380 1442253 |
Recalling Firm/ Manufacturer |
Stryker Endoscopy 5900 Optical Ct San Jose CA 95138-1400
|
For Additional Information Contact | 408-754-2000 |
Manufacturer Reason for Recall | The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods |
FDA Determined Cause 2 | Error in labeling |
Action | Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below:
www.stryker.com/IFU1000401070
Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314.
Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com.
For questions regarding this recall call 408-754-2000. |
Quantity in Commerce | 180,573 units total |
Distribution | Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela. |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|