  Class 2 Device Recall PKG, 3MM INSERT, FUNDUS GRASPER, 20CM, see related information
Date Initiated by Firm November 17, 2014
Date Posted December 12, 2014
Recall Status1 Terminated 3 on July 17, 2015
Recall Number Z-0761-2015
Recall Event ID 69838
Product Classification Endoscope and/or accessories - Product Code KOG
Product PKG, 3MM INSERT, FUNDUS GRASPER, 20CM, P/N 0250282003
Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Code Information 343779 344866 747275 942145 942647 942684 942770 945466 1043503 1246165 
Recalling Firm/
Manufacturer
Stryker Endoscopy
5900 Optical Ct
San Jose CA 95138-1400
For Additional Information Contact
408-754-2000
Manufacturer Reason
for Recall
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
FDA Determined
Cause 2
Error in labeling
Action Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
Quantity in Commerce 180,573 units total
Distribution Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-