| Class 2 Device Recall Sysmex XN10 and XN20 Hematology Analyzers | |
Date Initiated by Firm | December 18, 2014 |
Date Posted | June 10, 2016 |
Recall Status1 |
Terminated 3 on July 25, 2016 |
Recall Number | Z-1945-2016 |
Recall Event ID |
70054 |
510(K)Number | K112605 |
Product Classification |
Counter, differential cell - Product Code GKZ
|
Product | The XN-Series modules (XN-10, XN-20) are quantitative multi-parameter automated hematology analyzers intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. |
Code Information |
1) Model: XN-10; Catalog Number: AP795756; Serial Numbers: 111001-15483 except 15287, 15298, 15308, 15408, 15442, 15454, 15470 and 15484; 2) Model: XN-20; Catalog Number: AE797961; Serial Numbers: 11001-12118 except 11960, 12039, 12078, 12114, and 12117. |
Recalling Firm/ Manufacturer |
Sysmex Corporation 1 5 1 Wakinohamakaigandoori CHUO-KU Kobe-shi Japan
|
For Additional Information Contact | Sysmex Technical Assistance Center 888-879-7639 |
Manufacturer Reason for Recall | There is a potential for elevated hemoglobin (Hgb) and impedance platelet count (PLT-I) values. |
FDA Determined Cause 2 | Device Design |
Action | Sysmex Corporation sent a Product Notification Field Correction letter were to all affected customers of record. The letters included instructions for customers to: 1) Review results with an unexpected increase in Hgb, MCH, MCHC, and/or PLT-I. The increase may result in a low or normal value being artificially elevated to either a normal result or a result above the reference ranges, or a high value being artificially increased. There is negligble impact to the RBC when the issue occurs; 2) The relationship with an elevated Hgb would cause the indices (MCH and MCHC) values to appear to be abnormally elevated and may result in the results being flagged for verification or indicated to be outside of reference ranges; 3) Review any "Turbidity/Hgb Interf?" flag which will automatically trigger at an MCHC of
37.5 g/dL; 4) If Hgb is affected, the PLT-I may be also. Review any PLT-I result when an unexpected increase in MCH/MCHC or "Turbidity/Hgb Interf?" flag occurs; 5) Gross interference is likely to trigger a flag, however there is a potential any interference could increase the Hgb or the PLT-I count from a lower level to an "normal" level. Follow laboratory protocol for confirming unexpected results. Customers were further advised that they will be contacted by a Sysmex Field Representative within 180 days to schedule an appointment to replace the rinse cup on the analyzer(s) and that version 00-16 software will be installed to change the rinsing sequence for the aspiration probe.
Customers with questions are requested to e-mail the Sysmex Technical Assistance Center. Urgent questions can be answered by calling the Technical Assistance Center at 888-879-7639 (24 hours a day / 7 days a week). |
Quantity in Commerce | 1,137 units (US) 54 units (Canada) 9 inventory |
Distribution | Worldwide Distribution - US to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WY + District of Columbia + Guam; Internationally to Australia, Bahrain, Bangladesh, Cambodia, Canada, China, Germany, Hong Kong, India, Indonesia, Malaysia, Maldives, Mongolia, Mynamar, Nepal, New Zealand, Pakistan, Philippines, Singapore, South Korea, Sri Lanka, Taiwan, Thailand, and Vietnam.
THE CORRECTION IS LIMITED TO XN-10/XN-20 UNITS DISTRIBUTED BY SYSMEX AMERICA, INC., INSTALLED AT END-USER SITES IN US, PUERTO RICO, AND CANADA. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = GKZ
|
|
|
|