| Class 2 Device Recall Siemens ADVIA Centaur XP and ADVIA Centaur CP Progesterone Assay | |
Date Initiated by Firm | January 27, 2015 |
Date Posted | March 02, 2015 |
Recall Status1 |
Terminated 3 on March 31, 2016 |
Recall Number | Z-1231-2015 |
Recall Event ID |
70470 |
510(K)Number | K932955 |
Product Classification |
Radioimmunoassay, progesterone - Product Code JLS
|
Product | ADVIA Centaur Systems Progesterone Kit, (5-pack Ref); In Vitro Diagnostic
Catalog number: 02177364; SMN: 10333111. |
Code Information |
Lot Numbers: 42590268 43742268 44847268 Exp. Date: May 22, 2015 |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc 333 Coney St East Walpole MA 02032-1516
|
For Additional Information Contact | 508-668-5000 |
Manufacturer Reason for Recall | ADVIA Centaur Systems Progesterone Kit Lots Ending in 268 have a high bias on results greater than 30 ng/mL
(95.4 nmol/L) |
FDA Determined Cause 2 | Process control |
Action | Siemens issued an Urgent Medical Device Correction (UMDC) was sent to affected Siemens Healthcare Diagnostics customers in the United States on January 27, 2015 to be delivered to customers on January 28, 2015. An Urgent Field Safety Notice (U FSN) was sent out for distribution to all customers outside the United States on January 27, 2015. These notices inform the customer of the potential to observe a high bias on patient sample with results greater than 30 ng/ml when using ADVIA Centaur Systems Progesterone kit lots ending in 268. To mitigate the bias, Siemens instructs customers to dilute samples with values greater than 30 ng/ml when using ADVIA Centaur Systems Progesterone kit lots ending in 268. In addition, Siemens informs customers to continue to follow Urgent Medical Device Correction 10819675, Rev. A and Urgent Field Safety notice 10819674, Rev A (Information Regarding the ADVIA Centaur Systems Calibrator E), dated September 2014.
Fax-back forms to be returned to confirm that the customers received the
Urgent Medical Device Correction. |
Quantity in Commerce | 1719 kits |
Distribution | Worldwide Distribution-US (nationwide), Austria, Bahrain, Belgium, Bulgaria, Brazil, Canada, Chile, Colombia, Croatia, Cyprus, Czech, Republic, Denmark, Egypt, Estonia, France, Georgia, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kazakhstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Serbia, Slovakia, Spain, South America, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and Vatican. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JLS
|
|
|
|