| Class 2 Device Recall GE Cabinet Xray system |  |
Date Initiated by Firm | January 13, 2015 |
Date Posted | March 20, 2015 |
Recall Status1 |
Terminated 3 on August 17, 2015 |
Recall Number | Z-1211-2015 |
Recall Event ID |
70478 |
| |
Product Classification |
Cabinet x-ray, industrial - Product Code RCE
|
Product | GE Inspection Technologies x|tome|x L240 an Cabinet X-ray system |
Code Information |
SN - PA1260 |
Recalling Firm/ Manufacturer |
GE Inspection Technologies, LP 50 Industrial Park Rd Lewistown PA 17044-9312
|
For Additional Information Contact | 717-447-1278 |
Manufacturer Reason for Recall | It was discovered that the upper left corner-shaped lead shield in the window in one of the doors of the v|tome|x L240 installed had lost adhesion and fall out of position, which could lead to gaps in shielding. |
FDA Determined Cause 2 | Device Design |
Action | GE Inspection Technologies, Planned Action:
1. GE will notify customers of the affected cabinet x-ray systems through a Technical Information Letter. GE notified its affected customers via email on January 13, 2015, via letter dated January 14, 2015, and via telephone on January 19, 2015.
2.GE will install an interim insert within 30 days of receipt of this letter. The insert will cover any existing or interim gaps in the original lead shielding on the door.
3. GE will replace the doors of affected cabinet x-ray systems no later than 180 day of receipt of this letter.
4. These corrections will be made free of charge.
The Technical Information Letter used for notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter
For further questions, please call (717) 447-1278 |
Quantity in Commerce | 2 |
Distribution | US Distribution to the states of OH and PA. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|