  Class 2 Device Recall Zimmer see related information
Date Initiated by Firm December 17, 2014
Date Posted February 20, 2015
Recall Status1 Terminated 3 on September 23, 2015
Recall Number Z-1136-2015
Recall Event ID 70401
510(K)Number K974728  
Product Classification Prosthesis, hip, semi-constrained (metal uncemented acetabular component) - Product Code KWA
Product Metasul Femoral Head - indicated for total hip arthroplasty for treatment of patient conditions of non-inflammatory degenerative joint disease.
Code Information Item numbers: 734028800  Lot #'s:  61487614
Recalling Firm/
Manufacturer
Zimmer Gmbh
Sulzer Allee 8
Winterthur Switzerland
For Additional Information Contact Consumer Relations Call Center
800-447-5633
Manufacturer Reason
for Recall
Potential compromise of the inner sterile tray. If a compromised tray is detected, a delay in surgery may result. If undetected breach in both the inner and outer trays were to occur, and there is contamination present, a periprosthetic infection is possible, which could lead to revision.
FDA Determined
Cause 2
Package design/selection
Action On 12/17/2014, URGENT MEDICAL DEVICE RECALL notifications were sent to the affected distributors and hospital Risk Managers with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-800-348-2759.
Quantity in Commerce 16
Distribution Domestic: AK, AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MS, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY International: Canada, Japan
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = KWA and Original Applicant = SULZER ORTHOPEDICS, INC.
-
-