| | Class 2 Device Recall VERTEX EPIC MCD, GENESYS II |  |
| Date Initiated by Firm | March 13, 2015 |
| Date Posted | May 12, 2015 |
| Recall Status1 |
Terminated 3 on June 26, 2018 |
| Recall Number | Z-1598-2015 |
| Recall Event ID |
71190 |
| 510(K)Number | K922080 K952684 |
| Product Classification |
System, tomography, computed, emission - Product Code KPS
|
| Product | Nuclear Gamma Cameras (ADAC Vertex Classic, ADAC VERTEX PLUS/SOLUS/CARDIO, ADAC VERTEX V60) Philips Medical Systems, Cleveland, OH |
| Code Information |
Vertex Classic; Model #: 890122; Serial #s: 9404100, 9601238, 9609084, 0407V, 0682V, 0795V, 0822VP2, 103, 20, 500, 5405V, 64, 9308007, 9309020, 9310032, 93-10660-1292, 9401059, 9401061, 9403076, 9403083, 9404106, 9407122, 9409148, 9410002, 9411006, 9411007, 9411016, 9501049, 9501210, 9502067, 9502086, 9503003, 9503079, 9503109, 9504019, 9505021, 9505124, 9506140, 9508169, 9508171, 9508184, 9508193, 9509198, 9509206, 9510222, 9510226, 9601146, 9602249, 9608042, 9901004, B0106154, B02120033, M@10229AMT, M@207JD, M@4404V, M@4431V, M@4432V, M@4480V, M@4623V, M@4630V, M@4639V, M@4686V, M@4692V, M@4858V, M@4893V, M@4959V, M@5032V, M@5033V, M@5038V, M@5105V, M@5145V, M@5158V, M@5189V, M@5280V, M@5289V, M@5338V, M@5360V, M@5373V, M@5398V, M@5407V, M@5412V, M@5431V, M@5439V, M@5467V, M@5476V, M@5497V, M@5498V, M@5515V, M@5526V, M@5590V, M@5617V, M@5627V, M@5661V, M@5662V, M@5663V, M@5665V, M@5681V, M@5688V, M@5689V, M@5751V, M@5752V, M@5782V, M@5811V, M@5835V, M@5869V, M@5900V, M@5913V, M@5915V, M@5960V, M@5970V, M@6004V, M@6007V, M@6010V, M@6056V, M@6900V, M@9304RV, M@9645RV, M@MVS5236V, M@MVS5852V, M@O1325, UNK-AMT1238A. VERTEX PLUS/SOLUS/CARDIO: Model #: 88297; Serial #s: 9603015, 9605036, 9610076, 9610077, 9703171, 9706192, 9710103, 9711277, 9802304, 9803109, 9807121, 9808122, 9809056, 9809124, 9809415, 9809424, 9912461, _0002032, ~9608052, 0007001M, 0008009B, 0011002B, 0102019V, 0103020V, 01069VM, 01070VM, 0111001V, 01295VM, 01296K, 01864VM, 03026VM, 0939VM, 100098-01N, 1006, 1008, 10252, 1027, 1037, 10630, 1092, 1097, 11, 130234-01NVERTEX, 2035, 2038, 2047, 3059, 3060, 3064, 4007, 4078, 5934V, 6052, 6261, 7373VM, 7376VM, 7386VM, 7560VM, 7561VM, 8002, 802080322, 85247, 9054VM/V, 9056VM/V, 9073152, 9130VM, 9205345, 940384, 9408136, 9503084, 9504096, 9506016, 9506145, 9507163-CN, 9511123, 9512008, 9512009, 9512018, 9601015, 9601036, 9601240-CN, 9602161, 9602245, 9602246, 9602254, 9603336, 9604016, 9604021, 9605026, 9605037, 9606211, 9607015, 9607024, 9608046, 9608049, 9608050, 9608052, 9608064, 9609063, 9609081, 9610058, 9611003, 9611077, 9611082, 9612072, 9701127, 9702061-CN, 9702095, 9702150, 9703154, 9704044, 9704178, 9704183, 9705004, 9705186, 9706005, 9706182, 9706189, 9707002, 9707047, 9707050, 9707054, 9708062, 9708065, 9708068-CN, 9708069, 9708076, 9708078, 9708084, 9708227, 9709022, 9710121, 9710127, 9710244, 9711025, 9711034, 9711036, 9711074, 9711078, 9711261, 9711271, 9801047, 9801227, 9801232, 9801297, 9801299-CN, 9802050, 9802255, 9802260, 9802300, 9802306, 9802397, 9803003, 9803014, 9803067, 9803083, 9803269, 9803272, 9803288, 9803290, 9803293, 9803295, 9803312, 9803316, 9803322, 9804069, 9804330, 9804333, 9804342, 9805344, 9805350, 9806006, 9806014, 9806363, 9806377, 9806388, 98066002, 9807002, 9807391, 9807421, 9808036, 9808090, 9808108, 9808451, 9808452, 9808457, 9808480, 9808829, 9809062 (-CN), 9809065-CN, 9809067, 9809113, 9809122, 9809125, 9809498, 9809687, 9810075, 9810136, 9810428ADAC#7973VM, 9810442, 9810551, 9810557, 9811020, 9811093, 9811569, 9811737, 9812582, 9901597, 9901605, 9902150, 9902622, 9902630, 9903632, 9903637, 9903641, 9904089, 9905012, 9905659, 9905667, 9906670, 9907016, 9907017, 9907019, 9907029, 9909161, 9910031, 9911458, 9911698, 9911702, 9912181, 9912463, AD1261, AMT00848VM, B0006129, B0007273, B0008003, B0008012, B0008021, B0008030, B001200212, B0105138, B0106169, B01070191, B01070192, B0107179, B01080204, B01080207, B01090227, B01100231, B01100233, B01100238, B01110251, B02050071, B02050262, B02060167, B02060270, B02060279, B0206027S, B02070290, B02070300, B0207296, B02080318, B0208312, B02090327, B02090333, B02090334, B02100345, B03010355, B03010359, B03020048, B03080380, C0008070, C01080152, C01080154, C03050230, C03060193, M@10458, M@10573, M@10582, M@10589, M@10772, M@10837, M@10850, M@10927, M@10967, M@11020, M@11036, M@11109, M@11140, M@11178, M@11498, M@11525, M@11571, M@11590, M@11881, M@11949, M@11957, M@11972, M@12219, M@12412, M@12591, M@12711, M@12787, M@13010, M@13296.M@13466, M@13498, M@13543, M@13549, M@13665, M@13688, M@13691, M@13748, M@13813, M@22752, M@28051, M@4750019856, M@5828J, M@5906N, M@5919J, M@5936N, M@5956J, M@5961K, M@6002J, M@6012J, M@6075VM, M@6108VA, M@6123VA, M@6160VM, M@6183VA, M@6258J, M@6323VM, M@6324J, M@6330VA, M@6341VA, M@6345VA, M@6346VM, M@6348VM, M@6349VM, M@6365VM, M@6368J, M@6370VA, M@6375J, M@6376J, M@6460VM, M@6464VM, M@6470VA, M@6481VA, M@6496VM, M@6528VA, M@6541VA, M@6542VM, M@6566VM, M@6573VM, M@6596VM, M@6600VA, M@6638VM, M@6639J, M@6675VA, M@6716VM, M@6717VM, M@6762VM, M@6773VM, M@6776VM, M@6797K, M@6805VM, M@6823VM, M@6840K, M@6841VM, M@6867VM, M@6877VM, M@6886VM, M@6891VA, M@6924VM, M@6931VM, M@6932VM, M@6943VM, M@6946VM, M@6947J, M@6985J, M@7003VA, M@7004VA, M@7005VA, M@7017VM, M@7024VM, M@7034K, M@7042VM, M@7053VM, M@7073VM, M@7085K, M@7086J, M@7095VM, M@7100VM, M@7101VM, M@7102VM, M@7151VM, M@7153VM, M@7159VM, M@7166K, M@7171VM, M@7172K, M@7181J, M@7190VM, M@7198VA, M@7215K, M@7216VM, M@7217VM, M@7221VA, M@7222VA, M@7229VM, M@7233VM, M@7237VM, M@7245J, M@7246VM, M@7247VM, M@7250VM, M@7300J, M@7308VM, M@7376VM, M@7380VM, M@7384VM, M@7390VM, M@7393VM, M@7394VM, M@7410VM, M@7420VM, M@7464VA, M@7470VM, M@7490VM, M@7494K, M@7501VM, M@7504VM, M@7505VM1, M@7514VM, M@7515VM, M@7536VM, M@7537VM, M@7538VM, M@7556VM, M@7557VM, M@7569VM, M@7572VM, M@7591VM, M@7592VM, M@7603K, M@7605VM, M@7606VM, M@7623K, M@7626K, M@7627K, M@7628VA, M@7638VM, M@7643VM, M@7654VM, M@7664VM, M@7665K, M@7673VM, M@7674K, M@7679VM, M@7682K, M@7685VM, M@7691VM, M@7695VM, M@7697VM, M@7723VM, M@7729VM, M@7733VM, M@7739K, M@7760VM, M@7761VM, M@7770VM, M@7776VM, M@7789K, M@7801VM, M@7810VM, M@7811VM, M@7812VM, M@7819VA, M@7823K, M@7824VM, M@7830K, M@7855J, M@7857VM, M@7860VM, M@7861VM, M@7874VM, M@7877VM, M@7918K, M@7924K, M@7925K, M@7929VM, M@7930VM, M@7932VM, M@7946J, M@7957VM, M@7961VM, M@7962VM, M@7965VM, M@7968VM, M@7975K, M@7987VM, M@7995VM, M@7997VM, M@8004VM, M@8013VM, M@8018J, M@8021VM, M@8044VM, M@8060VM, M@8064VM, M@8076K, M@8089VM, M@8096VM, M@8105VM, M@8124VA, M@8124VA, M@8146VM, M@8155VM, M@8178VM, M@8203VM, M@8219VA, M@8220K, M@8240K, M@8241VM, M@8248VM, M@8252VM, M@8284K, M@8301K, M@8302VM, M@8306K, M@8337VM, M@8356K, M@8380VM, M@8399VM, M@8484VM, M@8528VM, M@8537VM, M@8551VM, M@8561K, M@8606K, M@8609VM, M@8651VM, M@8683VM, M@8686K, M@8693VM, M@8726VM, M@8738K, M@8759K, M@8771VM, M@8781VM, M@8797K, M@8799RV, M@8800VM, M@8803K, M@8812VM, M@8814VM, M@8882K, M@8888K, M@8897K, M@8903K, M@8906RK, M@8920VM, M@8922VM, M@8935K, M@8948RV, M@8955K, M@8990K, M@8996VM, M@9032K, M@9046K, M@9049VM, M@9081K, M@9091VM, M@9107RV, M@9166VM, M@9185K, M@9194VM, M@9221VM, M@9222VM, M@9247RV, M@9252VM, M@9267VM, M@9309RV, M@9316VM, M@9327K, M@9338VM, M@9340VM, M@9360VM, M@9369VM, M@9418VM, M@9426K, M@9438K, M@9471RV, M@9515VM, M@9532VM, M@9542VM, M@9543K, M@9583VM, M@9607VM, M@9653VM, M@9672VM, M@9724VM, M@9730K, M@9816K, M@9831VM, M@AMT1258, M@MCD6416VM, M@MCD7036VM, M@R372820, M@S5195VA, M@S5365-V, M@S5820VM, M@S5907V11, M@S6155VA, M@S7156VM, M@S7223K, M@S7559VA, M@S7571VM, M@S7856K, M@S7864K, M@S7904VM, M@S7941VA, M@S8687K, M0101023, R03010363, S0011003, S6235VM, STOR_QUO_47521360, UNK-6597VM, UNK-7590VM, UNK-8533VM1, UNK-8860VM, UNK-9868K, V0007070, V0010090, V0012099, V0104109, V0104110, V0105112, V0106121, V0106123, V0106124, V01070133, V01090146, V01100149, V01100151, V01100156, V02050166, V02070174, V02070179, V02080190, V02080194, V02080199, V02080200, V02090206, V03060224, V03070229, V03091241. VERTEX V60 Model #: 882030; Serial #s: 9601243, 9602253, 9603008, 9603013, 9604028, 9604041, 9605033, 9606038, 9610256, 9704175, 9704177, 9801284, 9803324, 9807089, 9810434, 9904005, 970001-9610187, M0007018, V0009076, V0011096, V0106122, V0106125, V02050163, V02050165, V02070181, 9705007, 9711262, 00559VM, 00634V, 00657VA, 00970VA, 01190VM, 01697VM, 01894VM, 0371V, 10588, 1189-17, 12001, 1273-06, 1274-06, 9601004, 9607063, 9707206, 9710247, 9712001, 9712008, 9712009, 9803003, 9806019, 9808021, 9902027, 9909039, M0007001-CN, V0005062, V01090141, V01090142, V03080235, V03080236, V2070177, VETEJ, X0005002, X0010004, X0101006, X0101007, X03030012, Z00657VA, Z01798VM. |
Recalling Firm/ Manufacturer |
Philips Medical Systems (Cleveland) Inc 595 Miner Rd Cleveland OH 44143-2131
|
| For Additional Information Contact | Mr. Yuchol Kim 440-483-2997 |
Manufacturer Reason for Recall | During clinical use of a Vertex Plus Gamma Camera, as the detector heads were in the relative 180 degree position, Detector #1 drifted down and made contact with the patients chin during a whole body study acquisition. |
FDA Determined Cause 2 | Employee error |
| Action | Philips sent an Customer Information letter dated March 16, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Customers were instructed to :
1. Do not lubricate the clutch(es) with oil, grease or any other type of lubricant..
2. Follow product Instructions For Use (IFU), CARDIO. SOLUS, ADAC VERTEX, AND ADAC VERTEX PLUS WITH ATLAS USERS MANUAL, 9201-0227E-ENG, REV A- General Precautions that instructs user to monitor the patient during scanning.
3. Follow product Instructions For Use (IFU), CARDIC, SOLUS, ADAC VERTEX, AND ADAC VERTEX PLUS WITH ATLAS USER'SMANUAL, 9201-0227E-ENG, REV A- Emergency Patient Removal, Section Safety and Regulatory Agency Compliance, page 2-11, instruction(s) on emergency
patient removal.
4. Ensure completion of product Preventive Maintenance (PM) procedure requiring inspection of the radius drive clutches every 3 months for contamination and replacement if contaminated and, testing of the gantry system.
If you need any further information or support concerning this issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377. |
| Quantity in Commerce | 918 units |
| Distribution | Worldwide Distribution - US (nationwide) including Puerto Rico and to the states of : AL, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI and WV., and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Costa Rica, Denmark, Egypt, France, Germany, Ireland, Italy, Japan, Lebanon, Malaysia, Netherlands, Norway, Pakistan, Portugal, Republic of Korea, Russia, South Korea, Spain, Sweden, Syrian Arab Republic, Thailand and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = KPS
|
|
|
|